Incorporating Cone-Beam CT Into the Diagnostic Algorithm for Suspected Radiocarpal Fractures: A New Standard of Care?

被引:28
作者
Gibney, Brian [1 ]
Smith, Michelle [2 ]
Moughty, Adrian [3 ]
Kavanagh, Eoin C. [1 ,4 ]
Hynes, Darragh [2 ,4 ]
MacMahon, Peter J. [1 ,4 ]
机构
[1] Mater Misericordiae Univ Hosp, Dept Radiol, Whitty Bldg,N Circular Rd, Dublin D07 R2WY, Ireland
[2] Mater Misericordiae Univ Hosp, Dept Orthopaed Surg, Dublin, Ireland
[3] Mater Misericordiae Univ Hosp, Emergency Dept, Dublin, Ireland
[4] Univ Coll Dublin, Sch Med, Dublin, Ireland
关键词
cone-beam CT; diagnosis; trauma; wrist; OCCULT SCAPHOID FRACTURES; CONVENTIONAL RADIOGRAPHY; COMPUTED-TOMOGRAPHY; COST-EFFECTIVENESS; WRIST FRACTURES; SENSITIVITY; SYSTEM; MDCT; MRI;
D O I
10.2214/AJR.19.21478
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to assess the result of adding cone-beam CT to the standard imaging algorithm for patients with suspected radiographically occult traumatic radiocarpal fractures. SUBJECTS AND METHODS. A prospective review was performed on all patients who had cone-beam CT investigation of acute wrist pain after normal initial radiographs. Patients with no identified fractures were clinically reassessed and referred for MRI if concern for a fracture persisted. RESULTS. In all, 117 patients were assessed; 50.4% had fractures identified with a total of 67 radiographically occult fractures. One fracture was identified on MRI that was not seen on cone-beam CT. Cone-beam CT had sensitivity of 98.3% (95% CI, 91.1-100%), specificity of 100% (95% CI, 93.7-100%), positive predictive value of 100%, and negative predictive value of 98.3% (95% CI, 89.1-100%). Accuracy was 99.1% (95% CI, 95.3-100%). CONCLUSION. Incorporating cone-beam CT into routine clinical practice as part of a standardized diagnostic algorithm yielded a 50% fracture detection rate in patients with negative wrist radiographs but ongoing clinical concern for radiocarpal fracture. Cone-beam CT provides more diagnostic information than radiographs at a lower radiation dose than conventional MDCT. Given the poor accuracy of radiographs for acute radiocarpal fractures and the high fracture prevalence in this cohort, we feel that cone-beam CT should be regarded as the new standard of care in the investigation of these patients.
引用
收藏
页码:1117 / 1123
页数:7
相关论文
共 28 条
[1]   Wrist fractures: sensitivity of radiography, prevalence, and patterns in MDCT [J].
Balci A. ;
Basara I. ;
Çekdemir E.Y. ;
Tetik F. ;
Aktaş G. ;
Acarer A. ;
Özaksoy D. .
Emergency Radiology, 2015, 22 (3) :251-256
[2]   Comparison of conventional radiography and MDCT in suspected scaphoid fractures [J].
Behzadi, Cyrus ;
Karul, Murat ;
Henes, Frank Oliver ;
Laqmani, Azien ;
Catala-Lehnen, Philipp ;
Lehmann, Wolfgang ;
Nagel, Hans-Dieter ;
Adam, Gerhard ;
Regier, Marc .
WORLD JOURNAL OF RADIOLOGY, 2015, 7 (01) :22-27
[3]   Diagnostic value of cone beam computed tomography (CBCT) in occult scaphoid and wrist fractures [J].
Borel, Christophe ;
Larbi, Ahmed ;
Delclaux, Stephanie ;
Lapegue, Franck ;
Chiavassa-Gandois, Helene ;
Sans, Nicolas ;
Faruch-Bilfeld, Marie .
EUROPEAN JOURNAL OF RADIOLOGY, 2017, 97 :59-64
[4]   The suspected scaphoid injury: resource implications in the absence of magnetic resonance imaging [J].
Burns, M. J. ;
Aitken, S. A. ;
McRae, D. ;
Duckworth, A. D. ;
Gray, A. .
SCOTTISH MEDICAL JOURNAL, 2013, 58 (03) :143-148
[5]   Direct comparison of conventional radiography and cone-beam CT in small bone and joint trauma [J].
De Smet, E. ;
De Praeter, G. ;
Verstraete, K. L. A. ;
Wouters, K. ;
De Beuckeleer, Luc ;
Vanhoenacker, F. M. H. M. .
SKELETAL RADIOLOGY, 2015, 44 (08) :1111-1117
[6]   MRI as a reference standard for suspected scaphoid fractures [J].
De Zwart, A. D. ;
Beeres, F. J. P. ;
Ring, D. ;
Kingma, L. M. ;
Coerkamp, E. G. ;
Meylaerts, S. A. G. ;
Rhemrev, S. J. .
BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1016) :1098-1101
[7]   Cone-Beam CT in diagnosis of scaphoid fractures [J].
Edlund, Rolf ;
Skorpil, Mikael ;
Lapidus, Gunilla ;
Backlund, Jenny .
SKELETAL RADIOLOGY, 2016, 45 (02) :197-204
[8]   Prototype metal artefact reduction algorithm in flat panel computed tomography - evaluation in patients undergoing transarterial hepatic radioembolisation [J].
Hamie, Qeumars Mustafa ;
Kobe, Adrian Raoul ;
Mietzsch, Leif ;
Manhart, Michael ;
Puippe, Gilbert Dominique ;
Pfammatter, Thomas ;
Guggenberger, Roman .
EUROPEAN RADIOLOGY, 2018, 28 (01) :265-273
[9]   Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects [J].
Huang, Ambrose J. ;
Chang, Connie Y. ;
Thomas, Bijoy J. ;
MacMahon, Peter J. ;
Palmer, William E. .
SKELETAL RADIOLOGY, 2015, 44 (06) :797-809
[10]   Diagnosis of Occult Scaphoid Fractures A Cost-Effectiveness Analysis [J].
Karl, John W. ;
Swart, Eric ;
Strauch, Robert J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2015, 97A (22) :1860-1868