Real-time intraocular pressure measurement in standard and microcoaxial phacoemulsification

被引:26
作者
Kreutzer, Thomas C. [1 ]
Al Saeidi, Rashid [1 ]
Kampik, Anselm [1 ]
Grueterich, Martin [1 ]
机构
[1] Univ Munich, Univ Eye Hosp, D-80336 Munich, Germany
关键词
D O I
10.1016/j.jcrs.2009.07.035
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate intraocular pressure (IOP) in the vitreous cavity during various stages of cataract surgery. SETTING: University Eye Hospital, Ludwig-Maximilians University, Munich, Germany. METHODS: In consecutive eyes having combined phacoemulsification, intraocular lens implantation, and pars plana vitrectomy, IOP was monitored in real time through a 25-gauge pars plana cannula connected to an external pressure transducer. Surgery was performed by standard clear corneal phacoemulsification with a 2.5 mm incision and a Mega-Tip (1.26 mm aperture) (Group 1) or by microcoaxial phacoemulsification with a CMP-Tip (0.80 mm aperture) (Group 2). RESULTS: The 2 groups had 5 eyes each. The mean IOP in Group 1 and in Group 2, respectively, was 15.9 mm Hg +/- 9.5 (SD) and 17.0 +/- 13.5 mm Hg preoperatively (P = .442), 40.1 +/- 12.7 mm Hg and 36.5 +/- 17.2 mm Hg during lens removal (P<.001), 17.6 +/- 14.2 mm Hg and 22.6 +/- 8.6 mm Hg during irrigation/aspiration (P<.001), 13.3 +/- 13.2 mm Hg and 16.3 +/- 13.1 mm Hg during IOL implantation (P = .005), and 22.9 +/- 7.0 mm Hg and 21.5 +/- 10.0 mm Hg after IOL implantation through the end of surgery (P = .329). CONCLUSIONS: Although the IOP levels were significantly lower than those in previous studies, both phacoemulsification techniques had safe IOP profiles during various steps of surgery. Real-time IOP monitoring may prevent the surgeon from inducing excessive IOP elevation during intraocular manipulation.
引用
收藏
页码:53 / 57
页数:5
相关论文
共 7 条
[1]  
BLUMENTHAL M, 1992, OPHTHALMIC SURG LAS, V23, P132
[2]   Comparison of intraocular pressure profiles during cataract surgery and by phacoemulsification and extracapsular cataract extraction [J].
Grinbaum, A ;
Blumenthal, M ;
Assia, E .
OPHTHALMIC SURGERY LASERS & IMAGING, 2003, 34 (03) :182-186
[3]   Intraocular pressure during phacoemulsification [J].
Khng, C ;
Packer, M ;
Fine, H ;
Hoffman, RS ;
Moreira, FB .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (02) :301-308
[4]   More than 4 years' experience with electronic intraocular needle tonometry [J].
Marx, W ;
Madjlessi, F ;
Reinhard, T ;
Althaus, C ;
Sundmacher, R .
OPHTHALMOLOGE, 1999, 96 (08) :498-502
[5]  
MOORHEAD LC, 1986, ARCH OPHTHALMOL-CHIC, V104, P269
[6]  
PAREL JM, 1987, OPHTHALMIC SURG LAS, V18, P371
[7]   EVALUATION OF INTRAOCULAR-PRESSURE FLUCTUATIONS WITH DIFFERING PHACOEMULSIFICATION APPROACHES [J].
WILBRANDT, HR ;
WILBRANDT, TH .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1993, 19 (02) :223-231