Ratiometry of transmembrane voltage-sensitive fluorescent dye emission in hearts

被引:84
作者
Knisley, SB [1 ]
Justice, RK [1 ]
Kong, W [1 ]
Johnson, PL [1 ]
机构
[1] Univ Alabama, Sch Engn, Dept Biomed Engn, Birmingham, AL 35294 USA
来源
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY | 2000年 / 279卷 / 03期
关键词
potentiometric dye; pyridinium; 4-(2-(6-(dibutylamino)-2-naphthalenyl) ethenyl)-1-(3-sulfopropyl)-; hydroxide; inner salt; potassium; action potential;
D O I
10.1152/ajpheart.2000.279.3.H1421
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Transmembrane voltage-sensitive fluorescence measurements are limited by baseline drift that can obscure changes in resting membrane potential and by motion artifacts that can obscure repolarization. Voltage-dependent shift of emission wavelengths may allow reduction of drift and motion artifacts by emission ratiometry. We have tested this for action potentials and potassium-induced changes in resting membrane potential in rabbit hearts stained with di-4-ANEPPS [Pyridinium, 4-(2-(6-(dibutylamino)-2-naphthalenyl) ethenyl)-1-(3-sulfopropyl)-, hydroxide, inner salt] using laser excitation (488 nm) and a two-photomultiplier tube system or spectrofluorometer (resolution of 500-1,000 Hz and <1 mm). Green and red emissions produced upright and inverted action potentials, respectively. Ratios of green emission to red emission followed action potential contours and exhibited larger fractional changes than either emission alone (P < 0.001). The largest changes and signal-to-noise ratio (signal/noise) were obtained with numerator wavelengths of 525-550 nm and denominator wavelengths of 650-700 nm. Ratiometry lessened drift 56-66% (P < 0.015) and indicated decreases in resting membrane potential. Ratiometry lessened motion artifacts and increased magnitudes of deflections representing phase-zero depolarizations relative to total deflections by 123-188% in intact hearts (P < 0.02). Durations of action potentials at different pacing rates, temperatures, and potassium concentrations were independent of whether they were measured ratiometrically or with microelectrodes (P greater than or equal to 0.65). The ratiometric calibration slope was 0.017/100 mV and decreased with time. Thus emission ratiometry lessens the effects of motion and drift and indicates resting membrane potential changes and repolarization.
引用
收藏
页码:H1421 / H1433
页数:13
相关论文
共 26 条
[1]   EXPERIMENTAL ELECTROPHYSIOLOGY AND ARRHYTHMOGENICITY - ANISOTROPY AND VENTRICULAR-TACHYCARDIA [J].
ALLESSIE, MA ;
SCHALIJ, MJ ;
KIRCHHOF, CJHJ ;
BOERSMA, L ;
HUYBERS, M ;
HOLLEN, J .
EUROPEAN HEART JOURNAL, 1989, 10 :2-8
[2]   Ratiometric measurement of endothelial depolarization in arterioles with a potential-sensitive dye [J].
Beach, JM ;
McGahren, ED ;
Xia, J ;
Duling, BR .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1996, 270 (06) :H2216-H2227
[3]   Differential effects of cytochalasin D and 2,3 butanedione monoxime on isometric twitch force and transmembrane action potential in isolated ventricular muscle: Implications for optical measurements of cardiac repolarization [J].
Biermann, M ;
Rubart, M ;
Moreno, A ;
Wu, JS ;
Josiah-Durant, A ;
Zipes, DP .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (12) :1348-1357
[4]   SUPPRESSION OF MOTION ARTIFACTS IN FLUORESCENCE SPECTROSCOPY OF PERFUSED HEARTS [J].
BRANDES, R ;
FIGUEREDO, VM ;
CAMACHO, SA ;
MASSIE, BM ;
WEINER, MW .
AMERICAN JOURNAL OF PHYSIOLOGY, 1992, 263 (03) :H972-H980
[5]   High-speed, random-access fluorescence microscopy: II. Fast quantitative measurements with voltage-sensitive dyes [J].
Bullen, A ;
Saggau, P .
BIOPHYSICAL JOURNAL, 1999, 76 (04) :2272-2287
[6]   A NEW LASER SCANNING SYSTEM FOR MEASURING ACTION-POTENTIAL PROPAGATION IN THE HEART [J].
DILLON, S ;
MORAD, M .
SCIENCE, 1981, 214 (4519) :453-456
[7]   SYNCHRONIZED REPOLARIZATION AFTER DEFIBRILLATION SHOCKS - A POSSIBLE COMPONENT OF THE DEFIBRILLATION PROCESS DEMONSTRATED BY OPTICAL RECORDINGS IN RABBIT HEART [J].
DILLON, SM .
CIRCULATION, 1992, 85 (05) :1865-1878
[8]   OPTICAL MAPPING OF REPOLARIZATION AND REFRACTORINESS FROM INTACT HEARTS [J].
EFIMOV, IR ;
HUANG, DT ;
RENDT, JM ;
SALAMA, G .
CIRCULATION, 1994, 90 (03) :1469-1480
[9]   Activation and repolarization patterns are governed by different structural characteristics of ventricular myocardium: Experimental study with voltage-sensitive dyes and numerical simulations [J].
Efimov, IR ;
Ermentrout, B ;
Huang, DT ;
Salama, G .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (06) :512-530
[10]   SPECTRA, MEMBRANE-BINDING, AND POTENTIOMETRIC RESPONSES OF NEW CHARGE SHIFT PROBES [J].
FLUHLER, E ;
BURNHAM, VG ;
LOEW, LM .
BIOCHEMISTRY, 1985, 24 (21) :5749-5755