Effect of source blur on digital breast tomosynthesis reconstruction

被引:10
|
作者
Zheng, Jiabei [1 ,2 ,3 ]
Fessler, Jeffrey A. [1 ,2 ]
Chan, Heang-Ping [1 ]
机构
[1] Univ Michigan, Dept Radiol, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Elect & Comp Engn, Ann Arbor, MI 48109 USA
[3] Intel Corp, 5000 W Chandler Blvd, Chandler, AZ 85226 USA
基金
美国国家卫生研究院;
关键词
digital breast tomosynthesis; geometric unsharpness; image reconstruction; spatial resolution; x-ray focal spot blur; FOCAL SPOT SIZE; X-RAY TUBES; DETECTOR BLUR; PINHOLE;
D O I
10.1002/mp.13801
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Most digital breast tomosynthesis (DBT) reconstruction methods neglect the blurring of the projection views caused by the finite size or motion of the x-ray focal spot. This paper studies the effect of source blur on the spatial resolution of reconstructed DBT using analytical calculation and simulation, and compares the influence of source blur over a range of blurred source sizes. Methods Mathematically derived formulas describe the point spread function (PSF) of source blur on the detector plane as a function of the spatial locations of the finite-sized source and the object. By using the available technical parameters of some clinical DBT systems, we estimated the effective source sizes over a range of exposure time and DBT scan geometries. We used the CatSim simulation tool (GE Global Research, NY) to generate digital phantoms containing line pairs and beads at different locations and imaged with sources of four different sizes covering the range of potential source blur. By analyzing the relative contrasts of the test objects in the reconstructed images, we studied the effect of the source blur on the spatial resolution of DBT. Furthermore, we simulated a detector that rotated in synchrony with the source about the rotation center and calculated the spatial distribution of the blurring distance in the imaged volume to estimate its influence on source blur. Results Calculations demonstrate that the PSF is highly shift-variant, making it challenging to accurately implement during reconstruction. The results of the simulated phantoms demonstrated that a typical finite-sized focal spot (similar to 0.3 mm) will not affect the reconstructed image resolution if the x-ray tube is stationary during data acquisition. If the x-ray tube moves during exposure, the extra blur due to the source motion may degrade image resolution, depending on the effective size of the source along the direction of the motion. A detector that rotates in synchrony with the source does not reduce the influence of source blur substantially. Conclusions This study demonstrates that the extra source blur due to the motion of the x-ray tube during image acquisition substantially degrades the reconstructed image resolution. This effect cannot be alleviated by rotating the detector in synchrony with the source. The simulation results suggest that there are potential benefits of modeling the source blur in image reconstruction for DBT systems using continuous-motion acquisition mode.
引用
收藏
页码:5572 / 5592
页数:21
相关论文
共 50 条
  • [1] Detector Blur and Correlated Noise Modeling for Digital Breast Tomosynthesis Reconstruction
    Zheng, Jiabei
    Fessler, Jeffrey A.
    Chan, Heang-Ping
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (01) : 116 - 127
  • [2] Effects of detector blur and correlated noise on digital breast tomosynthesis reconstruction
    Zheng, Jiabei
    Fessler, Jeffrey A.
    Chan, Heang-Ping
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [3] Open-Source Reconstruction Toolbox for Digital Breast Tomosynthesis
    Vimieiro, Rodrigo B.
    Borges, Lucas R.
    Vieira, Marcelo A. C.
    XXVI BRAZILIAN CONGRESS ON BIOMEDICAL ENGINEERING, CBEB 2018, VOL. 2, 2019, 70 (02): : 349 - 354
  • [4] X-ray source motion blur modeling and deblurring with generative diffusion for digital breast tomosynthesis
    Gao, Mingjie
    Fessler, Jeffrey A.
    Chan, Heang-Ping
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (11):
  • [5] Deep Convolutional Neural Network Regularized Digital Breast Tomosynthesis Reconstruction with Detector Blur and Correlated Noise Modeling
    Gao, Mingjie
    Fessler, Jeffrey A.
    Chan, Heang-Ping
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [6] Local dynamic reconstruction in digital breast tomosynthesis
    Barbieri, Matteo
    Jailin, Clement
    Vancamberg, Laurence
    Roux, Stephane
    17TH INTERNATIONAL WORKSHOP ON BREAST IMAGING, IWBI 2024, 2024, 13174
  • [7] The reconstruction of microcalcification clusters in digital breast tomosynthesis
    Ho, Candy P. S.
    Tromans, Chris E.
    Schnabel, Julia A.
    Brady, Michael
    MEDICAL IMAGING 2010: COMPUTER - AIDED DIAGNOSIS, 2010, 7624
  • [8] Combined Reconstruction and Registration of Digital Breast Tomosynthesis
    Yang, Guang
    Hipwell, John H.
    Clarkson, Matthew J.
    Tanner, Christine
    Mertzanidou, Thomy
    Gunn, Spencer
    Ourselin, Sebastien
    Hawkes, David J.
    Arridge, Simon R.
    DIGITAL MAMMOGRAPHY, 2010, 6136 : 760 - +
  • [9] A Spectral Reconstruction Method for Digital Breast Tomosynthesis
    Feng, S. S. J.
    Nagy, J.
    Chung, J.
    Mejia-Bustamante, V.
    Sechopoulos, I.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction
    Erhard, Klaus
    Grass, Michael
    Hitziger, Sebastian
    Iske, Armin
    Nielsen, Tim
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313