Biomechanical evaluation of anterior cervical spine stabilization

被引:79
作者
Grubb, MR [1 ]
Currier, BL [1 ]
Shih, JS [1 ]
Bonin, V [1 ]
Grabowski, JJ [1 ]
Chao, EYS [1 ]
机构
[1] Mayo Clin & Mayo Fdn, Dept Orthoped, Rochester, MN 55905 USA
关键词
biomechanics; cervical spine; spinal instrumentation;
D O I
10.1097/00007632-199804150-00009
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. An in vitro biomechanical study. Objectives. To simulate a severe compressive flexion injury for determination of the relative stability of different anterior instrumentation systems in a porcine model and to validate this model in human cadaveric specimen. Summary of Background Data. Anterior plate fixation is useful for high-grade mechanical insufficiency of the cervical spine and may prevent the need for a second procedure. Methods. The cervical spines of 45 porcine and 12 cadaveric specimens were subjected to nondestructive flexion, lateral bending, and torsional testing on a modified universal testing machine. A corpectomy was performed with release of the posterior ligamentous structures. The specimens were stabilized with one of three anterior plate constructs. The nondestructive testing was repeated to evaluate structural stability (stiffness and neutral zone). Finally, destructive testing examined failure moment, energy to failure, and mechanism of failure. Results. The instrumented specimens had flexural and lateral bending and torsional stiffness values that were similar to or greater than those of their paired intact specimens. The cervical spine locking plate had a significantly higher flexural stiffness ratio (plated:intact), torsional stiffness ratio, lower flexural neutral zone ratio, higher failure moment, and higher energy to failure than did the Caspar plate. Conclusions. The cervical spine locking plate is theoretically safer than the Caspar system because the posterior vertebral body cortex is not breached by the fixation screws, and the screws are less likely to back out anteriorly and irritate the esophagus. According to these results, the cervical spine locking plate system is biomechanically equivalent to and in some cases more stable than the Caspar system for fixation of a severe compressive flexion injury.
引用
收藏
页码:886 / 892
页数:7
相关论文
共 50 条
  • [21] Biomechanical testing of an artificial cervical joint and an anterior cervical plate
    DiAngelo, DJ
    Roberston, JT
    Metcalf, NH
    McVay, BJ
    Davis, RC
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2003, 16 (04): : 314 - 323
  • [22] Biomechanical evaluation of anterior and posterior fixations in an unstable calf spine model
    Lim, TH
    An, HS
    Hong, JH
    Ahn, JY
    You, JW
    Eck, J
    McGrady, LM
    SPINE, 1997, 22 (03) : 261 - 266
  • [23] Biomechanical Evaluation of Anterior Thoracic Salvage Screws in the Osteoporotic Thoracic Spine
    Behrend, Caleb
    George, Stephen
    Molinari, Robert
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2013, 26 (06): : E235 - E239
  • [24] SUB-AXIAL CERVICAL DISSOCIATION - ANATOMIC AND BIOMECHANICAL PRINCIPLES OF STABILIZATION
    MCLAIN, RF
    ARETAKIS, A
    MOSELEY, TA
    SER, P
    BENSON, DR
    SPINE, 1994, 19 (06) : 653 - 659
  • [25] Comparative studies of cervical spine anterior stabilization systems - Finite element analysis
    Mackiewicz, A.
    Banach, M.
    Denisiewicz, A.
    Bedzinski, R.
    CLINICAL BIOMECHANICS, 2016, 32 : 72 - 79
  • [26] Cervical spine stabilization by plate-cage. Clinical evaluation
    Hakalo, Jerzy
    Wronski, Jerzy
    NEUROLOGIA I NEUROCHIRURGIA POLSKA, 2007, 41 (01) : 44 - 54
  • [27] Biomechanical comparison of cervical spine interbody fusion cages
    Kandziora, F
    Pflugmacher, R
    Schäfer, J
    Born, C
    Duda, G
    Haas, NP
    Mittlmeier, T
    SPINE, 2001, 26 (17) : 1850 - 1857
  • [28] Biomechanical Analysis of the Intact and Destabilized Sheep Cervical Spine
    DeVries, Nicole A.
    Gandhi, Anup A.
    Fredericks, Douglas C.
    Grosland, Nicole M.
    Smucker, Joseph D.
    SPINE, 2012, 37 (16) : E957 - E963
  • [29] Biomechanical evaluation of hydroxyapatite intervertebral graft and anterior cervical plating in a porcine cadaveric model
    Takahashi, T
    Tominaga, T
    Yoshimoto, T
    Koshu, K
    Yokobori, AT
    Aizawa, Y
    BIO-MEDICAL MATERIALS AND ENGINEERING, 1997, 7 (02) : 121 - 127
  • [30] BIOMECHANICAL STUDY OF CERVICAL SPINE WITH PATHOLOGY
    Teixeira, Tatiana
    Sousa, Luisa
    Parente, Marco
    Natal Jorge, R.
    Maia Goncalves, J.
    Freitas, Rolando
    M2D2015: PROCEEDINGS OF THE 6TH INTERNATIONAL CONFERENCE ON MECHANICS AND MATERIALS IN DESIGN, 2015, : 1595 - 1600