Tonometers and infectious risk: myth or reality? Efficacy of different disinfection regimens on tonometer tips

被引:23
作者
Cillino, S.
Casuccio, A.
Giammanco, G. M.
Mammina, C.
Morreale, D.
Di Pace, F.
Lodato, G.
机构
[1] Univ Palermo, Dipartimento Neurol Oftalmol Otorinolaringoiatria, Sez Oftalmol, I-90127 Palermo, Italy
[2] Univ Palermo, Dipartimento Igiene & Microbiol, Palermo, Italy
关键词
disposable tonometer prisms; ophthalmic device disinfection; cross-infection ophthalmic transmission;
D O I
10.1038/sj.eye.6702269
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To evaluate the adequacy of common disinfection regimens for disposable tonometer tips and assess if disinfection of reusable prisms or the use of disposable tips is preferable. Methods We used disposable tonometer tips, using the same material and tip diameter of standard Goldmann tonometer prism. Strains of Pseudomonas aeruginosa, Staphylococcus aureus, Bacillus subtilis and Candida albicans were tested according to the European standard guidelines for disinfectants test. Antimicrobial effectiveness of the following disinfection practices has been assessed: dry wipe, Minuten wipes (Alpro (R)), soaking in 3% hydrogen peroxide, 0.5% benzalkonium chloride, and 0.5% Pantasept (R) for 1, 5, and 15 min. All tests have been performed three times and all conditions tested in duplicate. Results Dry wiping and 1 min soak in 3% hydrogen peroxide were ineffective on all microrganisms. Minuten wipes, 1 min soak in 0.5% benzalkonium chloride or 3% hydrogen peroxide were ineffective on B. subtilis. 0.5% Pantasept (R) soak was effective in 1 min for all microrganisms tested, whereas 3% hydrogen peroxide and 0.5% benzalkonium chloride soaks were effective when performed for at least 5 min. B. subtilis was the most resistant organism to disinfectant regimes at 1 min time. Conclusions Results of our study demonstrate a relative disinfection efficacy for the different evaluated regimens, provided that correct exposure times are adopted for the chosen disinfectants, a condition difficult to ensure in a busy clinic setting. We conclude that disposable prism tonometry provides a safe alternative to Goldmann tonometry.
引用
收藏
页码:541 / 546
页数:6
相关论文
共 53 条
[31]  
231::AID-JMV15&gt
[32]  
3.0.CO
[33]  
2-N
[34]  
*MICR IND INC, 1996, SURVEY
[35]  
MURRAH WF, 1988, ANN OPHTHALMOL, V20, P36
[36]  
Parment PA, 1996, ACTA OPHTHALMOL SCAN, V74, P235
[37]   INVESTIGATIONS OF PATIENTS OF HEALTH-CARE WORKERS INFECTED WITH HIV - THE CENTERS-FOR-DISEASE-CONTROL AND PREVENTION DATABASE [J].
ROBERT, LM ;
CHAMBERLAND, ME ;
CLEVELAND, JL ;
MARCUS, R ;
GOOCH, BF ;
SRIVASTAVA, PU ;
CULVER, DH ;
JAFFE, HW ;
MARIANOS, DW ;
PANLILIO, AL ;
BELL, DM .
ANNALS OF INTERNAL MEDICINE, 1995, 122 (09) :653-657
[38]  
Rosenthal Ruth A., 1995, CLAO Journal, V21, P99
[39]  
SALVI SM, 2004, EYE, P1
[40]   The epidemic of Acanthamoeba keratitis: Where do we stand? [J].
Schaumberg, DA ;
Snow, KK ;
Dana, MR .
CORNEA, 1998, 17 (01) :3-10