Primary Fixation and Cyclic Performance of Single-Stitch All-Inside and Inside-Out Meniscal Devices for Repairing Vertical Longitudinal Meniscal Tears

被引:13
作者
Bachmaier, Samuel [1 ,2 ]
Krych, Aaron J. [1 ,3 ,4 ]
Smith, Patrick A. [1 ,5 ]
Herbort, Mirco [1 ,6 ]
Ritter, Daniel [1 ,2 ]
LaPrade, Robert F. [1 ,7 ]
Wijdicks, Coen A. [1 ,2 ]
机构
[1] Arthrex Dept Orthoped Res, Munich, Germany
[2] Arthrex, Dept Orthoped Res, Munich, Germany
[3] Mayo Clin, Dept Orthoped Surg, Rochester, MN USA
[4] Mayo Clin, Ctr Sports Med, Rochester, MN USA
[5] Columbia Orthopaed Grp, Columbia, MO USA
[6] OCM Clin, Munich, Germany
[7] Twin Cities Orthoped, Edina, MN USA
关键词
meniscal repair; all-inside; inside-out; primary fixation; biomechanics; BIOMECHANICAL EVALUATION; SUTURE TECHNIQUE; FAILURE; MENISCECTOMY; STRENGTH; LESIONS; ANCHORS; LOAD;
D O I
10.1177/03635465221107086
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Primary device fixation and the resistance against gap formation during repetitive loading influence the quality of meniscal repair. There are limited biomechanical data comparing primary tensioning and cyclic behavior of all-inside versus inside-out repair. Hypothesis: All-inside devices provide higher initial load on the meniscal repair than inside-out fixation, and stiffer constructs show higher resistance against gap formation during cyclic loading. Study Design: Controlled laboratory study. Methods: In total, 60 longitudinal bucket-handle tears in human cadaveric menisci were created and repaired with a single stitch and randomly assigned to 4 all-inside groups (TrueSpan, FastFix 360, Stryker AIR, FiberStich) and 2 inside-out groups (suture repair [IO-S], suture tape [IO-ST]). Residual load after repair tensioning (50 N) and relief displacement were measured. Constructs underwent cyclic loading between 2 and 20 N over 500 cycles (0.75 Hz) with cyclic stiffness, gap formation, and final peak elongation measured. Ultimate load and stiffness were analyzed during pull to failure (3.15 mm/s). Results: All-inside repair demonstrated significantly higher primary fixation strength than inside-out repair. The significantly highest load (mean +/- SD; 20.1 +/- 0.9 N; P < .037) and relief displacement (-2.40 +/- 0.32 mm; P < .03) were for the knotless soft anchoring FiberStich group. The lowest initial load (9.0 +/- 1.5 N; P < .001) and relief displacement (-1.39 +/- 0.26 mm; P < .045) were for the IO-S repair group. The final gap formation (500th cycle) of FiberStich (0.75 +/- 0.37 mm; P < .02) was significantly smaller than others and that of the IO-S (1.47 +/- 0.33 mm; P < .045) significantly larger. The construct stiffness of the FiberStich and IO-ST groups was significantly greater at the end of cyclic testing (16.7 +/- 0.80 and 15.5 +/- 1.42 N/mm; P < .042, respectively) and ultimate failure testing (23.4 +/- 3.6 and 20.6 +/- 2.3 N/mm; P < .005). The FastFix 360 (86.4 +/- 4.8 N) and Stryker AIR (84.4 +/- 4.6 N) groups failed at a significantly lower load than the IO-S group (P < .02) with loss of anchor support. The FiberStich (146.8 +/- 23.4 N), TrueSpan (142.0 +/- 17.8 N), and IO-ST (139.4 +/- 7.3 N) groups failed at significantly higher loads (P < .02) due to suture tearing. Conclusion: Overall, primary fixation strength of inside-out meniscal repair was significantly lower than all-inside repair in this cadaveric tissue model. Although absolute differences among groups were small, meniscal repairs with higher construct stiffness (IO-ST, FiberStich) demonstrated increased resistance against gap formation and failure load.
引用
收藏
页码:2705 / 2713
页数:9
相关论文
共 41 条
[1]   Treatment of Acute Proximal Anterior Cruciate Ligament Tears-Part 1: Gap Formation and Stabilization Potential of Repair Techniques [J].
Bachmaier, Samuel ;
DiFelice, Gregory S. ;
Sonnery-Cottet, Bertrand ;
Douoguih, Wiemi A. ;
Smith, Patrick A. ;
Pace, Lee J. ;
Ritter, Daniel ;
Wijdicks, Coen A. .
ORTHOPAEDIC JOURNAL OF SPORTS MEDICINE, 2020, 8 (01)
[2]   Meniscal repair with the RapidLoc meniscal repair device [J].
Barber, F. Alan ;
Coons, David A. ;
Ruiz-Suarez, Michell .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (09) :962-966
[3]   The Biomechanical Performance of the Latest All-Inside Meniscal Repair Devices [J].
Barber, F. Alan ;
Howard, Michael S. ;
Ashraf, Wasik ;
Spenciner, David B. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2020, 36 (12) :3001-3007
[4]   Biomechanical Testing of Suture-Based Meniscal Repair Devices Containing Ultrahigh-Molecular-Weight Polyethylene Suture: Update 2011 [J].
Barber, F. Alan ;
Herbert, Morley A. ;
Bava, Eric D. ;
Drew, Otis R. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2012, 28 (06) :827-834
[5]   Load to failure testing of new meniscal repair devices [J].
Barber, FA ;
Herbert, MA ;
Richards, DP .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2004, 20 (01) :45-50
[6]   Analysis of a New All-Inside Versus Inside-Out Technique for Repairing Radial Meniscal Tears [J].
Beamer, Brandon S. ;
Masoudi, Aidin ;
Walley, Kempland C. ;
Harlow, Ethan R. ;
Manoukian, Ohan S. ;
Hertz, Ben ;
Haeussler, Claudia ;
Olson, Jeffrey J. ;
Deangelis, Joseph P. ;
Nazarian, Ara ;
Ramappa, Arun J. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2015, 31 (02) :293-298
[7]   Biomechanical evaluation of meniscal repair systems - A comparison of the Meniscal Viper Repair System, the vertical mattress FasT-Fix device, and vertical mattress ethibond sutures [J].
Chang, HC ;
Nyland, J ;
Caborn, DNM ;
Burden, R .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2005, 33 (12) :1846-1852
[8]   Clinical Outcomes of Inside-Out Meniscal Repair According to Anatomic Zone of the Meniscal Tear [J].
Cinque, Mark E. ;
DePhillipo, Nicholas N. ;
Moatshe, Gilbert ;
Chahla, Jorge ;
Kennedy, Mitchell, I ;
Doman, Grant J. ;
LaPrade, Robert F. .
ORTHOPAEDIC JOURNAL OF SPORTS MEDICINE, 2019, 7 (07)
[9]   Chondral injury after arthroscopic meniscal repair using bioabsorbable Mitek RapidLoc meniscal fixation [J].
Cohen, SB ;
Anderson, MW ;
Miller, MD .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2003, 19 (07) :e73-e75
[10]   Poisson's ratio of bovine meniscus determined combining unconfined and confined compression [J].
Danso, E. K. ;
Julkunen, P. ;
Korhonen, R. K. .
JOURNAL OF BIOMECHANICS, 2018, 77 :233-237