Evaluation of the i-STAT Alinity v in a veterinary clinical setting

被引:4
作者
Burke, Jasper E. [1 ]
Nguyen, Thuy Hien T. [2 ]
Davis, Tanya [2 ]
Koenig, Amie [1 ]
Lane, Selena L. [1 ]
Good, Jennifer [1 ]
Brainard, Benjamin M. [1 ]
机构
[1] Univ Georgia, Coll Vet Med, Dept Small Anim Med & Surg, Athens, GA USA
[2] Zoetis Inc, Parsippany, NJ USA
关键词
blood gas; cats; dogs; electrolytes; emergency; point-of-care; BLOOD-GASES; ANALYZER; PH; ELECTROLYTES; GUIDELINES; ACCURACY;
D O I
10.1177/10406387211019710
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Many point-of-care (POC) analyzers are available for the measurement of electrolytes and acid-base status in animals. We assessed the precision of the i-STAT Alinity v, a recently introduced POC analyzer, and compared it to 2 commonly used and previously validated POC analyzers (i-STAT 1, Stat Profile pHOx Ultra). Precision was evaluated by performing multiple analyses of whole blood samples from healthy dogs, cats, and horses on multiple i-STAT Alinity v analyzers. For comparison between analyzers, whole blood samples from dogs and cats presented to the emergency room were run concurrently on all 3 POC instruments. Reported values were compared by species (dogs and cats only) using Pearson correlation, and all values from all species were analyzed together for the Bland-Altman analysis. Results suggested that the i-STAT Alinity v precision was very good, with median coefficients of variability <2.5% for all measured parameters (except the anion gap), with variable ranges of coefficients of variation. In addition, good-to-excellent correlation was observed between the i-STAT Alinity v and i-STAT 1, and between the i-STAT Alinity v and Stat Profile pHOx Ultra for all parameters in both cats and dogs, respectively. In this cohort, the i-STAT Alinity v had clinically acceptable bias compared to the currently marketed analyzers and can be used for monitoring measured analytes in cats and dogs, although serial measurements in a single animal should be performed on the same analyzer whenever possible.
引用
收藏
页码:703 / 710
页数:8
相关论文
共 50 条
  • [21] Analytical performances of the i-STAT portable clinical analyzer for the measurement of PO2 and PCO2
    Sachs, C
    Finetti, P
    Rabouine, P
    Abdoulaye, F
    ANNALES DE BIOLOGIE CLINIQUE, 2002, 60 (04) : 411 - 420
  • [22] Next-generation Diagnostics for Melioidosis: Evaluation of a Prototype i-STAT Cartridge to Detect Burkholderia pseudomallei Biomarkers
    Schully, Kevin L.
    Young, Charles C.
    Mayo, Mark
    Connolly, Amy L.
    Rigas, Vanessa
    Spall, Ammarah
    Chan, Alyssa A.
    Salvador, Mark G.
    Lawler, James V.
    Opdyke, Jason A.
    Clark, Danielle V.
    Currie, Bart J.
    CLINICAL INFECTIOUS DISEASES, 2019, 69 (03) : 421 - 427
  • [23] The rise and fall of i-STAT point-of-care blood gas testing in an acute care hospital
    Ng, VL
    Kraemer, R
    Hogan, C
    Eckman, D
    Siobal, M
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 114 (01) : 128 - 138
  • [24] High altitude arterialised capillary earlobe blood gas measurement using the Abbott i-STAT
    Lewis, Christopher T.
    Malein, W. L.
    Chesner, I.
    Clarke, S.
    JOURNAL OF THE ROYAL ARMY MEDICAL CORPS, 2018, 164 (05) : 335 - 337
  • [25] Validation of the i-STAT system for the analysis of blood gases and acid-base status in juvenile sandbar shark (Carcharhinus plumbeus)
    Harter, T. S.
    Morrison, P. R.
    Mandelman, J. W.
    Rummer, J. L.
    Farrell, A. P.
    Brill, R. W.
    Brauner, C. J.
    CONSERVATION PHYSIOLOGY, 2015, 3
  • [26] A review of temperature-related challenges and solutions for the Abbott i-STAT and Siemens Healthineers epoc devices
    Fuzery, Anna K.
    Elian, Fahed A.
    Kost, Gerald J.
    CLINICAL BIOCHEMISTRY, 2023, 115 : 49 - 66
  • [27] Measurement of Activated Coagulation Time in Children: Evaluation of the Blood-Saving Kaolin i-STAT Activated Coagulation Time Technique in Pediatric Cardiac Anesthesia
    Ulmer, Francis F.
    Baulig, Werner
    Bettex, Dominique
    Spielmann, Nelly
    Buerki, Christoph
    Weiss, Markus
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2011, 25 (03) : 395 - 401
  • [28] Selected Blood Chemistry Values in Mobility-Impaired Broiler Breeder Hens with Suspected Calcium Tetany Using the i-STAT® Handheld Clinical Analyzer
    Martin, Michael P.
    Wineland, Michael
    Fletcher, Oscar J.
    Barnes, H. John
    AVIAN DISEASES, 2011, 55 (03) : 340 - 345
  • [29] Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus
    Harter, T. S.
    Reichert, M.
    Brauner, C. J.
    Milsom, W. K.
    CONSERVATION PHYSIOLOGY, 2015, 3
  • [30] Reliability of the i-STAT for the determination of blood electrolyte (K+, Na+, and CI-) concentrations in cattle
    Yildirim, E.
    Karapinar, T.
    Hayirli, A.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2015, 29 (01) : 388 - 394