Determination of Chlorine by Inductively Coupled Plasma Atomic Emission Spectrometry Using the Oxidation Introduction Method

被引:4
作者
Hashimoto, Bunji [1 ]
Daidoji, Hidehiro [1 ]
Uchihara, Hiroshi [1 ]
机构
[1] HORIBA Ltd, Minami Ku, Kyoto 6018510, Japan
关键词
ICP-AES; chlorine; oxidation introduction; halogen;
D O I
10.2116/bunsekikagaku.59.627
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Using atomic emission spectrometry, it is difficult to achieve high-sensitivity analysis of chlorine, due to its high excitation energy. Only 5% of the total sample volume contributes to the emission intensity, using a nebulizing system of inductively coupled plasma atomic emission spectrometry (ICP-AES). In this study we propose a new oxidation vaporization method, based on the introduction of gaseous chlorine into the plasma. High-sensitivity chlorine analysis has been achieved by this method, due to an almost 100% conversion of chlorine dissolved in the sample into the gaseous state. As a result, the sensitivity improvement achieved, and this new method can be used with the commercial introduction system, with a 20-times improved sensitivity for chlorine, compared with the typical nebulizing system. Results obtained by the new method were confirmed using mineral drinking water and river water. The measured values were almost identical to those obtained by ion chromatograph, while the measurement took only one minute.
引用
收藏
页码:627 / 631
页数:5
相关论文
共 3 条
[1]  
Houk R., 1986, ANAL CHEM, V58, P97
[2]   Use of a charge-coupled device detector in the 120-190 nm range in axially-viewed inductively coupled plasma atomic emission spectrometry [J].
Houseaux, J ;
Mermet, JM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (08) :979-982
[3]   Optimization of the conditions of oxidation vapor generation for determining chlorine, bromine, and iodine in aqueous solutions by inductively coupled plasma atomic-emission spectrometry [J].
Vtorushina, E. A. ;
Saprykin, A. I. ;
Knapp, G. .