Elimination of False-Positive Results in a Luminex Assay for Pneumococcal Antibodies

被引:26
作者
Pickering, Jerry W. [1 ]
Larson, Matthew T. [1 ]
Martins, Thomas B. [1 ]
Copple, Susan S. [2 ]
Hill, Harry R. [1 ,3 ]
机构
[1] ARUP Inst Clin & Expt Pathol, Salt Lake City, UT 84108 USA
[2] ARUP Labs, Salt Lake City, UT USA
[3] Univ Utah, Sch Med, Dept Pathol Pediat & Med, Salt Lake City, UT USA
关键词
FLUORESCENT MICROSPHERE IMMUNOASSAY; LINKED-IMMUNOSORBENT-ASSAY; SIMULTANEOUS QUANTITATION; CAPSULAR POLYSACCHARIDES; NEUTRALIZING EPITOPES; MULTIPLEX IMMUNOASSAY; BORDETELLA-PERTUSSIS; SERUM-ALBUMIN; VALIDATION; DIPHTHERIA;
D O I
10.1128/CVI.00329-09
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
In our 14-valent Luminex assay for pneumococcal antibodies, we identified two groups of sera that caused false-positive results. The first group bound nonspecifically to the Luminex microspheres. The second group reacted specifically with bovine serum albumin (BSA). We describe here methods that eliminated the false-positive reactivity of both groups.
引用
收藏
页码:185 / 189
页数:5
相关论文
共 19 条
[1]   Method for simultaneous measurement of antibodies to 23 pneumococcal capsular polysaccharides [J].
Biagini, RE ;
Schlottmann, SA ;
Sammons, DL ;
Smith, JP ;
Snawder, JC ;
Striley, CAF ;
MacKenzie, BA ;
Weissman, DN .
CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2003, 10 (05) :744-750
[2]   Improved multiplex immunoassay performance in human plasma and synovial fluid following removal of interfering heterophilic antibodies [J].
de Jager, W ;
Prakken, BJ ;
Bijlsma, JWJ ;
Kuis, W ;
Rijkers, GT .
JOURNAL OF IMMUNOLOGICAL METHODS, 2005, 300 (1-2) :124-135
[3]   Simultaneous Detection of Haemophilus influenzae Type b Polysaccharide-Specific Antibodies and Neisseria meningitidis Serogroup A, C, Y, and W-135 Polysaccharide-Specific Antibodies in a Fluorescent-Bead-Based Multiplex Immunoassay [J].
de Voer, Richarda M. ;
Schepp, Rutger M. ;
Versteegh, Florens G. A. ;
van der Klis, Fiona R. M. ;
Berbers, Guy A. M. .
CLINICAL AND VACCINE IMMUNOLOGY, 2009, 16 (03) :433-436
[4]   Optimization and validation of a multiplexed Luminex assay to quantify antibodies to neutralizing epitopes on human papillomaviruses 6, 11, 16, and 18 [J].
Dias, D ;
Van Doren, J ;
Schlottmann, S ;
Kelly, S ;
Puchalski, D ;
Ruiz, W ;
Boerckel, P ;
Kessler, J ;
Antonello, JM ;
Green, T ;
Brown, M ;
Smith, J ;
Chirmule, N ;
Barr, E ;
Jansen, KU ;
Esser, MT .
CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2005, 12 (08) :959-969
[5]  
Kricka LJ, 1999, CLIN CHEM, V45, P942
[6]   Development and validation of a nonaplex assay for the simultaneous quantitation of antibodies to nine Streptococcus pneumoniae serotypes [J].
Lal, G ;
Balmer, P ;
Stanford, E ;
Martin, S ;
Warrington, R ;
Borrow, R .
JOURNAL OF IMMUNOLOGICAL METHODS, 2005, 296 (1-2) :135-147
[7]   Heterophile antibody interference in a multiplexed fluorescent microsphere immunoassay for quantitation of cytokines in human serum [J].
Martins, TB ;
Pasi, BM ;
Litwin, CM ;
Hill, HR .
CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2004, 11 (02) :325-329
[8]   Development of a multiplexed fluorescent immunoassay for the quantitation of antibody responses to group A streptococci [J].
Martins, Thomas B. ;
Augustine, Nancy H. ;
Hill, Hany R. .
JOURNAL OF IMMUNOLOGICAL METHODS, 2006, 316 (1-2) :97-106
[9]   Sensitization following LVAD implantation using leucodepleted blood is not due to HLA antibodies [J].
Newell, H. ;
Smith, J. D. ;
Rogers, P. ;
Birks, E. ;
Danskine, A. J. ;
Fawson, R. E. ;
Rose, M. L. .
AMERICAN JOURNAL OF TRANSPLANTATION, 2006, 6 (07) :1712-1717
[10]   Simultaneous quantitation of antibodies to neutralizing epitopes on virus-like particles for human papillomavirus types 6, 11, 16, and 18 by a multiplexed luminex assay 2 [J].
Opalka, D ;
Lachman, CE ;
MacMullen, SA ;
Jansen, KU ;
Smith, JF ;
Chirmule, N ;
Esser, MT .
CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2003, 10 (01) :108-115