Investigation of 18F-FDG 3D mode PET image quality versus acquisition time

被引:13
作者
Brown, Colin [1 ]
Dempsey, Mary-Frances [1 ]
Gillen, Gerry [1 ]
Elliott, Alex T. [1 ]
机构
[1] Gartnavel Royal Hosp, W Scotland PET Ctr, Glasgow G12 0YN, Lanark, Scotland
关键词
acquisition time; image quality; positron emission tomography; IMPACT; 2D;
D O I
10.1097/MNM.0b013e3283355c5d
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Three-dimensional (3D) mode positron emission tomography (PET) is being used increasingly for clinical PET imaging. However, as yet, optimal acquisition parameters have not been established. The aim of this study was to investigate the effect of varying acquisition time on 3D image quality using standard clinical activities of F-18-fluorodeoxyglucose (FDG). Methods F-18-FDG phantom and patient PET images were acquired with varying acquisition times on a GE Discovery-STE PET/CT system. The NEMA Image-Quality phantom was imaged with four hot lesions in a uniform background. Images were acquired for 1, 2, 3, 4, and 5-min frames with three different lesion-to-background contrast ratios. Patient data were investigated using list mode acquisition to obtain comparable 2, 3, and 4-min frames. Qualitative analysis involved grading image quality and lesion detectability. Quantitative analysis of phantom images involved assessing the coefficient of variation (COV) of background areas as a measure of noise, and lesion over background variability as a measure of image quality. Patient data were also assessed using COV analysis of liver uptake. Results Qualitative and quantitative analysis showed no significant difference in image quality between 4 and 5-min acquisition frames for 3D mode F-18-FDG PET imaging with standard clinical activities. The observers noted no difference in perceived image quality. This finding was supported by COV analysis. Conclusion This study shows that GE Discovery-STE acquisition frame time can be reduced to 3 min for standard 3D mode imaging at standard clinical activities of F-18-FDG. Nucl Med Commun 31: 254-259 (C) 2010 Wolters Kluwer Health | Lippincott Williams & Wilkins.
引用
收藏
页码:254 / 259
页数:6
相关论文
共 9 条
  • [1] Bettinardi V, 2007, Q J NUCL MED MOL IM, V51, P214
  • [2] The Rose model, revisited
    Burgess, AE
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1999, 16 (03): : 633 - 646
  • [3] Impact of acquisition geometry, image processing, and patient size on lesion detection in whole-body 18F-FDG PET
    El Fakhri, Georges
    Santos, Paula A.
    Badawi, Ramsey D.
    Holdsworth, Clay H.
    Van den Abbeele, Annick D.
    Kijewski, Marie Foley
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (12) : 1951 - 1960
  • [4] Halpern BS, 2004, J NUCL MED, V45, P797
  • [5] Kemp BJ, 2006, J NUCL MED, V47, P1960
  • [6] Lodge MA, 2006, J NUCL MED, V47, P23
  • [7] Impact of body habitus on quantitative and qualitative image quality in whole-body FDG-PET
    Tatsumi, M
    Clark, PA
    Nakamoto, Y
    Wahl, RL
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (01) : 40 - 45
  • [8] Viera AJ, 2005, FAM MED, V37, P360
  • [9] Clinical evaluation of 2D versus 3D whole-body PET image quality using a dedicated BGO PET scanner
    Visvikis, D
    Griffiths, D
    Costa, DC
    Bomanji, J
    Ell, PJ
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 (09) : 1050 - 1056