Evaluation of Ultraviolet C for Disinfection of Endocavitary Ultrasound Transducers Persistently Contaminated despite Probe Covers

被引:55
作者
Kac, Guillaume [1 ]
Podglajen, Isabelle [2 ,6 ]
Si-Mohamed, Ali [3 ]
Rodi, Aurelia [1 ]
Grataloup, Christine [4 ]
Meyer, Guy [5 ,6 ]
机构
[1] Hop Europeen Georges Pompidou, APHP, Hyg Hosp, F-75015 Paris, France
[2] Hop Europeen Georges Pompidou, APHP, Bacteriol Lab, F-75015 Paris, France
[3] Hop Europeen Georges Pompidou, APHP, Virol Lab, F-75015 Paris, France
[4] Hop Europeen Georges Pompidou, APHP, Serv Radiol, F-75015 Paris, France
[5] Hop Europeen Georges Pompidou, APHP, Serv Pneumol, F-75015 Paris, France
[6] Univ Paris 05, Paris, France
关键词
PSEUDOMONAS-AERUGINOSA INFECTIONS; PROSTATE BIOPSIES; PERFORATION; SONOGRAPHY; OUTBREAK; CONDOMS;
D O I
10.1086/649794
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
OBJECTIVE. To determine the rate of bacterial and viral contamination of endocavitary ultrasound probes after endorectal or endovaginal examination with the use of probe covers and to evaluate the antimicrobial efficacy of a disinfection procedure consisting of cleaning with a disinfectant-impregnated towel followed by disinfection with ultraviolet C (UVC) light. METHODS. Endovaginal or endorectal ultrasound examinations were performed for 440 patients in 3 institutions. All probes were covered by a condom or sheath during the examination. For bacterial analysis, 1 swab was applied lengthwise across one-half the surface of the probe just after removal of the probe cover. The second swab was similarly applied over the probe immediately after the end of a 2-step process consisting of cleaning with a towel impregnated with a disinfectant spray and a 5-minute UVC disinfection cycle. Swabs were applied onto plates and incubated for 48 hours. The number of colony-forming units was counted, and organisms were identified. A similar protocol was used for viral detection of Epstein-Barr virus, human cytomegalovirus, and human papillomavirus, except that an additional swab was applied along the entire external surface of the probe cover before its removal. Viruses were detected by means of a polymerase chain reaction-based protocol. RESULTS. After removal of probe covers, contamination by pathogenic bacteria was found for 15 (3.4% [95% confidence interval, 2.0%-5.6%]) of 440 probes, and viral genome was detected on 5 (1.5% [ 95% confidence interval, 0.5%-3.5%]) of 336 probes. After cleaning with a towel impregnated with a disinfectant spray and disinfecting with UVC light, neither bacterial pathogenic flora nor viral genome was recovered from the probe. CONCLUSIONS. Endocavitary ultrasound probes may carry pathogens after removal of covers under routine conditions. A disinfection procedure consisting of cleaning with a disinfectant-impregnated towel followed by disinfection with UVC may provide a useful method for disinfecting endocavitary ultrasound probes. Infect Control Hosp Epidemiol 2010; 31:165-170
引用
收藏
页码:165 / 170
页数:6
相关论文
共 27 条
  • [1] *AM COLL RAD, 2006, PRACT GUID PERF ULTR
  • [2] Amis S, 2000, J CLIN ULTRASOUND, V28, P295, DOI 10.1002/1097-0096(200007/08)28:6<295::AID-JCU5>3.3.CO
  • [3] 2-7
  • [4] [Anonymous], GUID CLEAN PREP END
  • [5] *AUSTR SOC ULTR ME, 2005, GUID DIS INTR TRANSD
  • [6] Modeling the sexual transmissibility of human papillomavirus infection using stochastic computer simulation and empirical data from a cohort study of young women in Montreal, Canada
    Burchell, AN
    Richardson, H
    Mahmud, SM
    Trottier, H
    Tellier, PP
    Hanley, J
    Coutlée, F
    Franco, EL
    [J]. AMERICAN JOURNAL OF EPIDEMIOLOGY, 2006, 163 (06) : 534 - 543
  • [7] Epidemiology and transmission dynamics of genital HPV infection
    Burchell, Ann N.
    Winer, Rachel L.
    de Sanjose, Silvia
    Franco, Eduardo L.
    [J]. VACCINE, 2006, 24 : 52 - 61
  • [8] Centers for Disease Control and Prevention, 2006, MMWR-MORBID MORTAL W, V55, P1177
  • [9] Department of Health Government of West Australia, 2004, PREV CROSS INF DIAGN
  • [10] *FOOD DRUG ADM, FDA PUBL NOT REPR RE