Experimental verification of a commercial Monte Carlo-based dose calculation module for high-energy photon beams

被引:26
作者
Kuenzler, Thomas [1 ]
Fotina, Irina [1 ]
Stock, Markus [1 ]
Georg, Dietmar [1 ]
机构
[1] Med Univ Vienna AKH Wien, Dept Radiotherapy, Div Med Radiat Phys, Vienna, Austria
关键词
TREATMENT PLANNING SYSTEM; HETEROGENEOUS MULTILAYER PHANTOM; ANALYTICAL ANISOTROPIC ALGORITHM; ELECTRON-BEAM; DOSIMETRIC VERIFICATION; CLINICAL IMPLEMENTATION; LINEAR-ACCELERATOR; MV; RADIOCHROMIC FILM; IMRT TREATMENT;
D O I
10.1088/0031-9155/54/24/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The dosimetric performance of a Monte Carlo algorithm as implemented in a commercial treatment planning system (iPlan, BrainLAB) was investigated. After commissioning and basic beam data tests in homogenous phantoms, a variety of single regular beams and clinical field arrangements were tested in heterogeneous conditions (conformal therapy, arc therapy and intensity-modulated radiotherapy including simultaneous integrated boosts). More specifically, a cork phantom containing a concave-shaped target was designed to challenge the Monte Carlo algorithm in more complex treatment cases. All test irradiations were performed on an Elekta linac providing 6, 10 and 18 MV photon beams. Absolute and relative dose measurements were performed with ion chambers and near tissue equivalent radiochromic films which were placed within a transverse plane of the cork phantom. For simple fields, a 1D gamma (gamma) procedure with a 2% dose difference and a 2 mm distance to agreement (DTA) was applied to depth dose curves, as well as to inplane and crossplane profiles. The average gamma value was 0.21 for all energies of simple test cases. For depth dose curves in asymmetric beams similar gamma results as for symmetric beams were obtained. Simple regular fields showed excellent absolute dosimetric agreement to measurement values with a dose difference of 0.1% +/- 0.9% (1 standard deviation) at the dose prescription point. A more detailed analysis at tissue interfaces revealed dose discrepancies of 2.9% for an 18 MV energy 10 x 10 cm(2) field at the first density interface from tissue to lung equivalent material. Small fields (2 x 2 cm2) have their largest discrepancy in the re-build-up at the second interface (from lung to tissue equivalent material), with a local dose difference of about 9% and a DTA of 1.1 mm for 18 MV. Conformal field arrangements, arc therapy, as well as IMRT beams and simultaneous integrated boosts were in good agreement with absolute dose measurements in the heterogeneous phantom. For the clinical test cases, the average dose discrepancy was 0.5% +/- 1.1%. Relative dose investigations of the transverse plane for clinical beam arrangements were performed with a 2D gamma-evaluation procedure. For 3% dose difference and 3 mm DTA criteria, the average value for gamma(>1) was 4.7% +/- 3.7%, the average gamma(1%) value was 1.19 +/- 0.16 and the mean 2D gamma-value was 0.44 +/- 0.07 in the heterogeneous phantom. The iPlan MC algorithm leads to accurate dosimetric results under clinical test conditions.
引用
收藏
页码:7363 / 7377
页数:15
相关论文
共 19 条
  • [1] Advanced kernel methods vs. Monte Carlo-based dose calculation for high energy photon beams
    Fotina, Irina
    Winkler, Peter
    Kuenzler, Thomas
    Reiterer, Jochen
    Simmat, Isabell
    Georg, Dietmar
    RADIOTHERAPY AND ONCOLOGY, 2009, 93 (03) : 645 - 653
  • [2] Evaluation of a new commercial Monte Carlo dose calculation algorithm for electron beams
    Vandervoort, Eric J.
    Tchistiakova, Ekaterina
    La Russa, Daniel J.
    Cygler, Joanna E.
    MEDICAL PHYSICS, 2014, 41 (02) : 021711
  • [3] Experimental verification and clinical implementation of a commercial Monte Carlo electron beam dose calculation algorithm
    Fragosoa, Margarida
    Pillai, Sushakumari
    Solberg, Timothy D.
    Chetty, Indrin J.
    MEDICAL PHYSICS, 2008, 35 (03) : 1028 - 1038
  • [4] Monte Carlo-based dose calculation engine for minibeam radiation therapy
    Martinez-Rovira, I.
    Sempau, J.
    Prezado, Y.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (01): : 57 - 62
  • [5] Experimental validation of a Monte Carlo-based treatment-planning system for electron beams
    Mika, Sabine
    Christ, Gunter
    STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 (03) : 150 - 156
  • [6] Experimental validation of a Monte Carlo-based treatment-planning system for electron beams
    Mika S.
    Christ G.
    Strahlentherapie und Onkologie, 2007, 183 (3) : 150 - 156
  • [7] Monte Carlo-based correction factors for ion chamber dosimetry in heterogeneous phantoms for megavoltage photon beams
    Araki, Fujio
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22) : 7615 - 7627
  • [8] Development of a virtual source model for Monte Carlo-based independent dose calculation for varian linac
    Castle, James R.
    Duan, Jingwei
    Feng, Xue
    Chen, Quan
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05):
  • [9] Experimental and computational verifications of the dose calculation accuracy of PHITS for high-energy photon beam therapy
    Kuga, Naoya
    Shiiba, Takuro
    Sato, Tatsuhiko
    Hashimoto, Shintaro
    Kuroiwa, Yasuyoshi
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2024, 61 (01) : 136 - 145
  • [10] Performance of dose calculation algorithms from three generations in lung SBRT: comparison with full Monte Carlo-based dose distributions
    Ojala, Jarkko J.
    Kapanen, Mika K.
    Hyodynmaa, Simo J.
    Wigren, Tuija K.
    Pitkanen, Maunu A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 4 - 18