Efficacy of a novel bipolar radiofrequency ablation device on the beating heart for atrial fibrillation ablation: A long-term porcine study

被引:18
作者
Voeller, Rochus K. [1 ]
Zierer, Andreas [1 ]
Lall, Shelly C. [1 ]
Sakamoto, Shun-ichiro [1 ]
Schuessler, Richard B. [1 ]
Damiano, Ralph J., Jr. [1 ]
机构
[1] Washington Univ, Sch Med, Barnes Jewish Hosp, Div Cardiothorac Surg, St Louis, MO 63110 USA
基金
美国国家卫生研究院;
关键词
COX-MAZE PROCEDURE; ALTERNATIVE ENERGY-SOURCES; SURGICAL-TREATMENT; TECHNOLOGY; SURGERY; FLUTTER; TRIAL;
D O I
10.1016/j.jtcvs.2009.06.034
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Over recent years, a variety of energy sources have been used to replace the traditional incisions of the Cox maze procedure for the surgical treatment of atrial fibrillation. This study evaluated the safety and efficacy of a new bipolar radiofrequency ablation device for atrial ablation in a long-term porcine model. Methods: Six pigs underwent a Cox maze IV procedure on a beating heart off cardiopulmonary bypass using the AtriCure Isolator II bipolar ablation device (AtriCure, Inc, Cincinnati, Ohio). In addition, 6 pigs underwent median sternotomy and pericardiotomy alone to serve as a control group. All animals were allowed to survive for 30 days. Each pig underwent induction of atrial fibrillation and was then humanely killed to remove the heart en bloc for histologic assessment. Magnetic resonance imaging scans were also obtained preoperatively and postoperatively to assess atrial and ventricular function, pulmonary vein anatomy, valve function, and coronary artery patency. Results: All animals survived the operation. Electrical isolation of the left atrial appendage and the pulmonary veins was documented by pacing acutely and at 30 days in all animals. No animal that underwent the Cox maze IV procedure was able to be induced into atrial fibrillation at 30 days postoperatively, compared with all the sham animals. All 257 ablations examined were discrete, linear, and transmural, with a mean lesion width of 2.2 +/- 1.1 mm and a mean lesion depth of 5.3 +/- 3.0 mm. Conclusions: The AtriCure Isolator II device was able to create reliable long-term transmural lesions of the modified Cox maze procedure on a beating heart without cardiopulmonary bypass 100% of the time. There were no discernible effects on ventricular or valvular function. (J Thorac Cardiovasc Surg 2010; 140: 203-8)
引用
收藏
页码:203 / 208
页数:6
相关论文
共 35 条
[1]  
[Anonymous], 2004, OP TECH
[2]   Improved coronary artery definition with T2-weighted, free-breathing, three-dimensional coronary MRA [J].
Botnar, RM ;
Stuber, M ;
Danias, PG ;
Kissinger, KV ;
Manning, WJ .
CIRCULATION, 1999, 99 (24) :3139-3148
[3]  
COX JL, 1991, J THORAC CARDIOV SUR, V101, P584
[4]   THE SURGICAL-TREATMENT OF ATRIAL-FIBRILLATION .3. DEVELOPMENT OF A DEFINITIVE SURGICAL-PROCEDURE [J].
COX, JL ;
SCHUESSLER, RB ;
DAGOSTINO, HJ ;
STONE, CM ;
CHANG, BC ;
CAIN, ME ;
CORR, PB ;
BOINEAU, JP .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1991, 101 (04) :569-583
[5]   An 8 1/2-year clinical experience with surgery for atrial fibrillation [J].
Cox, JL ;
Schuessler, RB ;
Lappas, DG ;
Boineau, JP .
ANNALS OF SURGERY, 1996, 224 (03) :267-273
[6]   THE SURGICAL-TREATMENT OF ATRIAL-FIBRILLATION .2. INTRAOPERATIVE ELECTROPHYSIOLOGIC MAPPING AND DESCRIPTION OF THE ELECTROPHYSIOLOGIC BASIS OF ATRIAL-FLUTTER AND ATRIAL-FIBRILLATION [J].
COX, JL ;
CANAVAN, TE ;
SCHUESSLER, RB ;
CAIN, ME ;
LINDSAY, BD ;
STONE, C ;
SMITH, PK ;
CORR, PB ;
BOINEAU, JP .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1991, 101 (03) :406-426
[7]   MODIFICATION OF THE MAZE PROCEDURE FOR ATRIAL-FLUTTER AND ATRIAL-FIBRILLATION .1. RATIONALE AND SURGICAL RESULTS [J].
COX, JL ;
BOINEAU, JP ;
SCHUESSLER, RB ;
JAQUISS, RDB ;
LAPPAS, DG .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1995, 110 (02) :473-484
[8]   Alternative energy sources for the ablation of arrhythmias [J].
Cummings, JE ;
Pacifico, A ;
Drago, JL ;
Kilicaslan, F ;
Natale, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (05) :434-443
[9]   Surgical radiofrequency ablation induces coronary endothelial dysfunction in porcine coronary arteries [J].
Demaria, RG ;
Pagé, P ;
Leung, TK ;
Dubuc, M ;
Malo, O ;
Carrier, M ;
Perrault, LP .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2003, 23 (03) :277-282
[10]  
Doll N, 2003, THORAC CARDIOV SURG, V51, P267