Profile monitor SEMs for the NuMI beam at Fermilab

被引:0
作者
Indurthy, D [1 ]
Kopp, S [1 ]
Proga, M [1 ]
Pavlovich, Z [1 ]
机构
[1] Univ Texas, Dept Phys, Austin, TX 78712 USA
来源
BEAM INSTRUMENTATION WORKSHOP 2004 | 2004年 / 732卷
关键词
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Neutrinos at the Main Injector (NuMI) project will extract 120-GeV protons from the FNAL Main Injector in 8.56-musec spills of 4 x 10(13) protons every 1.9 sec. We have designed secondary emission monitor (SEM) detectors to measure beam profile and halo along the proton beam transport line. The SEMs are Ti foils 5 mum in thickness segmented in either 1-mm or 0.5-mm pitch strips, resulting in beam loss similar to5 x 10(-6). We discuss aspects of the mechanical design, calculations of expected beam heating, and results of a beam test at the 8-GeV transport line to MiniBoone at FNAL.