Computer-Aided Volumetry of Pulmonary Nodules Exhibiting Ground-Glass Opacity at MDCT

被引:82
作者
Oda, Seitaro [1 ]
Awai, Kazuo [1 ]
Murao, Kohei [2 ]
Ozawa, Akio [2 ]
Yanaga, Yumi [1 ]
Kawanaka, Koichi [1 ]
Yamashita, Yasuyuki [1 ]
机构
[1] Kumamoto Univ, Grad Sch Med Sci, Dept Diagnost Radiol, Kumamoto 8608556, Japan
[2] Fujitsu, Bio IT Business Dev Grp, Tokyo, Japan
关键词
computer-aided diagnosis; ground-glass opacity; high-resolution CT; pulmonary nodules; volumetry; THIN-SECTION CT; DOSE HELICAL CT; HISTOLOGIC CHARACTERISTICS; AUTOMATED VOLUMETRY; LUNG NODULES; GROWTH-RATE; ADENOCARCINOMA; SEGMENTATION; ACCURACY; REPRODUCIBILITY;
D O I
10.2214/AJR.09.2583
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to investigate the accuracy and reproducibility of results acquired with computer-aided volumetry software during MDCT of pulmonary nodules exhibiting ground-glass opacity. MATERIALS AND METHODS. To evaluate the accuracy of computer-aided volumetry software, we performed thin-section helical CT of a chest phantom that included simulated 3-, 5-, 8-, 10-, and 12-mm-diameter ground-glass opacity nodules with attenuation of -800, -630, and -450 HU. Three radiologists measured the volume of the nodules and calculated the relative volume measurement error, which was defined as follows: ( measured nodule volume minus assumed nodule volume divided by assumed nodule volume) x 100. Two radiologists performed two independent measurements of 59 nodules in humans. Intraobserver and inter-observer agreement was evaluated with Bland-Altman methods. RESULTS. The relative volume measurement error for simulated ground-glass opacity nodules measuring 3 mm ranged from 51.1% to 85.2% and for nodules measuring 5 mm or more in diameter ranged from -4.1% to 7.1%. In the clinical study, for intraobserver agreement, the 95% limits of agreement were -14.9% and -13.7% and -16.6% to 15.7% for observers A and B. For interobserver agreement, these values were -16.3% to 23.7% for nodules 8 mm in diameter or larger. CONCLUSION. With computer-aided volumetry of ground-glass opacity nodules, the relative volume measurement error was small for nodules 5 mm in diameter or larger. Intraobserver and interobserver agreement was relatively high for nodules 8 mm in diameter or larger.
引用
收藏
页码:398 / 406
页数:9
相关论文
共 32 条
[1]   Evolution of peripheral lung adenocarcinomas: CT findings correlated with histology and tumor doubling time [J].
Aoki, T ;
Nakata, H ;
Watanabe, H ;
Nakamura, K ;
Kasai, T ;
Hashimoto, H ;
Yasumoto, K ;
Kido, M .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (03) :763-768
[2]   Peripheral lung adenocarcinoma: Correlation of thin-section CT findings with histologic prognostic factors and survival [J].
Aoki, T ;
Tomoda, Y ;
Watanabe, H ;
Nakata, H ;
Kasai, T ;
Hashimoto, H ;
Kodate, M ;
Osaki, T ;
Yasumoto, K .
RADIOLOGY, 2001, 220 (03) :803-809
[3]   Pulmonary nodules: Estimation of malignancy at thin-section helical CT - Effect of computer-aided diagnosis on performance of radiologists [J].
Awai, K ;
Murao, K ;
Ozawa, A ;
Nakayama, Y ;
Nakaura, T ;
Liu, D ;
Kawanaka, K ;
Funama, Y ;
Morishita, S ;
Yamashita, Y .
RADIOLOGY, 2006, 239 (01) :276-284
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Accuracy of automated volumetry of pulmonary nodules across different multislice CT scanners [J].
Das, Marco ;
Ley-Zaporozhan, Julia ;
Gietema, H. A. ;
Czech, Andre ;
Muehlenbruch, Georg ;
Mahnken, Andreas H. ;
Katoh, Markus ;
Bakai, Annemarie ;
Salganicoff, Marcos ;
Diederich, Stefan ;
Prokop, Mathias ;
Kauczor, Hans-Ulrich ;
Guenther, Rolf W. ;
Wildberger, Joachim E. .
EUROPEAN RADIOLOGY, 2007, 17 (08) :1979-1984
[6]   Automated volumetry of solid pulmonary nodules in a phantom -: Accuracy across different CT scanner technologies [J].
Das, Marco ;
Muehlenbruch, Georg ;
Katoh, Markus ;
Bakai, Annemarie ;
Salganicoff, Marcos ;
Stanzel, Sven ;
Mahnken, Andreas H. ;
Guenther, Rolf W. ;
Wildberger, Joachim E. .
INVESTIGATIVE RADIOLOGY, 2007, 42 (05) :297-302
[7]   A comparison of six software packages for evaluation of solid lung nodules using semi-automated volumetry: What is the minimum increase in size to detect growth in repeated CT examinations [J].
de Hoop, Bartjan ;
Gietema, Hester ;
van Ginneken, Bram ;
Zanen, Pieter ;
Groenewegen, Gerard ;
Prokop, Mathias .
EUROPEAN RADIOLOGY, 2009, 19 (04) :800-808
[8]  
DOUGHERTY E, 1994, DIGITAL IMAGE PROCES, P77
[9]   Pulmonary nodules: Interscan variability of semiautomated volume measurements with multisection CT-influence of inspiration level, nodule size, and segmentation performance [J].
Gietema, Hester A. ;
Schaefer-Prokop, Cornelia M. ;
Mali, Willem P. T. M. ;
Groenewegen, Gerard ;
Prokop, Mathias .
RADIOLOGY, 2007, 245 (03) :888-894
[10]   Volumetric measurement of synthetic lung nodules with multi-detector row CT: Effect of various image reconstruction parameters and segmentation thresholds on measurement accuracy [J].
Goo, JM ;
Tongdee, T ;
Tongdee, R ;
Yeo, K ;
Hildebolt, CF ;
Bae, KT .
RADIOLOGY, 2005, 235 (03) :850-856