A hybrid volumetric dose verification method for single-isocenter multiple-target cranial SRS

被引:14
作者
Ahmed, Saeed [1 ,2 ]
Kapatoes, Jeff [3 ]
Zhang, Geoffrey [1 ]
Moros, Eduardo G. [1 ]
Feygelman, Vladimir [1 ]
机构
[1] H Lee Moffitt Canc Ctr & Res Inst, Dept Radiat Oncol, Tampa, FL 33612 USA
[2] Univ S Florida, Dept Phys, Tampa, FL 33620 USA
[3] Sun Nucl Corp, Melbourne, FL USA
关键词
Patient-specific SRS QA; BRAIN METASTASES JLGK0901; MODULATED ARC THERAPY; STEREOTACTIC RADIOSURGERY; QUANTITATIVE-EVALUATION; DOSIMETRY; SYSTEM; SBRT; DISTRIBUTIONS; VALIDATION; REFINEMENT;
D O I
10.1002/acm2.12430
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A commercial semi-empirical volumetric dose verification system (PerFraction [PF], Sun Nuclear Corp.) extracts multi-leaf collimator positions from the electronic portal imaging device movies collected during a pre-treatment run, while the rest of the delivered control point information is harvested from the accelerator log files. This combination is used to reconstruct dose on a patient CT dataset with a fast superposition/convolution algorithm. The method was validated for single-isocenter multitarget SRS VMAT treatments against absolute radiochromic film measurements in a cylindrical phantom. The targets ranged in size from 0.8 to 3.6 cm and in number from 3 to 10 per plan. A total of 17 films rotated at different angles around the cylinder axis were analyzed. Each of 27 total targets was intercepted by at least one film, and 2-4 different films were analyzed per plan. Film dose was always scaled to the ion chamber measurement in a high-dose, low-gradient area deliberately created at the isocenter. The planar dose agreement between PF and film using 3%(Global dose-difference normalization)/1 mm gamma analysis was on average 99.2 +/- 1.1%. The point dose difference in the low-gradient area in the middle of every target was below 3%, while PF-reconstructed and film dose centroids for individual targets showed submillimeter agreement when measured on a well aligned accelerator. Volumetrically, all voxels in all plans agreed between PF and the primary treatment planning system at the 3%/1 mm level. With proper understanding of its advantages and shortcomings, the tool can be applied to patient-specific QA in routine radiosurgical clinical practice.
引用
收藏
页码:651 / 658
页数:8
相关论文
共 39 条
[1]   Validation of a GPU-Based 3D dose calculator for modulated beams [J].
Ahmed, Saeed ;
Hunt, Dylan ;
Kapatoes, Jeff ;
Hayward, Robert ;
Zhang, Geoffrey ;
Moros, Eduardo G. ;
Feygelman, Vladimir .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (03) :73-82
[2]  
[Anonymous], 2013, TRUEBEAM DEV MOD US
[3]   Evaluation of volumetric modulated arc therapy for cranial radiosurgery using multiple noncoplanar arcs [J].
Audet, Chantal ;
Poffenbarger, Brett A. ;
Chang, Pauling ;
Jackson, Paul S. ;
Lundahl, Robert E. ;
Ryu, Stephen I. ;
Ray, Gordon R. .
MEDICAL PHYSICS, 2011, 38 (11) :5863-5872
[4]   Polymer gel dosimetry [J].
Baldock, C. ;
De Deene, Y. ;
Doran, S. ;
Ibbott, G. ;
Jirasek, A. ;
Lepage, M. ;
McAuley, K. B. ;
Oldham, M. ;
Schreiner, L. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :R1-R63
[5]   Long-term evaluation and cross-checking of two geometric calibrations of kV and MV imaging systems for Linacs [J].
Chiu, Tsuicheng D. ;
Yan, Yulong ;
Foster, Ryan ;
Mao, Weihua .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (04) :306-310
[6]   FEASIBILITY OF SINGLE-ISOCENTER VOLUMETRIC MODULATED ARC RADIOSURGERY FOR TREATMENT OF MULTIPLE BRAIN METASTASES [J].
Clark, Grant M. ;
Popple, Richard A. ;
Young, P. Edward ;
Fiveash, John B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (01) :296-302
[7]   A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation [J].
Depuydt, T ;
Van Esch, A ;
Huyskens, DP .
RADIOTHERAPY AND ONCOLOGY, 2002, 62 (03) :309-319
[8]   The history and principles of chemical dosimetry for 3-D radiation fields: Gels, polymers and plastics [J].
Doran, Simon J. .
APPLIED RADIATION AND ISOTOPES, 2009, 67 (03) :393-398
[9]  
Ezzell G A, 2001, J Appl Clin Med Phys, V2, P138, DOI 10.1120/1.1386508
[10]   Motion as a perturbation: Measurement-guided dose estimates to moving patient voxels during modulated arc deliveries [J].
Feygelman, Vladimir ;
Stambaugh, Cassandra ;
Zhang, Geoffrey ;
Hunt, Dylan ;
Opp, Daniel ;
Wolf, Theresa K. ;
Nelms, Benjamin E. .
MEDICAL PHYSICS, 2013, 40 (02)