Skill dependence of radiation exposure for the orthopaedic surgeon during interlocking nailing of long-bone shaft fractures: a clinical study

被引:66
作者
Blattert, TR
Fill, UA
Kunz, E
Panzer, W
Weckbach, A
Regulla, DF
机构
[1] Wuerzburg Univ Hosp, D-97080 Wurzburg, Germany
[2] GSF Natl Res Ctr Environm & Hlth, Inst Radiat Protect, D-85764 Neuherberg, Germany
[3] Dist Hosp, D-76437 Rastatt, Germany
关键词
radiation; fluoroscopy; thermoluminescent dosimetry; intramedullary fracture fixation; clinical competence;
D O I
10.1007/s00402-004-0743-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction The objective of this clinical trial was to determine whether there is a skill dependence for the total amount of radiation exposure to orthopaedic surgeons caused by fluoroscopy during intramedullary fracture fixation. Materials and methods Surgical teams were assigned to either the 'Senior group' or the 'Junior group' according to their professional qualification and clinical appointment. Twenty-two long-bone shaft fractures were stabilized with intramedullary nails. The radiation exposure was measured at different body locations including fingers, trunk and head by means of thermoluminescent LiF: Mg, Cu, P detectors. The total time of fluoroscopy was registered for each operation. Results Mean time of fluoroscopy per operation was 4.43 min for the 'Senior group' and 6.95 min for the 'Junior group'. The surgeons' hands were exposed to markedly higher doses ( range 0 - 2.88 mSv 'Senior group'; 0 - 11.94 mSv 'Junior group') than their trunk and head ( range 0 - 0.27 mSv 'Senior group'; 0 - 0.38 mSv 'Junior group'). After analysis of variance, differences between both groups proved to be statistically significant for all fingers measured ( p less than or equal to 0.02) and for the total time of fluoroscopy ( p = 0.019). Conclusions Generally, the hands are at higher risk than are the trunk and the head, and this finding is independent of surgical skills. However, an additional hazard is created for the less experienced surgeon by a highly varying and poorly predictable exposure of the hands and time needed for fluoroscopy. Thus, the use of radiation is more consistent and standardized with a skilled surgeon.
引用
收藏
页码:659 / 664
页数:6
相关论文
共 21 条
[1]  
[Anonymous], 1992, P ACR FDA WORKSH FLU
[2]  
BARRY TP, 1984, CLIN ORTHOP RELAT R, P160
[3]  
Dosch J C, 1983, Hefte Unfallheilkd, V161, P36
[4]   Measurements with Chinese LiF:Mg,Cu,P (GR 200) TL detectors using a volume-type oven [J].
Fill, UA ;
Regulla, DF .
APPLIED RADIATION AND ISOTOPES, 1997, 48 (01) :133-134
[5]   Use of Chinese LiF:Mg,Cu,P (GR-200) TL detectors at therapy-level absorbed doses [J].
Fill, UA ;
Regulla, DF .
APPLIED RADIATION AND ISOTOPES, 1998, 49 (07) :791-793
[6]   Advanced LiF technology for the assessment of patient exposure in diagnostic radiology [J].
Fill, UA ;
Regulla, DF .
RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) :225-229
[7]  
FRANKENBERG D, 2002, RADIAT RES, V157, P208
[8]   Monitoring of trauma surgeons' exposure to radiation.: Measurements of hand, eye and thyroid with highly sensitive thermoluminescence detectors [J].
Fuchs, M ;
Modler, H ;
Schmid, A ;
Dumont, C ;
Stürmer, KM .
UNFALLCHIRURG, 1999, 102 (05) :371-376
[9]   IRRADIATION OF THE SURGEON DURING PINNING OF FEMORAL FRACTURES [J].
GIACHINO, AA ;
CHENG, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1980, 62 (02) :227-229
[10]   RADIATION EXPOSURE TO THE HANDS OF ORTHOPEDIC SURGEONS DURING PROCEDURES UNDER FLUOROSCOPIC X-RAY CONTROL [J].
GOLDSTONE, KE ;
WRIGHT, IH ;
COHEN, B .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) :899-901