Computational analysis of the high pass birdcage resonator: finite difference time domain simulations for high-field MRI

被引:63
作者
Ibrahim, TS [1 ]
Lee, R [1 ]
Baertlein, BA [1 ]
Yu, Y [1 ]
Robitaille, PML [1 ]
机构
[1] Ohio State Univ, Ctr Adv Biomed Imaging, Columbus, OH 43210 USA
关键词
RF coil; simulations; modeling; birdcage resonator; B-1; field;
D O I
10.1016/S0730-725X(00)00161-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this work, a finite difference time domain (FDTD) algorithm is validated at 1.5 tesla using the standard GE Signa quadrature head coil and a muscle phantom. The electrical characteristics of the birdcage head coil are then calculated for the linear and quadrature cases. Unlike previous computational analysis which assume idealized currents on the end rings and the struts of the resonator, a complete computational analysis is provided. This treatment considers the coupling between the resonator and the sample and includes a real coil excitation, a complete current derivation, and a thorough description of both B-1 fields and RF radiation. With this improvement, electromagnetic phenomena such as radiation, standing wave currents on the wires, and field inhomogeneities due to interactions between the coil and the load inside the coil are observed. At 200 MHz, it is demonstrated that this particular coil does not work well due to radiation and non-uniformities on the struts of the device. Also, at this frequency magnetic field inhomogeneities become large when the coil is loaded with a phantom.
引用
收藏
页码:835 / 843
页数:9
相关论文
共 16 条
[1]  
BERENGER JA, 1994, COMPUTATIONAL PH OCT
[2]   Numerical simulation of SAR and B1-field inhomogeneity of shielded RF coils loaded with the human head [J].
Chen, J ;
Feng, ZM ;
Jin, JM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1998, 45 (05) :650-659
[3]   SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil [J].
Collins, CM ;
Li, SZ ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (06) :847-856
[4]   MOMENT-METHOD SOLUTIONS AND SAR CALCULATIONS FOR INHOMOGENEOUS MODELS OF MAN WITH LARGE NUMBER OF CELLS [J].
DEFORD, JF ;
GANDHI, OP ;
HAGMANN, MJ .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1983, 31 (10) :848-851
[5]  
FIELD S, 1987, PHYSICS TECHNOLOGY H
[6]  
Gandhi OP, 1990, BIOL EFFECTS MED APP
[7]  
HAN Y, 1993, P SMRM 12 ANN M, P1327
[8]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[9]   MAPPING OF THE RADIOFREQUENCY FIELD [J].
INSKO, EK ;
BOLINGER, L .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1993, 103 (01) :82-85
[10]   Computation of electromagnetic fields for high-frequency magnetic resonance imaging applications [J].
Jin, JM ;
Chen, J ;
Chew, WC ;
Gan, H ;
Magin, RL ;
Dimbylow, PJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (12) :2719-2738