The aim of the study was to analyse BTEX compounds ( benzene, toluene, ethylbenzene, xylenes) in air by solid phase micro-extraction/gas chromatography/ mass spectrometry (SPME/GC/MS), and this article presents the features of the calibration method proposed. Examples of real-world air analysis are given. Standard gaseous mixtures of BTEX in air were generated by dynamic dilution. SPME sampling was carried out under non-equilibrium conditions using a Carboxen/ PDMS fibre exposed for 30 min to standard gas mixtures or to ambient air. The behaviour of the analytical response was studied from 0 to 65 mug/m(3) by adding increasing amounts of BTEX to the air matrix. Detection limits range from 0.05 to 0.1 mug/m(3) for benzene, depending on the fibre. Inter-fibre relative standard deviations ( reproducibility) are larger than 18%, although the repeatability for an individual fibre is better than 10%. Therefore, each fibre should be considered to be a particular sampling device, and characterised individually depending on the required accuracy. Sampling indoor and outdoor air by SPME appears to be a suitable short-delay diagnostic method for volatile organic compounds, taking advantage of short sampling time and simplicity.
引用
收藏
页码:824 / 830
页数:7
相关论文
共 36 条
[1]  
*AIRPARIF, 2000, AIRPARIF ACT SURV QU
[2]   Solid-phase microextraction and gas chromatography-mass spectrometry for determination of monoaromatic hydrocarbons in blood and urine: Application to people exposed to air pollutants [J].
Andreoli, R ;
Manini, P ;
Bergamaschi, E ;
Brustolin, A ;
Mutti, A .
CHROMATOGRAPHIA, 1999, 50 (3-4) :167-172
[3]  
[Anonymous], 2003, STAT QUALITY CONTROL
[4]   SOLID-PHASE MICROEXTRACTION WITH THERMAL-DESORPTION USING FUSED-SILICA OPTICAL FIBERS [J].
ARTHUR, CL ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1990, 62 (19) :2145-2148
[5]  
Asakawa F, 1999, B ENVIRON CONTAM TOX, V62, P109
[6]   Atmosphere generation system for the preparation of ambient air volatile organic compound standard mixtures [J].
Ballesta, PP ;
Baldan, A ;
Cancelinha, J .
ANALYTICAL CHEMISTRY, 1999, 71 (11) :2241-2245
[7]   Solid-phase microextraction coupled with high-performance liquid chromatography for the determination of alkylphenol ethoxylate surfactants in water [J].
BoydBoland, AA ;
Pawliszyn, JB .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1521-1529
[8]   ANALYSIS OF ENVIRONMENTAL AIR SAMPLES BY SOLID-PHASE MICROEXTRACTION AND GAS-CHROMATOGRAPHY ION-TRAP MASS-SPECTROMETRY [J].
CHAI, M ;
PAWLISZYN, J .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1995, 29 (03) :693-701
[9]   SOLID-PHASE MICROEXTRACTION COUPLED TO HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
CHEN, J ;
PAWLISZYN, JB .
ANALYTICAL CHEMISTRY, 1995, 67 (15) :2530-2533
[10]   Nomenclature in evaluation of analytical methods including detection and quantification capabilities (IUPAC Recommendations 1995) [J].
Currie, LA .