Direct determination of cyanides by potentiometric biosensors

被引:40
作者
Keusgen, M
Kloock, JP
Knobbe, DT
Jünger, M
Krest, I
Goldbach, M
Klein, W
Schöning, MJ
机构
[1] Univ Marburg, Inst Pharmaceut Chem, D-35032 Marburg, Germany
[2] Univ Appl Sci Aachen, Div Julich, Lab Chem Sensors & Biosensors, D-52428 Julich, Germany
[3] Res Ctr Julich GmbH, Inst Thin Films & Interfaces, ISG2, D-52425 Julich, Germany
[4] Univ Bonn, Inst Pharmaceut Biol, D-53115 Bonn, Germany
关键词
cyanidase; EIS structure; potentiometric ISE; silicon-based biosensor; cyanides;
D O I
10.1016/j.snb.2004.04.067
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cyanide (CN-) and its derivatives are widely used in the metal mining and metal plating industry. But also many medicinal and food plants are toxic for man because of their content of cyanogenic glycosides. In the present investigation, a cyanide-specific biosensor exploiting immobilized cyanidase (EC 3.5.5.1) has been developed. Enzymatically formed ammonia was either detected by a potentiometric sensor based on an ammonia electrode or by a pH-sensitive electrolyte/insulator/semiconductor (EIS) layer structure made of Al/p-Si/SiO2/Si3N4. It could be demonstrated with both methods that sensitivity of these biosensors is sufficient to detect cyanide in sub-toxic concentrations. The detection limit was determined to be in the micromolar concentration range. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:380 / 385
页数:6
相关论文
共 13 条
[1]  
BACLAY M, 1998, ENZYME MICROB TECHNO, V23, P321
[2]   SECONDARY METABOLITES IN PLANT DEFENSE-MECHANISMS [J].
BENNETT, RN ;
WALLSGROVE, RM .
NEW PHYTOLOGIST, 1994, 127 (04) :617-633
[3]   Determination of cyanide using flow-injection multisensor system [J].
Ipatov, A ;
Ivanov, M ;
Makarychev-Mikhailov, S ;
Kolodnikov, V ;
Legin, A ;
Vlasov, Y .
TALANTA, 2002, 58 (06) :1071-1076
[4]   DETERMINATION OF LOW-LEVELS OF CYANIDE WITH A SILVER SILVER SULFIDE WIRE ELECTRODE [J].
JOVANOVIC, VM ;
SAKBOSNAR, M ;
JOVANOVIC, MS .
ANALYTICA CHIMICA ACTA, 1987, 196 :221-227
[5]   Development of a biosensor specific for cysteine sulfoxides [J].
Keusgen, M ;
Jünger, M ;
Krest, I ;
Schöning, MJ .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (5-6) :805-812
[6]   Biosensoric flow-through method for the determination of cysteine sulfoxides [J].
Krest, I ;
Keusgen, M .
ANALYTICA CHIMICA ACTA, 2002, 469 (02) :155-164
[7]  
MIDGLEY D, 1978, POTENTIOMETRIC WATER, P298
[8]   Amperometric biosensor for formic acid in air [J].
Sandström, KJM ;
Newman, J ;
Sunesson, AL ;
Levin, JO ;
Turner, APF .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 70 (1-3) :182-187
[9]  
SCHONING MJ, 1999, MEASUREMENT INSTRUME, P70
[10]   A long-term stable penicillin-sensitive potentiometric biosensor with enzyme immobilized by heterobifunctional cross-linking [J].
Thust, M ;
Schoning, MJ ;
Vetter, J ;
Kordos, P ;
Luth, H .
ANALYTICA CHIMICA ACTA, 1996, 323 (1-3) :115-121