Airway Scope versus Macintosh laryngoscope: a manikin study

被引:27
作者
Hirabayashi, Yoshihiro [1 ]
机构
[1] Jichi Med Univ, Dept Anesthesiol & Crit Care Med, Shimotsuke, Tochigi 3290498, Japan
关键词
D O I
10.1136/emj.2006.045013
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To compare the performance of the Airway Scope for tracheal intubation with that of the Macintosh laryngoscope for direct laryngoscopy. Methods: 10 anaesthetists, 10 residents, 10 nurses and 10 students intubated the trachea of a manikin using the Airway Scope or the Macintosh laryngoscope in each of 4 laryngoscopy scenarios. Results: Within each group of operators, there was no significant difference in the time taken for intubation between the Airway Scope and the Macintosh laryngoscope. However, the overall number of teeth clicks was less in the Airway Scope, and the Airway Scope scored better on the visual analogue scale than did the Macintosh laryngoscope. Conclusion: The Airway Scope performed as well as and perhaps better than the Macintosh laryngoscope when used on a manikin.
引用
收藏
页码:357 / 358
页数:2
相关论文
共 1 条
  • [1] Description and first clinical application of AirWay Scope for tracheal intubation
    Koyama, Jun-ichi
    Aoyama, Tatsuro
    Kusano, Yoshikazu
    Seguchi, Tatsuya
    Kawagishi, Kyutaro
    Iwashita, Tomomi
    Okamoto, Kazufumi
    Okudera, Hiroshi
    Takasuna, Hiroshi
    Hongo, Kazuhiro
    [J]. JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 2006, 18 (04) : 247 - 250