A polarization state detector (PSD) measures the state of polarization of the detected light. The state of polarization is fully described by the Stokes vector containing four Stokes parameters. A division-of-amplitude photopolarimeter (DOAP) measures the four Stokes parameters by simultaneously acquiring four intensities using photodetectors. A key component of the DOAP is the first beam splitter, which splits up the incoming beam into two beams. The effect of the beam splitter on the state of polarization of the reflected (r) and transmitted (t) beam is determined by six parameters: R; T;Psi(r);Psi(t);Psi(r), and Delta(t). R and T are the reflectance and transmittance, and (Psi(r);Delta(r)) and (Psi(t);Delta(t)) are the ellipsometric parameters of the beam splitter in reflection and transmission, respectively. To measure the Stokes vector with high accuracy, the six optical parameters must be chosen appropriately. In previous work, the optimal parameters of the beam splitter have been determined as R = T = 1/2; cos(2) 2 Psi(r) = 1/3;Psi(t) = Pi=2 and (Delta(r) modulo Pi = Pi/2 by calculating the maximum of the absolute value of the determinant of the instrument matrix. Using additional quarter-wave plates eliminates the constraint on the retardance and hence simplifies the manufacturing process of the beam splitter, especially when broadband application is intended. To compensate a suboptimal value of Delta(r) - Delta(t), the azimuthal angles of the principal axes of the retarders must be adjusted, for which we provide analytic formulas. Hence, a DOAP with retarders is also optimal in the sense that the same values for the determinant and condition number of the instrument matrix are obtained. When using two additional retarders, it is necessary to mount both at the same light path in order to obtain an optimal DOAP. We will show that is also possible to get an optimal DOAP with only one additional quarter-wave plate instead of two, if one of the Wollaston prisms is rotated.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Polarimeter system of stokes parameters based on Division-of-Amplitude
    Cheng, Min-Xi
    He, Zhen-Jiang
    Huang, Zuo-Hua
    Guangdian Gongcheng/Opto-Electronic Engineering, 2008, 35 (05): : 93 - 97
  • [2] Signal, noise, and bias for a broadband, division-of-amplitude Stokes polarimeter
    Tyler, David W.
    Mudge, Jason D.
    OPTICS EXPRESS, 2010, 18 (17): : 18015 - 18034
  • [3] Broadband division-of-amplitude polarimeter based on uncoated prisms
    Compain, E
    Drevillon, B
    APPLIED OPTICS, 1998, 37 (25): : 5938 - 5944
  • [4] Review of division-of-amplitude photopolarimeter
    Du, Xi-Liang
    Dai, Jing-Min
    Jiliang Xuebao/Acta Metrologica Sinica, 2006, 27 (04): : 325 - 330
  • [5] Handy method to calibrate division-of-amplitude polarimeters for the first three Stokes parameters
    Morel, Olivier
    Seulin, Ralph
    Fofi, David
    OPTICS EXPRESS, 2016, 24 (12): : 13634 - 13646
  • [6] Homemade open-source full-Stokes polarimeter based on division of amplitude
    Del Rio-lima, A.
    Gutierrez-valdes, A.
    Mojica-casique, C.
    Ballesteros-flores, F.
    Villanueva-reyes, I. A.
    Colin-rodriguez, R.
    Gardea-flores, C. A.
    Poveda-cuevas, F. J.
    Seman, J. A.
    APPLIED OPTICS, 2024, 63 (27) : 7177 - 7187
  • [7] Research of Polarization Navigation System Based on Division-of-Amplitude
    He Xin
    Li Leilei
    Chen Jiabin
    Zhang Zhongyu
    Dai Yansheng
    2016 IEEE CHINESE GUIDANCE, NAVIGATION AND CONTROL CONFERENCE (CGNCC), 2016, : 306 - 311
  • [8] Recent developments of division-of-amplitude photopolarimeters
    Azzam, RMA
    INTERNATIONAL SYMPOSIUM ON POLARIZATION ANALYSIS AND APPLICATIONS TO DEVICE TECHNOLOGY, 1996, 2873 : 1 - 4
  • [9] DIVISION-OF-AMPLITUDE PHOTOPOLARIMETER (DOAP) FOR THE SIMULTANEOUS MEASUREMENT OF ALL 4 STOKES PARAMETERS OF LIGHT
    AZZAM, RMA
    OPTICA ACTA, 1982, 29 (05): : 685 - 689
  • [10] Static spectroscopic ellipsometer based on division-of-amplitude polarization demodulation
    Li, Siyuan
    Deng, Zhongxun
    Quan, Naicheng
    Zhang, Chunmin
    OPTICS COMMUNICATIONS, 2024, 552