Simple high-performance liquid chromatographic method for determination of atropine and obidoxime in a parenteral injection device

被引:26
作者
Gören, AC [1 ]
Bilsel, G [1 ]
Bilsel, M [1 ]
Yenisoy-Karakas, S [1 ]
Karakas, D [1 ]
机构
[1] Natl Metrol Inst, TUBITAK UME, TR-41470 Gebze, Turkey
关键词
atropine; obidoxime; method validation; uncertainty of the method;
D O I
10.1016/j.chroma.2004.09.048
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Atropine and obidoxime in a parenteral injection device are determined by simple HPLC method simultaneously without any pretreatment at 228 nm. The relative standard deviations (R.S.D.) were below 1.6% for the compounds. The correlation coefficient was greater than 0.999 for both compounds in the calibration range. The recoveries at 5 mg/L concentration averaged as 95% for atropine and 102% for obidoxime. The uncertainty of the measurements for atropine and obidoxime was 2.8% and 2.4%, respectively. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:237 / 239
页数:3
相关论文
共 10 条
[1]  
[Anonymous], 2000, EURACHEM CITAC GUID
[2]   DETERMINATION OF SOME PYRIDINIUM ALDOXIME COMPOUNDS BY MEANS OF ION-PAIR REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - APPLICATION IN BIOLOGICAL-MATERIAL [J].
BENSCHOP, HP ;
KONINGS, KAG ;
KOSSEN, SP ;
LIGTENSTEIN, DA .
JOURNAL OF CHROMATOGRAPHY, 1981, 225 (01) :107-114
[3]   Chemical weapons: documented use and compounds on the horizon [J].
Bismuth, C ;
Borron, SW ;
Baud, FJ ;
Barriot, P .
TOXICOLOGY LETTERS, 2004, 149 (1-3) :11-18
[4]   GAS-CHROMATOGRAPHIC ASSAY OF ATROPINE IN FORMULATIONS CONTAINING ATROPINE SULFATE AND CHOLINESTERASE REACTIVATORS [J].
BRIGGS, CJ ;
SIMONS, KJ .
JOURNAL OF CHROMATOGRAPHY, 1983, 257 (01) :132-136
[5]  
BROWN ND, 1978, J CHROMATOGR, V148, P453
[6]   Internal standard high-performance liquid chromatography method for the determination of obidoxime in urine of organophosphate-poisoned patients [J].
Grasshoff, C ;
Thiermann, H ;
Gillessen, T ;
Zilker, T ;
Szinicz, L .
JOURNAL OF CHROMATOGRAPHY B, 2001, 753 (02) :203-208
[7]   SIMPLE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR ASSESSING THE DETERIORATION OF ATROPINE-OXIME MIXTURES EMPLOYED AS ANTIDOTES IN THE TREATMENT OF NERVE AGENT POISONING [J].
PADDLE, BM ;
DOWLING, MH .
JOURNAL OF CHROMATOGRAPHY, 1993, 648 (02) :373-380
[8]   DETERMINATION OF ATROPINE AND OBIDOXIME IN AUTOMATIC INJECTION DEVICES USED AS ANTIDOTES AGAINST NERVE AGENT INTOXICATION [J].
POHJOLA, J ;
HARPF, M .
JOURNAL OF CHROMATOGRAPHY A, 1994, 686 (02) :350-354
[9]   SEPARATION OF GEOMETRICAL SYN AND ANTI ISOMERS OF OBIDOXIME BY ION-PAIR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
SPOHRER, U ;
EYER, P .
JOURNAL OF CHROMATOGRAPHY A, 1995, 693 (01) :55-61
[10]   Modern strategies in therapy of organophosphate poisoning [J].
Thiermann, H ;
Szinicz, L ;
Eyer, F ;
Worek, F ;
Eyer, P ;
Felgenhauer, N ;
Zilker, T .
TOXICOLOGY LETTERS, 1999, 107 (1-3) :233-239