The effect of sterilization on bending and torsional properties of K-files manufactured with different metallic alloys

被引:23
作者
Canalda-Sahli, C [1 ]
Brau-Aguade, E [1 ]
Sentis-Vilalta, J [1 ]
机构
[1] Univ Barcelona, Fac Dent, Dept Dent Pathol & Therapeut, E-08007 Barcelona, Spain
关键词
mechanical properties; nickel-titanium files; stainless-steel files; sterilization; titanium files;
D O I
10.1046/j.1365-2591.1998.t01-1-00114.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to assess the effect of dry-heat or autoclave sterilization on the resistance to fracture in torque and angular deflection and the resistance to bending of It-type files made of nickel-titanium (Nitiflex, Naviflex), titanium (Microtitane) or stainless steel (Flexofile, Flex-R). Ten K-type files of each sort, from size 25 to 40, were tested, according to ANSI/ADA specification 28 (1988) and ISO specification 3630 (1992), Sterilization with dry heat and autoclave slightly decreased the flexibility of files made of stainless steel and nickel-titanium for most of the sizes, although the values obtained satisfied ISO specifications. The files made of titanium showed an increased flexibility after sterilization with autoclave (sizes 30 and 35) and dry heat (sizes 30, 35 and 40), Resistance to fracture varied amongst the five groups of files tested as follows: it decreased in some sizes of stainless-steel instruments, decreased in all sizes of titanium files assessed by the torsional moment, and either increased or decreased in some sizes of nickel-titanium files. All files tested, however, satisfied relevant standards for angular deflection after being subjected to sterilization with an autoclave or dry heat.
引用
收藏
页码:48 / 52
页数:5
相关论文
共 12 条
[1]  
*AM NAT STAND I, 1988, 28 AM NAT STAND I
[2]   COMMENTARY ON RESISTANCE TO FRACTURE BY TWISTING - ANSI ADA SPECIFICATION NO-28 AND NO-58 [J].
BOLGER, WL ;
GOUGH, RW .
JOURNAL OF ENDODONTICS, 1985, 11 (06) :245-248
[3]   A comparison of bending and torsional properties of K-files manufactured with different metallic alloys [J].
CanaldaSahli, C ;
BrauAguade, E ;
BerasteguiJimeno, E .
INTERNATIONAL ENDODONTIC JOURNAL, 1996, 29 (03) :185-189
[4]   Effects of cleaning, chemical disinfection, and sterilization procedures on the mechanical properties of endodontic instruments [J].
Haikel, Y ;
Serfaty, R ;
Bleicher, P ;
Lwin, TTC ;
Allemann, C .
JOURNAL OF ENDODONTICS, 1997, 23 (01) :15-18
[5]  
*INT ORG STAND, 1992, 36302 ISODIS
[6]   THE EFFECTS OF VARIOUS STERILIZATION METHODS ON THE TORSIONAL STRENGTH OF ENDODONTIC FILES [J].
IVERSON, GW ;
VONFRAUNHOFER, JA ;
HERRMANN, JW .
JOURNAL OF ENDODONTICS, 1985, 11 (06) :266-268
[7]   LOAD-DEFLECTION CHARACTERISTICS OF NICKEL-TITANIUM ALLOY WIRES AFTER CLINICAL RECYCLING AND DRY HEAT STERILIZATION [J].
KAPILA, S ;
HAUGEN, JW ;
WATANABE, LG .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1992, 102 (02) :120-126
[8]   THE EFFECT OF AUTOCLAVE STERILIZATION ON ENDODONTIC FILES [J].
MITCHELL, BF ;
JAMES, GA ;
NELSON, RC .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1983, 55 (02) :204-207
[10]   TORSIONAL PROPERTIES OF TWISTED AND MACHINED ENDODONTIC FILES [J].
SETO, BG ;
NICHOLLS, JI ;
HARRINGTON, GW .