Evaluation of the i-STAT Alinity v in a veterinary clinical setting

被引:4
|
作者
Burke, Jasper E. [1 ]
Nguyen, Thuy Hien T. [2 ]
Davis, Tanya [2 ]
Koenig, Amie [1 ]
Lane, Selena L. [1 ]
Good, Jennifer [1 ]
Brainard, Benjamin M. [1 ]
机构
[1] Univ Georgia, Coll Vet Med, Dept Small Anim Med & Surg, Athens, GA USA
[2] Zoetis Inc, Parsippany, NJ USA
关键词
blood gas; cats; dogs; electrolytes; emergency; point-of-care; BLOOD-GASES; ANALYZER; PH; ELECTROLYTES; GUIDELINES; ACCURACY;
D O I
10.1177/10406387211019710
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Many point-of-care (POC) analyzers are available for the measurement of electrolytes and acid-base status in animals. We assessed the precision of the i-STAT Alinity v, a recently introduced POC analyzer, and compared it to 2 commonly used and previously validated POC analyzers (i-STAT 1, Stat Profile pHOx Ultra). Precision was evaluated by performing multiple analyses of whole blood samples from healthy dogs, cats, and horses on multiple i-STAT Alinity v analyzers. For comparison between analyzers, whole blood samples from dogs and cats presented to the emergency room were run concurrently on all 3 POC instruments. Reported values were compared by species (dogs and cats only) using Pearson correlation, and all values from all species were analyzed together for the Bland-Altman analysis. Results suggested that the i-STAT Alinity v precision was very good, with median coefficients of variability <2.5% for all measured parameters (except the anion gap), with variable ranges of coefficients of variation. In addition, good-to-excellent correlation was observed between the i-STAT Alinity v and i-STAT 1, and between the i-STAT Alinity v and Stat Profile pHOx Ultra for all parameters in both cats and dogs, respectively. In this cohort, the i-STAT Alinity v had clinically acceptable bias compared to the currently marketed analyzers and can be used for monitoring measured analytes in cats and dogs, although serial measurements in a single animal should be performed on the same analyzer whenever possible.
引用
收藏
页码:703 / 710
页数:8
相关论文
共 50 条
  • [1] Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus)
    Steinmetz, Hanspeter W.
    Vogt, Rainer
    Kastner, Sabine
    Riond, Barbara
    Hatt, Jean-Michel
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2007, 19 (04) : 382 - 388
  • [2] Evaluation of i-STAT portable clinical analyzer in a neonatal and pediatric intensive care unit
    Murthy, JN
    Hicks, JM
    Soldin, SJ
    CLINICAL BIOCHEMISTRY, 1997, 30 (05) : 385 - 389
  • [3] Evaluation of the i-STAT Portable Clinical Analyzer in dogs
    Matburger, C
    Henke, J
    Hirschberger, J
    Matburger, S
    Erhardt, W
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2000, 28 (02): : 132 - 137
  • [4] Evaluation of i-STAT portable clinical analyzer in a hemodialysis unit
    Gault, MH
    Harding, CE
    CLINICAL BIOCHEMISTRY, 1996, 29 (02) : 117 - 124
  • [5] Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer
    Larcher, Romaric
    Lottelier, Maxence
    Badiou, Stephanie
    Dupuy, Anne-Marie
    Bargnoux, Anne-Sophie
    Cristol, Jean-Paul
    DIAGNOSTICS, 2023, 13 (02)
  • [6] Evaluation of the portable blood gas analyser i-STAT
    Tschudi, PR
    SCHWEIZER ARCHIV FUR TIERHEILKUNDE, 1998, 140 (12): : 507 - 512
  • [7] EVALUATION OF A NOVEL POINT-OF-CARE SYSTEM, THE I-STAT PORTABLE CLINICAL ANALYZER
    ERICKSON, KA
    WILDING, P
    CLINICAL CHEMISTRY, 1993, 39 (02) : 283 - 287
  • [8] Analysis of Blood Biochemistry of Free Ranging and Human-Managed Southern White Rhinoceros (Ceratotherium simum simum) Using the i-STAT Alinity v®
    Trivedi, Shweta
    Burnham, Christina M.
    Capobianco, Christian M.
    Boshoff, Christiaan
    Zheng, Yaxin
    Pettiglio, Jordan Wood
    Ange-van Heugten, Kimberly
    Bissell, Heidi D.
    Minter, Larry J.
    VETERINARY MEDICINE INTERNATIONAL, 2021, 2021
  • [9] ANALYTICAL EVALUATION OF I-STAT PORTABLE CLINICAL ANALYZER AND USE BY NONLABORATORY HEALTH-CARE PROFESSIONALS
    JACOBS, E
    VADASDI, E
    SARKOZI, L
    COLMAN, N
    CLINICAL CHEMISTRY, 1993, 39 (06) : 1069 - 1074
  • [10] Evaluation of the i-STAT hand-held chemical analyser during treadmill and endurance exercise
    Silverman, S. C.
    Birks, E. K.
    EQUINE VETERINARY JOURNAL, 2002, 34 : 551 - 554