The New Lead (II) Ion Selective Electrode Based On Free Plasticizer Film of pTHFA Photopolymer

被引:11
作者
Alva, Sagir [1 ]
Widinugroho, Arsyad [1 ]
Adrian, Melny [1 ]
Khaerudini, Deni Shidqi [1 ,2 ]
Pratiwi, Swandya Eka [1 ]
Abd Aziz, Aiman Sajidah [3 ]
机构
[1] Univ Mercu Buana, Mech Engn Dept, West Jakarta 11650, Indonesia
[2] Indonesian Inst Sci LIPI Kawasan Puspiptek, Res Ctr Phys, Serpong 15314, South Tangerang, Indonesia
[3] MIMOS Semicond Sdn Bhd, Flexible Elect Lab, Kuala Lumpur 57000, Malaysia
关键词
POLYMERIC MEMBRANE ELECTRODES; COEFFICIENTS; ACRYLATE); EXPOSURE; SENSORS; SAMPLES; CD(II); BLOOD; METAL; PH;
D O I
10.1149/2.0601915jes
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
This research has succeeded in developing a Lead (II) -Ion-selective electrode (Pb2+-ISE) sensor based on poly-tetrahydro furfuryl acrylate (pTHFA) films free plasticizers utilizing photo-polymer technique. The Pb2+-ISE has a performance that includes a linear range of 0.1-10(-5) M, Nernstian number 26.5-29.8 mV/dec and LOD 3.24-3.98 (x 10(-6)) M, where the optimum composition between KTpClPB lipophilic salt with leads (IV) ionophore is 1:1 mole ratio. The Pb2+-ISE sensor also shows a relatively good selectivity value (Log K-a,b(pot)) against several interfering ions including K+ -6.5 +/- 0.3, Na+ -6.2 +/- 0.1, NH4+ -6.7 +/- 0.1, Mg2+ -12.7 +/- 0.2, Cu2+ -4.2 +/- 0.5 and Cd2+ -6.1 +/- 0.2, where the response of the Pb2+-ISE sensor is stable at pH 3-8. In this study, the Pb2+-ISE sensor shows validation test results that are comparable to measurements using atomic absorption spectroscopy (AAS) devices using artificial solutions and real sample. The results of testing with artificial solutions successively produced Pb2+ concentrations of 9.5 +/- 0.3, 52.5 +/- 1.7, and 100.8 +/- 2.5 ppm, while for real samples the Cisadane river was 0.0025 +/- 0.45 ppm and the refueling station was 18.15 +/- 16.74 mg/kg soil sample. (C) 2019 The Electrochemical Society.
引用
收藏
页码:B1513 / B1519
页数:7
相关论文
共 53 条
[1]  
Alva S., 2016, INT J INNOV MECH ENG, V2, P10, DOI [10.22441/ijimeam.2016.1.002, DOI 10.22441/IJIMEAM.2016.1.002]
[2]   Solid-State Reference Electrode Based on Thin-Films of Tetrahydrofurfuryl Acrylate (pTHFA) Photopolymer [J].
Alva, Sagir ;
Ardiyansyah, Dego ;
Khaerudini, Deni Shidqi ;
Suherman, Robi .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2019, 166 (08) :B598-B603
[3]   Ag/AgCl Reference Electrode Based on Thin Film of Arabic Gum Membrane [J].
Alva, Sagir ;
Aziz, Aiman Sajidah Binti Abdul ;
bin Syono, Mohd Ismahadi ;
Jamil, Wan Adil bin Wan .
INDONESIAN JOURNAL OF CHEMISTRY, 2018, 18 (03) :479-485
[4]   Development of Ammonium-Selective Electrode Based on PVC/MB28 Membrane [J].
Alva, Sagir ;
Sundari, R. ;
Aziz, A. S. A. ;
Rashid, N. A. A. ;
Gunawan, W. .
INTERNATIONAL CONFERENCE ON DESIGN, ENGINEERING AND COMPUTER SCIENCES, 2018, 453
[5]   Disposable solid-contact ion-selective electrodes for environmental monitoring of lead with ppb limit-of-detection [J].
Anastasova, Salzitsa ;
Radu, Aleksandar ;
Matzeu, Giusy ;
Zuliani, Claudio ;
Mattinen, Ulriika ;
Bobacka, Johan ;
Diamond, Dermot .
ELECTROCHIMICA ACTA, 2012, 73 :93-97
[6]  
Arida HA, 2011, INT J ELECTROCHEM SC, V6, P3858
[7]   Anthracene based surface acoustic wave sensors for picomolar detection of lead ions. Correlation between experimental results and DFT calculations [J].
Attia, G. ;
Rahali, S. ;
Teka, S. ;
Fourati, N. ;
Zerrouki, C. ;
Seydou, M. ;
Chehimi, S. ;
Hayouni, S. ;
Mbakidi, J. -P. ;
Bouquillon, S. ;
Majdoub, M. ;
Ben Chaabane, R. .
SENSORS AND ACTUATORS B-CHEMICAL, 2018, 276 :349-355
[8]   Enhancing ion-selective polymeric membrane electrodes by instrumental control [J].
Bakker, Eric .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2014, 53 :98-105
[9]   How to assess the limits of ion-selective electrodes:: Method for the determination of the ultimate span, response range, and selectivity coefficients of neutral carrier-based cation selective electrodes [J].
Bereczki, R ;
Takács, B ;
Langmaier, J ;
Neely, M ;
Gyurcsányi, RE ;
Tóth, K ;
Nagy, G ;
Lindner, E .
ANALYTICAL CHEMISTRY, 2006, 78 (03) :942-950
[10]  
Buhlmann P., 2012, SUPRAMOL CHEM, V5, P2539