Closed-cycle cryocooled SQUID system with superconductive shield for biomagnetism

被引:8
作者
Yu, Kwon Kyu [1 ]
Lee, Yong Ho [1 ]
Lee, Seong Joo [1 ]
Shim, Jeong Hyun [1 ]
Hwang, Seong Min [1 ]
Kim, Jin Mok [1 ]
Kwon, Hyuckchan [1 ]
Kim, Kiwoong [1 ]
机构
[1] Korea Res Inst Stand & Sci, Ctr Biosignals, Taejon 305340, South Korea
关键词
SQUID magnetometer; magnetocardiogram; biomagnetism; cryocooler; superconductive shield; ferromagnetic shield; MAGNETOENCEPHALOGRAPHY; MAGNETOCARDIOGRAPHY; GRADIOMETERS;
D O I
10.1088/0953-2048/27/10/105007
中图分类号
O59 [应用物理学];
学科分类号
摘要
We developed a cryocooled SQUID system with which human magnetocardiogram (MCG) and possibly magnetoenceparogram (MEG) can be measured. To reduce cyclic magnetic noises originating from the regenerator of the cold heads of the cryocooler, a superconductive shield (99.5% Pb) was used to protect the SQUID sensors, and a ferromagnetic shield (78% Ni alloy) was used to screen the cold head. In addition, the SQUID sensors' chamber was placed at a distance of 1.8 m from the cold head chamber to install the cold-head chamber outside the magnetically shielded room (MSR) for future development. The loss in cooling power due to the increased distance was compensated by increasing the number of thermal rods, and thus the SQUID sensor and superconductive shield could be refrigerated to 4.8 K and 5 K, respectively. The superconductive shield successfully rejected thermal noise emitted from metallic blocks used to improve thermal conduction. The noise of the SQUID system was 3 fT/Hz(1/2), and the cyclic magnetic noise could be reduced to 1.7 pT. We could obtain a clear MCG signal while the entire cryogenics was in operation without any special digital processing.
引用
收藏
页数:9
相关论文
共 30 条
[1]   SQUID-Based Microwave Cavity Search for Dark-Matter Axions [J].
Asztalos, S. J. ;
Carosi, G. ;
Hagmann, C. ;
Kinion, D. ;
van Bibber, K. ;
Hotz, M. ;
Rosenberg, L. J. ;
Rybka, G. ;
Hoskins, J. ;
Hwang, J. ;
Sikivie, P. ;
Tanner, D. B. ;
Bradley, R. ;
Clarke, J. .
PHYSICAL REVIEW LETTERS, 2010, 104 (04)
[2]   Operation of a high-Tc DC-SQUID-gradiometer on a non-metallic pulse tube refrigerator [J].
Becker, C. ;
Steppke, A. ;
Koettig, T. ;
Gerster, J. ;
Doerrer, L. ;
Thuerk, M. ;
Schmidl, F. ;
Seidel, P. .
8TH EUROPEAN CONFERENCE ON APPLIED SUPERCONDUCTIVITY (EUCAS'07), 2008, 97
[3]   DISCRETE SPATIAL-FILTERING WITH SQUID GRADIOMETERS IN BIOMAGNETISM [J].
BRUNO, AC ;
RIBEIRO, PC ;
VONDERWEID, JP ;
SYMKO, OG .
JOURNAL OF APPLIED PHYSICS, 1986, 59 (07) :2584-2589
[4]   The perception of pain in others suppresses somatosensory oscillations: A magnetoencephalography study [J].
Cheng, Yawei ;
Yang, Chia-Yen ;
Lin, Ching-Po ;
Lee, Po-Lei ;
Decety, Jean .
NEUROIMAGE, 2008, 40 (04) :1833-1840
[5]  
Clem TR, 1996, NATO ADV SCI I E-APP, V329, P517
[6]   Characterization of a low frequency magnetic noise from a two-stage pulse tube cryocooler [J].
Eshraghi, M. J. ;
Sasada, I. ;
Kim, J. M. ;
Lee, Y. H. .
CRYOGENICS, 2009, 49 (07) :334-339
[7]   Superconducting quantum interference device instruments and applications [J].
Fagaly, R. L. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (10)
[8]   COOLING OF SQUIDS USING A GIFFORD-MCMAHON CRYOCOOLER CONTAINING MAGNETIC REGENERATIVE MATERIAL TO MEASURE BIOMAGNETISM [J].
FUJIMOTO, S ;
KAZAMI, K ;
TAKADA, Y ;
YOSHIDA, T ;
OGATA, H ;
KADO, H .
CRYOGENICS, 1995, 35 (02) :143-148
[9]   MAGNETOENCEPHALOGRAPHY - THEORY, INSTRUMENTATION, AND APPLICATIONS TO NONINVASIVE STUDIES OF THE WORKING HUMAN BRAIN [J].
HAMALAINEN, M ;
HARI, R ;
ILMONIEMI, RJ ;
KNUUTILA, J ;
LOUNASMAA, OV .
REVIEWS OF MODERN PHYSICS, 1993, 65 (02) :413-497
[10]   A DRUNG-TYPE MAGNETOMETER MOUNTED ON A GM CRYOCOOLER [J].
KAZAMI, K ;
TAKADA, Y ;
FUJIMOTO, S ;
YOSHIDA, T ;
OGATA, H ;
KADO, H .
SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1994, 7 (05) :256-259