Online electrochemical method for continuous and simultaneous monitoring of glucose and L-lactate in vivo with graphene hybrids as the electrocatalyst

被引:9
|
作者
Gu, Hui [1 ]
Yang, Yan [1 ]
Zhou, Xinguang [1 ]
Zhou, Tianshu [2 ]
Shi, Guoyue [1 ]
机构
[1] E China Normal Univ, Dept Chem, Shanghai 200241, Peoples R China
[2] E China Normal Univ, Dept Environm Sci, Shanghai 200241, Peoples R China
基金
中国国家自然科学基金; 上海市自然科学基金;
关键词
Pd nanoparticles; Graphene; In vivo; Striatum; Glucose; L-lactate; ENZYME ELECTRODE; STRIATUM GLUCOSE; QUANTUM DOTS; AU; MICRODIALYSIS; BIOSENSOR; NANOPARTICLES; METABOLISM; GLUTAMATE; OXIDASE;
D O I
10.1016/j.jelechem.2014.06.002
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study described a physiologically relevant on-line electrochemical method for simultaneous monitoring of glucose and L-lactate in vivo with good sensitivity and selectivity. The analytical system that constructed by dual-enzyme biosensor was based on a novel and effective graphene hybrids with the on-line microdialysis system, which made the continuous and simultaneous monitoring of striatum glucose and L-lactate in brain of freely moving rats possible. The dual-enzyme biosensor was accomplished by using oxidases as the specific sensing unit and the UV-assisted synthesized Graphene/ZnO@Pd as the electrocatalyst. Transmission electron microscopy, X-ray diffraction, energy-dispersive X-ray and X-ray photoelectron spectroscopy confirmed the denoted Pd nanoparticles (NPs) on the Graphene/ZnO composites. The analytical system showed good linearity for glucose and L-lactate concentrations from 20 mu M to 500 mu M with detection limit of 2.39 mu M and 2.52 mu M respectively. After post-calibration, the basal level concentrations of glucose and L-lactate in the microdialysate from the striatum of freely moving rats were calculated to be 0.323 +/- 0.021 mM and 0.501 mM +/- 0.057 mM (mean +/- s.d., n = 3). Moreover, the dual oxidase/Graphene/ZnO@Pd based biosensors suffered from little cross-talk and exhibit a good stability and reproducibility. (C) 2014 Published by Elsevier B.V.
引用
收藏
页码:41 / 47
页数:7
相关论文
共 41 条
  • [1] In vivo continuous monitoring of L-lactate coupling subcutaneous microdialysis and an electrochemical biocell
    Volpe, G.
    Moscone, D.
    Compagnone, D.
    Palleschi, G.
    Sensors and Actuators, B: Chemical, 1995, B24 (1 -3 pt 1): : 138 - 141
  • [2] IN-VIVO CONTINUOUS MONITORING OF L-LACTATE COUPLING SUBCUTANEOUS MICRODIALYSIS AND AN ELECTROCHEMICAL BIOCELL
    VOLPE, G
    MOSCONE, D
    COMPAGNONE, D
    PALLESCHI, G
    SENSORS AND ACTUATORS B-CHEMICAL, 1995, 24 (1-3) : 138 - 141
  • [3] Continuous monitoring of D-glucose and L-lactate by flow injection analysis
    Haouz, A
    Stieg, S
    ENZYME AND MICROBIAL TECHNOLOGY, 2002, 30 (01) : 129 - 133
  • [4] Physiologically Relevant Online Electrochemical Method for Continuous and Simultaneous Monitoring of Striatum Glucose and Lactate Following Global Cerebral Ischemia/Reperfusion
    Lin, Yuqing
    Zhu, Ningning
    Yu, Ping
    Su, Lei
    Mao, Lanqun
    ANALYTICAL CHEMISTRY, 2009, 81 (06) : 2067 - 2074
  • [5] DETERMINATION OF L-LACTATE BY AN ENZYMATIC-ELECTROCHEMICAL METHOD
    KRAGENINGS, I
    RACKWITZ, R
    ANNALES DE BIOLOGIE CLINIQUE, 1978, 36 (03) : 290 - 291
  • [6] Simultaneous Determination of Glucose and l-Lactate in Rat Brain by an Electrochemical in vivo Flow-Injection System with an On-line Microdialysis Sampling
    Toshio Yao
    Takayuki Yano
    Youko Nanjyo
    Hirohito Nishino
    Analytical Sciences, 2003, 19 : 61 - 65
  • [7] Simultaneous determination of glucose and L-lactate in rat brain by an electrochemical in vivo flow-injection system with an on-line microdialysis sampling
    Yao, T
    Yano, T
    Nanjyo, Y
    Nishino, H
    ANALYTICAL SCIENCES, 2003, 19 (01) : 61 - 65
  • [8] Simultaneous in vivo monitoring of glucose, L-lactate, and pyruvate concentrations in rat brain by a flow-injection biosensor system with an on-line microdialysis sampling
    Yao, T
    Yano, T
    Nishino, H
    ANALYTICA CHIMICA ACTA, 2004, 510 (01) : 53 - 59
  • [9] Blood withdrawal system for continuous ex vivo glucose and lactate monitoring
    Gfrerer, RJ
    Trajanoski, Z
    Skrabal, F
    Jobst, G
    Urban, G
    Kotanko, P
    Wach, P
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 75 - 76
  • [10] Multianalyte sensor for the simultaneous determination of glucose, L-lactate and uric acid based on a microelectrode array
    Frebel, H
    Chemnitius, GC
    Cammann, K
    Kakerow, R
    Rospert, M
    Mokwa, W
    SENSORS AND ACTUATORS B-CHEMICAL, 1997, 43 (1-3) : 87 - 93