Evaluation of the basic properties of the BANGkit™ gel dosimeter

被引:15
作者
Murakami, Y.
Nakashima, T.
Watanabe, Y.
Akimitsu, T.
Matsuura, K.
Kenjo, M.
Kaneyasu, Y.
Wadasaki, K.
Hirokawa, Y.
Ito, K.
机构
[1] Hiroshima Univ, Grad Sch Med, Dept Radiol, Hiroshima 7348551, Japan
[2] Hiroshima Univ, Grad Sch Med, Dept Oral & Maxillofacial Radiol, Hiroshima 7348551, Japan
[3] Univ Minnesota, Dept Therapeut Radiol, Minneapolis, MN 55455 USA
[4] Takanobashi Cent Hosp, Naka Ku, Hiroshima 7300042, Japan
[5] Hiroshima Heiwa Clin Hosp, Hiroshima 7300856, Japan
关键词
D O I
10.1088/0031-9155/52/8/017
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We evaluated the basic properties of a commercially available BANGkit(TM) gel dosimeter, which is a normoxic type of BANG((R)) gel. This gel-kit has the same composition as the BANG((R))3 gel, but is fully oxygenated. To exclude oxygen, oxygen scavenging ascorbic acid and copper sulfate as a catalyst are used. The properties that we examined are the effects of the concentrations of copper sulfate and ascorbic acid on the response, the reproducibility, the longterm stability, the temperature effect at irradiation and the dose-rate effect. In our results, the excellent linear fit of the R-2-dose response in a dose range for clinical use and its reproducibility were observed. The precision of a linear fit was preserved for about 3 weeks. The temperature at irradiation showed a significant effect on the dose response. Although the dose-rate dependence in the high-dose range was observed, it was negligible for the clinical dose range up to 270 cGy. In conclusion, this gel dosimeter is thought to be utilizable in clinical practice, while we have to pay attention to the temperature during the entire measurement processes, and additionally there is room for improvement in the linearity and the dose- rate dependence in the high-dose range.
引用
收藏
页码:2301 / 2311
页数:11
相关论文
共 22 条
[1]  
[Anonymous], [No title captured], Patent No. 5321357
[2]   Basic investigations on the performance of a normoxic polymer gel with tetrakis-hydroxy-methyl-phosphonium chloride as an oxygen scavenger: Reproducibility, accuracy, stability, and dose rate dependence [J].
Bayreder, Christian ;
Georg, Dietmar ;
Moser, Ewald ;
Berg, Andreas .
MEDICAL PHYSICS, 2006, 33 (07) :2506-2518
[3]   The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gels [J].
De Deene, Y ;
Vergote, K ;
Claeys, C ;
De Wagter, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (03) :653-673
[4]   Three-dimensional dosimetry using polymer gel and magnetic resonance imaging applied to the verification of conformal radiation therapy in head-and-neck cancer [J].
De Deene, Y ;
De Wagter, C ;
Van Duyse, B ;
Derycke, S ;
De Neve, W ;
Achten, E .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) :283-291
[5]  
De Deene Y, 2002, PHYS MED BIOL, V47, P2459, DOI 10.1088/0031-9155/47/14/307
[6]  
De Deene Y, 2002, PHYS MED BIOL, V47, P3441, DOI 10.1088/0031-9155/47/19/301
[7]  
De Deene Y, 2000, PHYS MED BIOL, V45, P859, DOI 10.1088/0031-9155/45/4/304
[8]   Polymer gels for magnetic resonance imaging of radiation dose distributions at normal room atmosphere [J].
Fong, PM ;
Keil, DC ;
Does, MD ;
Gore, JC .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (12) :3105-3113
[9]   MAGIC-type polymer gel for three-dimensional dosimetry:: Intensity-modulated radiation therapy verification [J].
Gustavsson, H ;
Karlsson, A ;
Bäck, SÅJ ;
Olsson, LE ;
Haraldsson, P ;
Engström, P ;
Nyström, H .
MEDICAL PHYSICS, 2003, 30 (06) :1264-1271
[10]   A study of a normoxic polymer gel dosimeter comprising methacrylic acid, gelatin and tetrakis (hydroxymethyl) phosphonium chloride (MAGAT) [J].
Hurley, C ;
Venning, A ;
Baldock, C .
APPLIED RADIATION AND ISOTOPES, 2005, 63 (04) :443-456