Analysis of the interventricular pressure waveform in the moving-actuator total artificial heart

被引:8
作者
Jo, YH
Choi, JS
Kim, WE
Park, JW
Choi, WW
Kim, HC
Kim, WG
Ahn, H
Rho, JR
Min, BG
机构
[1] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Inst Med & Biol Engn,Med Res Ctr,Chongno Gu, Seoul, South Korea
[2] Seoul Natl Univ, Coll Med, Biol Engn Major Interdisciplinary Program, Seoul, South Korea
[3] Seoul Natl Univ, Coll Med, Dept Thorac & Cardiovasc Surg, Seoul, South Korea
关键词
Actuators - Blood vessels - Pacemakers - Pressure - Sensors;
D O I
10.1097/00002480-200011000-00018
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Right and left filling pressures are important parameters in the automatic control of a total artificial heart (TAH) within normal physiologic ranges. Our TAH is composed of a moving actuator, right and left ventricles, and an interventricular space (IVS) enclosed by a semirigid housing. During operation of the TAH, the IVS volume is changed dynamically by the difference between the ejection volume of one ventricle and the inflow volume of the other. We measured the interventricular pressure (IVP) waveform by using a pressure sensor and analyzed the relationship between the IVP and the preload condition. From in vitro and in vivo experiments, we found that the measured filling pressures were linearly related to the negative peak value of the IVP. Additionally, we found that we could use the time interval from actuator start to the positive peak Value of the IVP (outflow valve opening) as a useful parameter to estimate the blood filling volume of the diastole ventricle.
引用
收藏
页码:749 / 755
页数:7
相关论文
共 7 条