The influence of the axial, antero-posterior and lateral positions of the center of rotation of a ball-and-socket disc prosthesis on the cervical spine biomechanics

被引:37
作者
Galbusera, Fabio [1 ]
Anasetti, Federica [1 ]
Bellini, Chiara Maria [1 ]
Costa, Francesco [1 ]
Fornari, Maurizio [1 ]
机构
[1] IRCCS Ist Ortoped Galeazzi, I-20161 Milan, Italy
关键词
Cervical biomechanics; Cervical disc replacement; Finite element; Total disc arthroplasty; Probabilistic model; FINITE-ELEMENT; ARTHROPLASTY; FUSION; IMPLANTATION; REPLACEMENT; DISKECTOMY; KINEMATICS;
D O I
10.1016/j.clinbiomech.2010.01.010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Previous studies documented the importance of the positioning and the design parameters of the prosthesis in determining the biomechanics of the implanted spine However, a comprehensive biomechanical evaluation of the significance of these parameters is still lacking. Therefore, the paper is aimed to the quantification of their influence on the flexibility of the implanted spine and the force transmitted through the facet Joints. Methods: A finite element model of the C5-C6 spine unit including a ball-and-socket disc prosthesis was built Three probabilistic variables were considered: the axial, antero-posterior and lateral positions of the center of rotation. Randomized input parameters were generated with the Monte Carlo method. Pure moments of 1.6 Nm in flexion, extension, lateral bending and axial rotation were imposed to the upper endplate of C5: 100 simulations were conducted for the each of the considered loading conditions Findings Axial position of the center of rotation influenced the spine flexibility in all loading conditions and the facet force in extension, lateral bending and axial rotation The antero-posterior position was found to influence the spine flexibility in flexion and extension, and the facet force in lateral bending and axial rotation. The lateral position was not significant. Interpretation: The effects of the positioning of a cervical disc prosthesis were estimated. A wide range of mechanical behaviors can be obtained by the manufacturers by appropriately manipulating the position of the center of rotation. A proper positioning of the artificial disc during the surgery, in particular in the antero-posterior direction, was found to be of critical importance. (C) 2010 Elsevier Ltd All rights reserved
引用
收藏
页码:397 / 401
页数:5
相关论文
共 19 条
[1]   Range of motion change after cervical arthroplasty with ProDisc-C and Prestige artificial discs compared with anterior cervical discectomy and fusion [J].
Chang, Ung-Kyu ;
Kim, Daniel H. ;
Lee, Max C. ;
Willenberg, Rafer ;
Kim, Se-Hoon ;
Lim, Jesse .
JOURNAL OF NEUROSURGERY-SPINE, 2007, 7 (01) :40-46
[2]  
DiAngelo Denis J, 2004, Neurosurg Focus, V17, pE7
[3]  
DiAngelo Denis J, 2004, Neurosurg Focus, V17, pE4
[4]   Adjacent level intradiscal pressure and segmental kinematics following a cervical total disc arthroplasty -: An In Vitro human cadaveric model [J].
Dmitriev, AE ;
Cunningham, BW ;
Hu, NB ;
Sell, G ;
Vigna, F ;
McAfee, PC .
SPINE, 2005, 30 (10) :1165-1172
[5]   Biomechanics of the C5-C6 spinal unit before and after placement of a disc prosthesis [J].
Galbusera, F. ;
Fantigrossi, A. ;
Raimondi, M. T. ;
Sassi, M. ;
Fornari, M. ;
Assietti, R. .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2006, 5 (04) :253-261
[6]   Biomechanical studies on cervical total disc arthroplasty: A literature review [J].
Galbusera, Fabio ;
Bellini, Chiara M. ;
Brayda-Bruno, Marco ;
Fornari, Maurizio .
CLINICAL BIOMECHANICS, 2008, 23 (09) :1095-1104
[7]   Cervical spine biomechanics following implantation of a disc prosthesis [J].
Galbusera, Fabio ;
Bellini, Chiara M. ;
Raimondi, Manuela T. ;
Fornari, Maurizio ;
Assietti, Roberto .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (09) :1127-1133
[9]   Intervertebral disc replacement for cervical degenerative disease -: clinical results and functional outcome at two years in patients implanted with the Bryan® cervical disc prosthesis [J].
Heidecke, V. ;
Burkert, W. ;
Brucke, M. ;
Rainov, N. G. .
ACTA NEUROCHIRURGICA, 2008, 150 (05) :453-459
[10]   Biomechanics of nonfusion implants [J].
Huang, RC ;
Wright, TM ;
Panjabi, MM ;
Lipman, JD .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2005, 36 (03) :271-+