Extraction of the photon spectra from measured beam parameters

被引:12
作者
Bloch, P [1 ]
McDonough, J [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
关键词
x-ray spectrum; radiotherapy; spectrum reconstruction; transmission; buildup;
D O I
10.1118/1.598257
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Knowledge of the photon spectrum of a radiotherapy beam is often needed for three-dimensional (3-D) dose calculations using Monte Carlo methods and/or algorithms employing energy deposition kernels. Direct measurement of the x-ray energy fluence spectrum is not feasible for the high-energy photon beams used clinically. In this paper, the spectrum is extracted from basic beam data that are readily obtained for a clinical beam. We describe the photon spectrum using just two parameters. One parameter, which determines the high-energy part of the spectrum, is obtained using the measured dose in the buildup region for a small field, where electron contamination of the beam can be neglected. The other parameter is extracted from the photon beam attenuation in water. The results compare favorably to spectra generated from Monte Carlo simulations. (C) 1998 American Association of Physicists in Medicine.
引用
收藏
页码:752 / 757
页数:6
相关论文
共 18 条
[1]   DETERMINATION OF EFFECTIVE BREMSSTRAHLUNG SPECTRA AND ELECTRON CONTAMINATION FOR PHOTON DOSE CALCULATIONS [J].
AHNESJO, A ;
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (10) :1451-1464
[2]   3D DOSE CALCULATIONS FOR ELECTRON AND PHOTON BEAMS [J].
ALTSCHULER, MD ;
BLOCH, P ;
BUHLE, EL ;
AYYALASOMAYAJULA, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (02) :391-411
[3]   RECONSTRUCTION OF MEGAVOLTAGE PHOTON SPECTRA BY ATTENUATION ANALYSIS [J].
BAKER, CR ;
AMAEE, B ;
SPYROU, NM .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04) :529-542
[4]   ANALYSIS OF CENTRAL-AXIS DOSES FOR HIGH-ENERGY X-RAYS [J].
BJARNGARD, BE ;
VADASH, P .
MEDICAL PHYSICS, 1995, 22 (07) :1191-1195
[5]  
BJARNGARD BE, 1994, MED PHYS, V21, P1069, DOI 10.1118/1.597349
[6]  
BJARNGARD BE, 1995, MED PHYS, V22, P465, DOI 10.1118/1.597477
[7]   RECONSTRUCTION OF 12-MV BREMSSTRAHLUNG SPECTRA FROM MEASURED TRANSMISSION DATA BY DIRECT RESOLUTION OF THE NUMERIC SYSTEM AF=T [J].
CATALA, A ;
FRANCOIS, P ;
BONNET, J ;
SCOUARNEC, C .
MEDICAL PHYSICS, 1995, 22 (01) :3-10
[8]  
DESBROY GE, 1991, MED PHYS, V18, P497
[9]   Validation of reconstructed bremsstrahlung spectra between 6 MV and 25 MV from measured transmission data [J].
Francois, P ;
Coste, F ;
Bonnet, J ;
Caselles, O .
MEDICAL PHYSICS, 1997, 24 (05) :769-773
[10]  
HALBLEIB JA, 1984, 840573 SAND NAT LAB