Applicability of supercritical fluid chromatography - mass spectrometry to metabolomics. I - Optimization of separation conditions for the simultaneous analysis of hydrophilic and lipophilic substances

被引:76
作者
Desfontaine, Vincent [1 ]
Losacco, Gioacchino Luca [1 ]
Gagnebin, Yoric [1 ]
Pezzatti, Julian [1 ]
Farrell, William P. [2 ]
Gonzalez-Ruiz, Victor [1 ]
Rudaz, Serge [1 ]
Veuthey, Jean-Luc [1 ]
Guillarme, Davy [1 ]
机构
[1] Univ Geneva, Univ Lausanne, Sch Pharmaceut Sci, CMU, Rue Michel Servet 1, CH-1211 Geneva 4, Switzerland
[2] Pfizer Inc, Worldwide Med Chem, La Jolla Labs, 10770 Sci Ctr Dr, San Diego, CA 92121 USA
关键词
Supercritical fluid chromatography; Polar compounds; Metabolomics; SFC-MS; LIQUID-CHROMATOGRAPHY; MOBILE-PHASE; STATIONARY PHASES; PHARMACEUTICAL COMPOUNDS; SOLUBLE VITAMINS; PERFORMANCE; MS; METABOLITES; SAMPLES; WATER;
D O I
10.1016/j.chroma.2018.05.055
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The aim of this study was to evaluate the suitability of SFC-MS for the analysis of a wide range of compounds including lipophilic and highly hydrophilic substances (log P values comprised between 6 and 11), for its potential application toward human metabolomics. For this purpose, a generic unified chromatography gradient from 2 to 100% organic modifier in CO2 was systematically applied. In terms of chemistry, the best stationary phases for this application were found to be the Agilent Poroshell HILIC (bare silica) and Macherey-Nagel Nucleoshell HILIC (silica bonded with a zwitterionic ligand). To avoid system overpressure at very high organic modifier proportion, columns of 100 x 3 mm I.D. packed with sub-3 mu m superficially porous particles were selected. In terms of organic modifier, a mixture of 95% MeOH and 5% water was selected, with 50 mM ammonium formate and 1 mM ammonium fluoride, to afford good solubility of analytes in the mobile phase, limited retention for the most hydrophilic metabolites and suitable peak shapes of ionizable species. A sample diluent containing 50%ACN/50% water was employed as injection solvent. These conditions were applied to a representative set of metabolites belonging to nucleosides, nucleotides, small organic acids, small bases, sulfated/sulfonated metabolites, poly-alcohols, lipid related substances, quaternary ammonium metabolites, phosphate-based substances, carbohydrates and amino acids. Among all these metabolites, 65% of the compounds were adequately analyzed with excellent peak shape, 23% provided distorted peak shapes, while only 12% were not detected (mostly metabolites having several phosphate or several carboxylic acid groups). (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:96 / 107
页数:12
相关论文
共 47 条
[1]   Effects of liquid post-column addition in electrospray ionization performance in supercritical fluid chromatography-mass spectrometry [J].
Akbal, Laura ;
Hopfgartner, Gerard .
JOURNAL OF CHROMATOGRAPHY A, 2017, 1517 :176-184
[2]   Is supercritical fluid chromatography hyphenated to mass spectrometry suitable for the quality control of vitamin D3 oily formulations? [J].
Andri, B. ;
Dispas, A. ;
Klinkenberg, R. ;
Streel, B. ;
Marini, R. D. ;
Ziemons, E. ;
Hubert, Ph. .
JOURNAL OF CHROMATOGRAPHY A, 2017, 1515 :209-217
[3]   Gradient enhanced-fluidity liquid hydrophilic interaction chromatography of ribonucleic acid nucleosides and nucleotides: A "green" technique [J].
Beilke, Michael C. ;
Beres, Martin J. ;
Olesik, Susan V. .
JOURNAL OF CHROMATOGRAPHY A, 2016, 1436 :84-90
[4]   RPLC-HILIC and SFC with Mass Spectrometry: Polarity-Extended Organic Molecule Screening in Environmental (Water) Samples [J].
Bieber, Stefan ;
Greco, Giorgia ;
Grosse, Sylvia ;
Letzel, Thomas .
ANALYTICAL CHEMISTRY, 2017, 89 (15) :7907-7914
[5]   Hydrophilic interaction liquid chromatography (HILIC)-a powerful separation technique [J].
Buszewski, Boguslaw ;
Noga, Sylwia .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 402 (01) :231-247
[6]  
C.I.o.H.R.A.I.-H.S.T.M.I. Centre, HUM MET DAT
[7]   Evaluation of supercritical fluid chromatography coupled to tandem mass spectrometry for pesticide residues in food [J].
Cutillas, Victor ;
Martinez Galera, Maria ;
Rajski, Lukasz ;
Fernandez-Alba, Amadeo R. .
JOURNAL OF CHROMATOGRAPHY A, 2018, 1545 :67-74
[8]  
Desfontaine V., 2017, Supercritical Fluid Chromatography, P213
[9]   Systematic evaluation of matrix effects in supercritical fluid chromatography versus liquid chromatography coupled to mass spectrometry for biological samples [J].
Desfontaine, Vincent ;
Capetti, Francesca ;
Nicoli, Raul ;
Kuuranne, Tiia ;
Veuthey, Jean-Luc ;
Guillarme, Davy .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2018, 1079 :51-61
[10]   Screening study of SFC critical method parameters for the determination of pharmaceutical compounds [J].
Dispas, Amandine ;
Lebrun, Pierre ;
Sacre, Pierre-Yves ;
Hubert, Philippe .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 125 :339-354