Calibration of CT Hounsfield units for radiotherapy treatment planning of patients with metallic hip prostheses: the use of the extended CT-scale

被引:98
作者
Coolens, C [1 ]
Childs, PJ
机构
[1] Canc Res Inst, Dept Phys, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden Hosp NHS Trust, Dept Phys, Sutton SM2 5PT, Surrey, England
关键词
D O I
10.1088/0031-9155/48/11/308
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Heterogeneity corrections for radiotherapy dose calculations are based on the electron density of the disturbing heterogeneity. However, when CT planning a radiotherapy treatment, where metallic hip implants are present, considerable artefacts are seen in the images. Often, an additional problem arises whereby no information regarding the artificial hip's composition and geometry is available. This study investigates whether the extended CT range can be used to determine the composition (hence electron density) of artificial hips in radiotherapy patients. Two CT-calibration methods were evaluated, one based on material substitution, the other a stoichiometric calibration. We also evaluate whether the physical dimensions of metal prostheses can be accurately imaged for subsequent use in treatment planning computers. Neither calibration method successfully predicted electron densities. However, the limited range of implant-materials used in patients means that the extended CT range can still successfully distinguish between implant densities. The physical dimensions can be determined to 2 mm by establishing the required windowing of displays for each material. The cross-sectional area of the prosthesis and the presence of other high-density objects in a CT slice can influence the generated CT number and careful design of calibration phantoms is essential.
引用
收藏
页码:1591 / 1603
页数:13
相关论文
共 29 条
[1]  
Alecu R, 1999, Med Dosim, V24, P33, DOI 10.1016/S0958-3947(98)00044-2
[2]   EFFECT OF A FEMORAL-HEAD PROSTHESIS ON MEGAVOLTAGE BEAM RADIOTHERAPY [J].
BIGGS, PJ ;
RUSSELL, MD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (03) :581-586
[3]  
*BS EN, 1998, 12010 BS EN
[4]   INVIVO ISOCENTER DOSE IN 2 HIP-PROSTHESIS PATIENTS [J].
BURLESON, WD ;
STUTZMAN, CD ;
STITT, JA ;
KARLSSON, UL ;
MIAN, TA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (06) :1347-1352
[5]  
Carolan M, 2000, Australas Radiol, V44, P290, DOI 10.1046/j.1440-1673.2000.00816.x
[6]   A study on beams passing through hip prosthesis for pelvic radiation treatment [J].
Ding, GX ;
Yu, CW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :1167-1175
[7]  
ENG TY, 2001, MED DOSIM, V25, P7
[8]   PLANNING OF RADIOTHERAPY FOR PATIENTS WITH HIP-PROSTHESIS [J].
ERLANSON, M ;
FRANZEN, L ;
HENRIKSSON, R ;
LITTBRAND, B ;
LOFROTH, PO .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (05) :1093-1098
[9]  
FISHMAN E, 1986, RADIOLOGY, V160, P375
[10]  
GOLDMAN LW, 1995, P 1995 SUMM SCH CT U