Spatially Variant Resolution Modelling for Iterative List-Mode PET Reconstruction

被引:0
作者
Bickell, Matthew G. [1 ]
Zhou, Lin [1 ]
Nuyts, Johan [1 ]
机构
[1] Katholieke Univ Leuven, Dept Nucl Med, Med Imaging Res Ctr, Leuven, Belgium
来源
2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2013年
关键词
IMAGE-RECONSTRUCTION; MATHEMATICAL REMOVAL; BIOGRAPH MCT; EM ALGORITHM; SYSTEM;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A spatially variant resolution modelling technique is presented which estimates the system matrix on-the-tly. The method randomly redistributes the line-of-response endpoints according to probability density functions describing the detector response and photon acollinearity effects. A list-mode OSEM algorithm is used for the reconstruction. We demonstrate that this model agrees with measured Pit's and we present results showing an improvement in resolution recovery as compared to using a Gaussian convolution to model the resolution, although with an increase in noise. We also present results from applying this method to event-by-event rigid motion correction with list-mode reconstruction using the microPET Focus220 seamier.
引用
收藏
页数:6
相关论文
共 50 条
[21]   A regularized relaxed ordered subset list-mode reconstruction algorithm and its preliminary application to undersampling PET imaging [J].
Cao, Xiaoqing ;
Xie, Qingguo ;
Xiao, Peng .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (01) :49-66
[22]   A Cube-Based Dual-GPU List-Mode Reconstruction Algorithm for PET Imaging [J].
Teimoorisichani, Mohammadreza ;
Goertzen, Andrew L. .
IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2022, 6 (04) :463-474
[23]   List-mode quantitative joint reconstruction of activity and attenuation maps in Time-of-Flight PET [J].
Hemmati, Hamidreza ;
Kamali-Asl, Alireza ;
Ghafarian, Pardis ;
Rahmim, Arman ;
Zaidi, Habib ;
Ay, Mohammad Reza .
JOURNAL OF INSTRUMENTATION, 2023, 18 (09)
[24]   An accurate probabilistic model with detector resolution and Doppler broadening correction in list-mode MLEM reconstruction for Compton camera [J].
Wu, Chuanpeng ;
Zhang, Siyuan ;
Li, Liang .
PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (12)
[25]   Parallel List-Mode Reconstruction and Calculation of the System Matrix for the High-Resolution ClearPET™ Neuro [J].
Musmann, Patrick ;
Pietrzyk, Uwe ;
Schramm, Nils ;
Weber, Simone .
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, :2783-2786
[26]   Hybrid PET-MR list-mode kernelized expectation maximization reconstruction for quantitative PET images of the carotid arteries [J].
Deidda, Daniel ;
Karakatsanis, Nicolas ;
Robson, Philip M. ;
Efthimiou, Nikos ;
Fayad, Zahi A. ;
Aykroyd, Robert G. ;
Tsoumpas, Charalampos .
2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
[27]   Fully 4D list-mode reconstruction applied to respiratory-gated PET scans [J].
Grotus, N. ;
Reader, A. J. ;
Stute, S. ;
Rosenwald, J. C. ;
Giraud, P. ;
Buvat, I. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (06) :1705-1721
[28]   Studies of DOI estimation method in LOR data for 3D PET list-mode reconstruction [J].
Wu, Jing ;
Ma, Tianyu ;
Liu, Hui ;
Liang, Juncheng ;
Wang, Shi ;
Liu, Yaqiang ;
Cheng, Jianping .
2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
[29]   Likelihood maximization for list-mode emission tomographic image reconstruction [J].
Byrne, C .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (10) :1084-1092
[30]   A Fast CPU/GPU Ray Projector for Fully 3D List-Mode PET Reconstruction [J].
Bert, Julien ;
Visvikis, Dimitris .
2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, :4126-4130