Enhancements to commissioning techniques and quality assurance of brachytherapy treatment planning systems that use model-based dose calculation algorithms

被引:54
|
作者
Rivard, Mark J. [1 ]
Beaulieu, Luc [4 ,5 ,6 ]
Mourtada, Firas [2 ,3 ]
机构
[1] Tufts Univ, Sch Med, Dept Radiat Oncol, Boston, MA 02111 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Expt Diagnost Imaging, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[4] Univ Laval, CHU Quebec, Dept Radiooncol, Quebec City, PQ G1R 2J6, Canada
[5] Univ Laval, CHU Quebec, Ctr Rech Cancerol, Quebec City, PQ G1R 2J6, Canada
[6] Univ Laval, Dept Phys Genie Phys & Opt, Quebec City, PQ G1R 2J6, Canada
关键词
brachytherapy; dose calculation algorithm; treatment planning system; QA; SHIELDED CERVICAL APPLICATORS; RATE ENDORECTAL BRACHYTHERAPY; PERMANENT PROSTATE IMPLANTS; FINITE PATIENT DIMENSIONS; EYE PLAQUE BRACHYTHERAPY; MONTE-CARLO CALCULATIONS; RADIATION-THERAPY; INTRACAVITARY BRACHYTHERAPY; INTERSTITIAL BRACHYTHERAPY; BREAST BRACHYTHERAPY;
D O I
10.1118/1.3429131
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The current standard for brachytherapy dose calculations is based on the AAPM TG-43 formalism. Simplifications used in the TG-43 formalism have been challenged by many publications over the past decade. With the continuous increase in computing power, approaches based on fundamental physics processes or physics models such as the linear-Boltzmann transport equation are now applicable in a clinical setting. Thus, model-based dose calculation algorithms (MBDCAs) have been introduced to address TG-43 limitations for brachytherapy. The MBDCA approach results in a paradigm shift, which will require a concerted effort to integrate them properly into the radiation therapy community. MBDCA will improve treatment planning relative to the implementation of the traditional TG-43 formalism by accounting for individualized, patient-specific radiation scatter conditions, and the radiological effect of material heterogeneities differing from water. A snapshot of the current status of MBDCA and AAPM Task Group reports related to the subject of QA recommendations for brachytherapy treatment planning is presented. Some simplified Monte Carlo simulation results are also presented to delineate the effects MBDCA are called to account for and facilitate the discussion on suggestions for (i) new QA standards to augment current societal recommendations, (ii) consideration of dose specification such as dose to medium in medium, collisional kerma to medium in medium, or collisional kerma to water in medium, and (iii) infrastructure needed to uniformly introduce these new algorithms. Suggestions in this Vision 20/20 article may serve as a basis for developing future standards to be recommended by professional societies such as the AAPM, ESTRO, and ABS toward providing consistent clinical implementation throughout the brachytherapy community and rigorous quality management of MBDCA-based treatment planning systems. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3429131]
引用
收藏
页码:2645 / 2658
页数:14
相关论文
共 31 条
  • [21] Dosimetric Impact of Air Pockets in the Vaginal Cuff Brachytherapy Using Model-based Dose Calculation Algorithm
    Puriparthi, Lakshmi Venkataramana
    Talluri, Anil Kumar
    Akkineni, Naga Prasanthi
    Bajwa, Harjot Kaur
    Tumu, Venkatappa Rao
    Sresty, N. V. N. Madhusudhana
    Alluri, Krishnam Raju
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (04) : 373 - 377
  • [22] Beam modeling and beam model commissioning for Monte Carlo dose calculation-based radiation therapy treatment planning: Report of AAPM Task Group 157
    Ma, Chang Ming Charlie
    Chetty, Indrin J.
    Deng, Jun
    Faddegon, Bruce
    Jiang, Steve B.
    Li, Jinsheng
    Seuntjens, Jan
    Siebers, Jeffrey V.
    Traneus, Erik
    MEDICAL PHYSICS, 2020, 47 (01) : E1 - E18
  • [23] Moving GPU-OpenCL-based Monte Carlo dose calculation toward clinical use: Automatic beam commissioning and source sampling for treatment plan dose calculation
    Tian, Zhen
    Li, Yongbao
    Hassan-Rezaeian, Nima
    Jiang, Steve B.
    Jia, Xun
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02): : 69 - 84
  • [24] Effect of a lead block on alveolar bone protection in image-guided high-dose-rate interstitial brachytherapy for tongue cancer: using model-based dose calculation algorithms to correct for inhomogeneity
    Akiyama, Hironori
    Yoshida, Ken
    Takenaka, Tadashi
    Kotsuma, Tadayuki
    Masui, Koji
    Monzen, Hajime
    Sumida, Iori
    Tsujimoto, Yutaka
    Miyao, Mamoru
    Okumura, Hiroki
    Shimbo, Taiju
    Takegawa, Hideki
    Murakami, Naoya
    Inaba, Koji
    Kashihara, Tairo
    Takacsi-Nagy, Zoltan
    Tselis, Nikolaos
    Yamazaki, Hideya
    Tanaka, Eiichi
    Nihei, Keiji
    Ariji, Yoshiko
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2022, 14 (01) : 87 - 95
  • [25] Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation
    Beaulieu, Luc
    Tedgren, Asa Carlsson
    Carrier, Jean-Francois
    Davis, Stephen D.
    Mourtada, Firas
    Rivard, Mark J.
    Thomson, Rowan M.
    Verhaegen, Frank
    Wareing, Todd A.
    Williamson, Jeffrey F.
    MEDICAL PHYSICS, 2012, 39 (10) : 6208 - 6236
  • [26] Comparison of radiography- and computed tomography-based treatment planning in cervix cancer in brachytherapy with specific attention to some quality assurance aspects
    Fellner, C
    Pötter, R
    Knocke, TH
    Wambersie, A
    RADIOTHERAPY AND ONCOLOGY, 2001, 58 (01) : 53 - 62
  • [27] Implement a knowledge-based automated dose volume histogram prediction module in Pinnacle3 treatment planning system for plan quality assurance and guidance
    Xu, Hao
    Lu, Jiayu
    Wang, Jiazhou
    Fan, Jiawei
    Hu, Weigang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (08): : 134 - 140
  • [28] A comparison of TG-43 and TG-1 86 dose calculation algorithms for treatment planning of intra-cavitary brachytherapy using tandem and ovoid applicator
    Shajid, Syed Mohamed
    Aggarwal, Lalit Mohan
    Mourya, Ankur
    Choudhary, Sunil
    Singh, Ankita
    Oommen, Saji
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2024, 16 (04) : 289 - 296
  • [29] Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer
    Kanani, Abolfazl
    Owrangi, Amir
    Yazdi, Mehran
    Fatemi-Ardekani, Ali
    Mosleh-Shirazi, Mohammad Amin
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2023, 15 (01) : 57 - 68
  • [30] Characterization of Dosimetric Differences in Strut-Adjusted Volume Implant Treatment Plans Calculated With TG-43 Formalism and a Model-Based Dose Calculation Algorithm
    Mazur, Thomas R.
    Hao, Yao
    Garcia-Ramirez, Jose
    Altman, Michael B.
    Li, H. Harold
    Thomas, Maria A.
    Zoberi, Imran
    Zoberi, Jacqueline E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 110 (04): : 1200 - 1209