Effectiveness of neonatal pulse oximetry screening for detection of critical congenital heart disease in daily clinical routine-results from a prospective multicenter study

被引:169
|
作者
Riede, Frank Thomas [1 ]
Woerner, Cornelia [1 ]
Daehnert, Ingo [1 ]
Moeckel, Andreas [2 ]
Kostelka, Martin [1 ]
Schneider, Peter [1 ]
机构
[1] Univ Leipzig, Ctr Heart, D-04289 Leipzig, Germany
[2] Helios Klinikum Borna, Dept Pediat, D-04552 Borna, Germany
关键词
Prenatal ultrasound; Physical examination; Pulse oximetry screening; Newborn; Critical congenital heart disease; ASYMPTOMATIC NEWBORNS; OXYGEN-SATURATION; DEFECTS; DIAGNOSIS; MALFORMATIONS; DISCHARGE; FAILURE; INFANTS; AGE;
D O I
10.1007/s00431-010-1160-4
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Pulse oximetry screening (POS) has been proposed as an effective, noninvasive, inexpensive tool allowing earlier diagnosis of critical congenital heart disease (cCHD). Our aim was to test the hypothesis that POS can reduce the diagnostic gap in cCHD in daily clinical routine in the setting of tertiary, secondary and primary care centres. We conducted a prospective multicenter trial in Saxony, Germany. POS was performed in healthy term and post-term newborns at the age of 24-72 h. If an oxygen saturation (SpO(2)) of a parts per thousand currency sign95% was measured on lower extremities and confirmed after 1 h, complete clinical examination and echocardiography were performed. POS was defined as false-negative when a diagnosis of cCHD was made after POS in the participating hospitals/at our centre. From July 2006-June 2008, 42,240 newborns from 34 institutions have been included. Seventy-two children were excluded due to prenatal diagnosis (n = 54) or clinical signs of cCHD (n = 18) before POS. Seven hundred ninety-five newborns did not receive POS, mainly due to early discharge after birth (n = 727; 91%). In 41,445 newborns, POS was performed. POS was true positive in 14, false positive in 40, true negative in 41,384 and false negative in four children (three had been excluded for violation of study protocol). Sensitivity, specificity, positive and negative predictive value were 77.78%, 99.90%, 25.93% and 99.99%, respectively. With POS as an adjunct to prenatal diagnosis, physical examination and clinical observation, the percentage of newborns with late diagnosis of cCHD was 4.4%. POS can substantially reduce the postnatal diagnostic gap in cCHD, and false-positive results leading to unnecessary examinations of healthy newborns are rare. POS should be implemented in routine postnatal care.
引用
收藏
页码:975 / 981
页数:7
相关论文
共 50 条
  • [1] Effectiveness of neonatal pulse oximetry screening for detection of critical congenital heart disease in daily clinical routine—results from a prospective multicenter study
    Frank Thomas Riede
    Cornelia Wörner
    Ingo Dähnert
    Andreas Möckel
    Martin Kostelka
    Peter Schneider
    European Journal of Pediatrics, 2010, 169 : 975 - 981
  • [2] Effectiveness of pulse oximetry screening for detection of critical congenital heart disease in daily clinical routine-results from a prospective multicenter study
    Riede, Frank Thomas
    Schneider, Peter
    EUROPEAN JOURNAL OF PEDIATRICS, 2010, 169 (12) : 1571 - 1572
  • [3] Effectiveness of pulse oximetry screening for detection of critical congenital heart disease in daily clinical routine—results from a prospective multicenter study
    Frank Thomas Riede
    Peter Schneider
    European Journal of Pediatrics, 2010, 169 : 1571 - 1572
  • [4] Evaluation of Critical Congenital Heart Disease Screening Results with Pulse Oximetry
    Kara, Nursu
    Arman, Didem
    Gul, Adem
    Simsek, Turkan
    Ceylan, Ozben
    Comert, Serdar
    ISTANBUL MEDICAL JOURNAL, 2022, 23 (02): : 102 - 106
  • [5] Critical Congenital Heart Disease Screening Using Pulse Oximetry
    Frank, Lowell H.
    Bradshaw, Elizabeth
    Beekman, Robert
    Mahle, William T.
    Martin, Gerard R.
    JOURNAL OF PEDIATRICS, 2013, 162 (03) : 445 - 453
  • [6] Effectiveness of pulse-oximetry in addition to routine neonatal examination in detection of congenital heart disease in asymptomatic newborns
    Oakley, Juliette L.
    Soni, Naharmal B.
    Wilson, Dirk
    Sen, Siddhartha
    JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE, 2015, 28 (14) : 1736 - 1739
  • [7] Universal Pulse Oximetry Screening for Early Detection of Critical Congenital Heart Disease
    Kumar, Praveen
    CLINICAL MEDICINE INSIGHTS-PEDIATRICS, 2016, 10 : 35 - 41
  • [8] Pulse oximetry screening in newborns to enhance detection of critical congenital heart disease
    Narvey, Michael
    Wong, Kenny K.
    Fournier, Anne
    PAEDIATRICS & CHILD HEALTH, 2017, 22 (08) : 494 - 498
  • [9] Newborn Pulse Oximetry Screening to Detect Critical Congenital Heart Disease
    Studer, Matthew A.
    Smith, Ashley E.
    Lustik, Michael B.
    Carr, Michael R.
    JOURNAL OF PEDIATRICS, 2014, 164 (03) : 505 - +
  • [10] Neonatal pulse oximetry screening improves detecting of critical congenital heart disease
    Ma Xiao-jing
    Huang Guo-ying
    CHINESE MEDICAL JOURNAL, 2013, 126 (14) : 2736 - 2740