Comparison of the performance of battery-operated fluid warmers

被引:14
作者
Lehavi, Amit [1 ,2 ]
Yitzhak, Avraham [3 ]
Jarassy, Refael [3 ]
Heizler, Rami [1 ]
Katz, Yeshayahu [1 ,2 ]
Raz, Aeyal [1 ,4 ]
机构
[1] Dept Anesthesiol, Rambam Healthcare Campus, IL-3109601 Haifa, Israel
[2] Technion Israel Inst Technol, Ruth & Bruce Rappaport Fac Med, Haifa, Israel
[3] Israel Def Forces, Med Corps, Surg Gen Headquarters, Tel Aviv, Israel
[4] Univ Wisconsin, Dept Anesthesiol, Madison, WI USA
关键词
BUDDY LITE(TM); DEVICES; FLOW;
D O I
10.1136/emermed-2017-207112
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives Warming intravenous fluids is essential to prevent hypothermia in patients with trauma, especially when large volumes are administered. Prehospital and transport settings require fluid warmers to be small, energy efficient and independent of external power supply. We compared the warming properties and resistance to flow of currently available battery-operated fluid warmers. Methods Fluid warming was evaluated at 50, 100 and 200 mL/min at a constant input temperature of 20 degrees C and 10 degrees C using a cardiopulmonary bypass roller pump and cooler. Output temperature was continuously recorded. Results Performance of fluid warmers varied with flows and input temperatures. At an input temperature of 20 degrees C and flow of 50 ml/min, the Buddy Lite, enFlow, Thermal Angel and Warrior warmed 3.4, 2.4, 1 and 3.6 L to over 35 degrees C, respectively. However, at an input temperature of 10 degrees C and flow of 200 ml/min, the Buddy Lite failed to warm, the enFlow warmed 3.3 L to 25.7 degrees C, the Thermal Angel warmed 1.5 L to 20.9 degrees C and the Warrior warmed 3.4 L to 34.4 degrees C (p<0.0001). Conclusion We found significant differences between the fluid warmers: the use of the Buddy Lite should be limited to moderate input temperature and low flow rates. The use of the Thermal Angel is limited to low volumes due to battery capacity and low output temperature at extreme conditions. The Warrior provides the best warming performance at high infusion rates, as well as low input temperatures, and was able to warm the largest volumes in these conditions.
引用
收藏
页码:564 / 570
页数:7
相关论文
共 8 条
[1]   A laboratory comparison of the performance of the buddy lite™ and enFlow™ fluid warmers [J].
Bruells, C. S. ;
Bruells, A. C. ;
Rossaint, R. ;
Stoppe, C. ;
Schaelte, G. ;
Zoremba, N. .
ANAESTHESIA, 2013, 68 (11) :1161-1164
[2]   Warming of intravenous and irrigation fluids for preventing inadvertent perioperative hypothermia [J].
Campbell, Gillian ;
Alderson, Phil ;
Smith, Andrew F. ;
Warttig, Sheryl .
COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2015, (04)
[3]   Evaluation of commercially available fluid-warming devices for use in forward surgical and combat areas [J].
Dubick, MA ;
Brooks, DE ;
Macaitis, JM ;
Bice, TG ;
Moreau, AR ;
Holcomb, JB .
MILITARY MEDICINE, 2005, 170 (01) :76-82
[4]   Flow rates through intravenous access devices: an in vitro study [J].
Khoyratty, Saleem I. ;
Gajendragadkar, Pushpaj R. ;
Polisetty, Kiran ;
Ward, Sue ;
Skinner, Tim ;
Gajendragadkar, Parag R. .
JOURNAL OF CLINICAL ANESTHESIA, 2016, 31 :101-105
[5]   The laboratory performance of the enFLOW®, buddy lite™ and ThermoSens® fluid warmers [J].
Kim, H. J. ;
Yoo, S. M. ;
Jung, J. S. ;
Lee, S. H. ;
Sun, K. ;
Son, H. S. .
ANAESTHESIA, 2015, 70 (02) :205-208
[6]   Needleless Connectors Substantially Reduce Flow of Crystalloid and Red Blood Cells During Rapid Infusion [J].
Lehn, Robert A. ;
Gross, Jeffrey B. ;
McIsaac, Joseph H. ;
Gipson, Keith E. .
ANESTHESIA AND ANALGESIA, 2015, 120 (04) :801-804
[7]  
Ndao S, 2013, J SPEC OPER MED, V13, P31
[8]   Hypothermia in the trauma patient [J].
Tsuei, BJ ;
Kearney, PA .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2004, 35 (01) :7-15