Technical Note: Development and performance of a software tool for quality assurance of online replanning with a conventional Linac or MR-Linac

被引:32
作者
Chen, Guang-Pei [1 ]
Ahunbay, Ergun [1 ]
Li, X. Allen [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
关键词
quality assurance; online replanning; secondary MU calculation; MR-Linac; RADIOTHERAPY;
D O I
10.1118/1.4943795
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop an integrated quality assurance (QA) software tool for online replanning capable of efficiently and automatically checking radiation treatment (RT) planning parameters and gross plan quality, verifying treatment plan data transfer from treatment planning system (TPS) to record and verify (R&V) system, performing a secondary monitor unit (MU) calculation with or without a presence of a magnetic field from MR-Linac, and validating the delivery record consistency with the plan. Methods: The software tool, named ArtQA, was developed to obtain and compare plan and treatment parameters from both the TPS and the R&V system database. The TPS data are accessed via direct file reading and the R& V data are retrieved via open database connectivity and structured query language. Plan quality is evaluated with both the logical consistency of planning parameters and the achieved dose-volume histograms. Beams in between the TPS and R& V system are matched based on geometry configurations. To consider the effect of a 1.5 T transverse magnetic field from MR-Linac in the secondary MU calculation, a method based on modified Clarkson integration algorithm was developed and tested for a series of clinical situations. Results: ArtQA has been used in their clinic and can quickly detect inconsistencies and deviations in the entire RT planning process. With the use of the ArtQA tool, the efficiency for plan check including plan quality, data transfer, and delivery check can be improved by at least 60%. The newly developed independent MU calculation tool for MR-Linac reduces the difference between the plan and calculated MUs by 10%. Conclusions: The software tool ArtQA can be used to perform a comprehensive QA check from planning to delivery with conventional Linac or MR-Linac and is an essential tool for online replanning where the QA check needs to be performed rapidly. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:1713 / 1719
页数:7
相关论文
共 16 条
[1]   An on-line replanning scheme for interfractional variations [J].
Ahunbay, Ergun E. ;
Peng, Cheng ;
Chen, Guang-Pei ;
Narayanan, Sreeram ;
Yu, Cedric ;
Lawton, Colleen ;
Li, X. Allen .
MEDICAL PHYSICS, 2008, 35 (08) :3607-3615
[2]  
[Anonymous], 2014, LEDRTP0024 IMPAC MED
[3]  
Bogdanich W., 2011, New York Times
[4]   Quality Control Quantification (QCQ): A Tool to Measure the Value of Quality Control Checks in Radiation Oncology [J].
Ford, Eric C. ;
Terezakis, Stephanie ;
Souranis, Annette ;
Harris, Kendra ;
Gay, Hiram ;
Mutic, Sasa .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03) :E263-E269
[5]   Investigation of variance reduction techniques for Monte Carlo photon dose calculation using XVMC [J].
Kawrakow, I ;
Fippel, M .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08) :2163-2183
[6]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[7]   A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assurance [J].
Kung, JH ;
Chen, GTY ;
Kuchnir, FK .
MEDICAL PHYSICS, 2000, 27 (10) :2226-2230
[8]   MRI/linac integration [J].
Lagendijk, Jan J. W. ;
Raaymakers, Bas W. ;
Raaijmakers, Alexander J. E. ;
Overweg, Johan ;
Brown, Kevin J. ;
Kerkhof, Ellen M. ;
van der Put, Richard W. ;
Hardemark, Bjorn ;
van Vutpen, Marco ;
van der Heide, Uutke A. .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) :25-29
[9]   Software tool for physics chart checks [J].
Li, H. Harold ;
Wu, Yu ;
Yang, Deshan ;
Mutic, Sasa .
PRACTICAL RADIATION ONCOLOGY, 2014, 4 (06) :E217-E225
[10]   The Dawn of a New Era: First Ever MR-IGRT Treatments - Initial Experiences and Future Implications [J].
Mutic, S. ;
Parikh, P. J. ;
Bradley, J. D. ;
Hallahan, D. E. ;
Hu, Y. ;
Kashani, R. ;
Kawrakow, I. ;
Li, H. ;
Michalski, J. M. ;
Olsen, J. R. ;
Robinson, C. G. ;
Rodriguez, V. ;
Santanam, L. ;
Tanderup, K. ;
Victoria, J. ;
Wooten, H. O. ;
Yang, D. ;
Zoberi, I. ;
Green, O. L. ;
Dempsey, J. F. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 :S94-S94