Multiple Overlapping k-Space Junctions for Investigating Translating Objects (MOJITO)

被引:11
作者
Bookwalter, Candice A. [1 ,2 ]
Griswold, Mark A. [1 ,2 ]
Duerk, Jeffrey L. [1 ,2 ]
机构
[1] Case Western Reserve Univ, Dept Radiol, Case Ctr Imaging Res, Cleveland, OH 44106 USA
[2] Case Western Reserve Univ, Dept Biomed Engn, Case Ctr Imaging Res, Cleveland, OH 44106 USA
关键词
Continuous sampling; motion correction; magnetic resonance imaging (MRI); MOTION CORRECTION; GRIDDING ALGORITHM; NAVIGATOR ECHOES; MRI; RECONSTRUCTION; EFFICIENT; SNR;
D O I
10.1109/TMI.2009.2029854
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
It is a well-known property in Fourier transform magnetic resonance imaging (MRI) that rigid body translational motion in image space results in linear phase accumulation in k-space. This work describes Multiple Overlapping k-space Junctions for Investigating Translating Objects (MOJITO), a correction scheme based on phase differences at trajectory intersections caused by 2-D alterations in the position of an object during MR imaging. The algorithm allows both detection and correction of motion artifacts caused by 2-D rigid body translational motion. Although similar in concept to navigator echoes, MOJITO does not require a repeating path through k-space, uses k-space data from a broader region of k-space, and uses the repeated data in image reconstruction; this provides the potential for a highly efficient self-navigating motion correction method. Here, the concept and theoretical basis of MOJITO is demonstrated using the continuous sampling BOWTIE trajectory in simulation and MR experiments. This particular trajectory is selected since it is well suited for such an algorithm due to numerous trajectory intersections. Specifically, the validity of the technique in the presence of noise and off-resonance effects is demonstrated through simulation.
引用
收藏
页码:339 / 349
页数:11
相关论文
共 37 条
[11]   ORBITAL NAVIGATOR ECHOES FOR MOTION MEASUREMENTS IN MAGNETIC-RESONANCE-IMAGING [J].
FU, ZW ;
WANG, Y ;
GRIMM, RC ;
ROSSMAN, PJ ;
FELMLEE, JP ;
RIEDERER, SJ ;
EHMAN, RL .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (05) :746-753
[12]   Correction of motion artifacts in linogram and projection reconstruction MRI using geometry and consistency constraints [J].
Gai, N ;
Axel, L .
MEDICAL PHYSICS, 1996, 23 (02) :251-262
[13]   PROJECTION RECONSTRUCTION TECHNIQUES FOR REDUCTION OF MOTION EFFECTS IN MRI [J].
GLOVER, GH ;
PAULY, JM .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (02) :275-289
[14]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[15]   VD-AUTO-SMASH imaging [J].
Heidemann, RM ;
Griswold, MA ;
Haase, A ;
Jakob, PM .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (06) :1066-1074
[16]  
JACKSON J, 1991, IEEE T MED IMAGING, V10, P476
[17]   Self-gated cardiac cine MRI [J].
Larson, AC ;
White, RD ;
Laub, G ;
McVeigh, ER ;
Li, DB ;
Simonetti, OP .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (01) :93-102
[18]  
Lustig M., 2007, P ISMRM 15 SCI M, P865
[19]   A new way of averaging with applications to MRI [J].
Madore, B ;
Henkelman, RM .
MEDICAL PHYSICS, 1996, 23 (01) :109-113
[20]   Multifrequency interpolation for fast off-resonance correction [J].
Man, LC ;
Pauly, JM ;
Macovski, A .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (05) :785-792