Assessment of carbonate-phosphoric acid analytical technique performed using GasBench II in continuous flow isotope ratio mass spectrometry

被引:79
|
作者
Paul, Debajyoti [1 ]
Skrzypek, Grzegorz [1 ]
机构
[1] Univ Texas, Dept Earth & Environm Sci, San Antonio, TX 78249 USA
关键词
stable isotope; carbonate; GasBench; continuous flow; phosphoric acid;
D O I
10.1016/j.ijms.2006.11.006
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The commonly used online automated analytical method (in particular Thermo GasBench 11) to determine stable carbon and oxygen isotope ratio composition of calcium carbonate in continuous flow isotope ratio mass spectrometry is evaluated. We have analyzed the isotopic composition of three in-house calcium carbonate standards as unknown samples for assessment of speed, accuracy and precision of isotopic measurements. Isotopic analyses Of CO2 extracted from carbonate-phosphoric acid reaction were performed, where temperature of reaction and the time allowed for the reaction was varied in the range 25.-90 degrees C and 3-101 min (and 12 h), respectively. Our results suggest that for isotopic analyses, carbonate-acid reaction may be conducted at any temperature in the range 50-90 degrees C for an hour; 50-70 degrees C seems to be a more restricted range. The results are reproduced accurately if the raw isotopic values are normalized against international standards also analyzed under the same experimental conditions as that of the samples. For isotopically homogeneous calcium carbonate samples, isotopic analyses may also be performed after 15 min of carbonate-acid reaction at 70 degrees C, which is significantly less time than what is routinely used. Care must be taken to ensure complete dissolution of carbonate sample as incomplete reaction or partial dissolution of isotopically heterogeneous grains would result in respective isotopic variability Of CO2 gas produced during carbonate-acid reaction. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:180 / 186
页数:7
相关论文
共 10 条
  • [1] Long-term performance of the Gasbench isotope ratio mass spectrometry system for the stable isotope analysis of carbonate microsamples
    Spoetl, Christoph
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2011, 25 (11) : 1683 - 1685
  • [2] An online method to determine chlorine stable isotope composition by continuous flow isotope ratio mass spectrometry (CF-IRMS) coupled with a Gasbench Ⅱ
    刘运德
    周爱国
    甘义群
    刘存富
    余婷婷
    李小倩
    JournalofCentralSouthUniversity, 2013, 20 (01) : 193 - 198
  • [3] Guidelines for Accurate and Precise Stable Isotope Analysis of Calcite, Dolomite, and Magnesite Using a Carbonate Device for Continuous Flow-Isotope Ratio Mass Spectrometry (CF-IRMS)
    Kim, Soujung
    Kim, Sang-Tae
    Knyf, Martin
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2025, 39 (05)
  • [4] Isotopic (δ13C, δ18O) Analysis of Small Amounts of Carbonate in Silicate Rocks by the Continuous Flow Isotope Ratio Mass Spectrometry Method
    Dubinina, E. O.
    Chizhova, Y. N.
    Kossova, S. A.
    GEOCHEMISTRY INTERNATIONAL, 2023, 61 (11) : 1164 - 1176
  • [5] Nitrogen Mass Fraction and Isotope Determinations in Geological Reference Materials Using Sealed-Tube Combustion Coupled with Continuous-Flow Isotope-Ratio Mass Spectrometry
    Feng, Lianjun
    Li, Hongwei
    Liu, Wei
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2018, 42 (04) : 539 - 548
  • [6] Sample matrix effects on measured carbon and oxygen isotope ratios during continuous-flow isotope-ratio mass spectrometry
    Levitt, Nicholas Paul
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2014, 28 (21) : 2259 - 2274
  • [7] Carbonate clumped isotope analysis (Δ47) of 21 carbonate standards determined via gas-source isotope-ratio mass spectrometry on four instrumental configurations using carbonate-based standardization and multiyear data sets
    Upadhyay, Deepshikha
    Lucarelli, Jamie
    Arnold, Alexandrea
    Flores, Randy
    Bricker, Hayley
    Ulrich, Robert N.
    Jesmok, Gregory
    Santi, Lauren
    Defliese, William
    Eagle, Robert A.
    Carroll, Hannah M.
    Bateman, Jesse Bloom
    Petryshyn, Victoria
    Loyd, Sean J.
    Tang, Jianwu
    Priyadarshi, Antra
    Elliott, Ben
    Tripati, Aradhna
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2021, 35 (17)
  • [8] Determination of clumped isotopes in carbonate using isotope ratio mass spectrometry: Toward a systematic evaluation of a sample extraction method using a static Porapak™ Q absorbent trap
    Wang, Xu
    Cui, Linlin
    Li, Yangyang
    Huang, Xiaofang
    Zhai, Jixuan
    Ding, Zhongli
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2016, 403 : 8 - 14
  • [9] Investigation of amino acid δ13C signatures in bone collagen to reconstruct human palaeodiets using liquid chromatography-isotope ratio mass spectrometry
    Choy, Kyungcheol
    Smith, Colin I.
    Fuller, Benjamin T.
    Richards, Michael P.
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2010, 74 (21) : 6093 - 6111
  • [10] D and 18O enrichment measurements in biological fluids in a continuous-flow elemental analyser with an isotope-ratio mass spectrometer using two configurations
    Ripoche, N.
    Ferchaud-Roucher, V.
    Krempf, M.
    Ritz, P.
    JOURNAL OF MASS SPECTROMETRY, 2006, 41 (09): : 1212 - 1218