Distortion of Rabi oscillations in a compact cold-atom clock

被引:2
作者
Ouyang, Xinchuan [1 ,2 ,3 ]
Yang, Bowen [1 ,2 ,3 ]
Hu, Qingqing [4 ]
Qi, Hanghang [1 ,2 ,3 ]
Xiao, Ling [1 ,2 ]
Wan, Jinyin [1 ,2 ]
Cheng, Huadong [1 ,2 ,3 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Opt & Fine Mech, Key Lab Quantum Opt, Shanghai 201800, Peoples R China
[2] Chinese Acad Sci, Ctr Cold Atom Phys, Shanghai Inst Opt & Fine Mech, Shanghai 201800, Peoples R China
[3] Univ Chinese Acad Sci, Ctr Mat Sci & Optoelect Engn, Beijing 100049, Peoples R China
[4] Natl Innovat Inst Def Technol, Adv Interdisciplinary Technol Res Ctr, Beijing 100010, Peoples R China
基金
中国国家自然科学基金;
关键词
QUANTIZATION;
D O I
10.1103/PhysRevA.103.043118
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report on experimental measurements and theoretical calculations of Rabi oscillation distortion during rubidium clock transition in a compact cold-atom clock. The distortion of the Rabi oscillations of the clock transition caused by the magnetically sensitive transitions between magnetic sublevels is related to the duration and amplitude of the microwave pulse, the density distribution of cold atoms, and the biased magnetic field inside the microwave cavity. The theoretical calculations agreed well with the experimental results. The Rabi oscillation measurements are beneficial for determining the optimal parameters of a cold-atom clock and identifying the uniformity of the microwave field inside the cavity. This measurement can be extended to the atomic fountain clock and the vapor-cell atomic clock.
引用
收藏
页数:6
相关论文
共 30 条
[1]   Quantum Rabi Oscillations in Coherent and in Mesoscopic Cat Field States [J].
Assemat, F. ;
Grosso, D. ;
Signoles, A. ;
Facon, A. ;
Dotsenko, I ;
Haroche, S. ;
Raimond, J. M. ;
Brune, M. ;
Gleyzes, S. .
PHYSICAL REVIEW LETTERS, 2019, 123 (14)
[2]   Rabi spectroscopy and excitation inhomogeneity in a one-dimensional optical lattice clock [J].
Blatt, S. ;
Thomsen, J. W. ;
Campbell, G. K. ;
Ludlow, A. D. ;
Swallows, M. D. ;
Martin, M. J. ;
Boyd, M. M. ;
Ye, J. .
PHYSICAL REVIEW A, 2009, 80 (05)
[3]   Quantum rabi oscillation: A direct test of field quantization in a cavity [J].
Brune, M ;
Schmidt-Kaler, F ;
Maali, A ;
Dreyer, J ;
Hagley, E ;
Raimond, JM ;
Haroche, S .
PHYSICAL REVIEW LETTERS, 1996, 76 (11) :1800-1803
[4]   RAMSEY RESONANCE IN A ZACHARIAS FOUNTAIN [J].
CLAIRON, A ;
SALOMON, C ;
GUELLATI, S ;
PHILLIPS, WD .
EUROPHYSICS LETTERS, 1991, 16 (02) :165-170
[5]   HORACE: A compact cold atom clock for Galileo [J].
Esnault, F. X. ;
Rossetto, N. ;
Holleville, D. ;
Delporte, J. ;
Dimarcq, N. .
ADVANCES IN SPACE RESEARCH, 2011, 47 (05) :854-858
[6]   High-stability compact atomic clock based on isotropic laser cooling [J].
Esnault, Francois-Xavier ;
Holleville, David ;
Rossetto, Nicolas ;
Guerandel, Stephane ;
Dimarcq, Noel .
PHYSICAL REVIEW A, 2010, 82 (03)
[7]   Three-dimensional cooling of cesium atoms in a reflecting copper cylinder [J].
Guillot, E ;
Pottie, PE ;
Dimarcq, M .
OPTICS LETTERS, 2001, 26 (21) :1639-1641
[8]   First accuracy evaluation of NIST-F2 [J].
Heavner, Thomas P. ;
Donley, Elizabeth A. ;
Levi, Filippo ;
Costanzo, Giovanni ;
Parker, Thomas E. ;
Shirley, Jon H. ;
Ashby, Neil ;
Barlow, Stephan ;
Jefferts, S. R. .
METROLOGIA, 2014, 51 (03) :174-182
[9]   Exceptional points and the ring laser gyroscope [J].
Horstman, Luke ;
Hsu, Ning ;
Hendrie, James ;
Smith, David ;
Diels, Jean-Claude .
PHOTONICS RESEARCH, 2020, 8 (03) :252-256
[10]   Power dependence of distributed cavity phase-induced frequency biases in atomic fountain frequency standards [J].
Jefferts, SR ;
Shirley, JH ;
Ashby, N ;
Burt, EA ;
Dick, GJ .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (12) :2314-2321