The impact of workload on the ability to localize audible alarms

被引:18
作者
Edworthy, Judy [1 ]
Reid, Scott [1 ]
Peel, Katie [1 ]
Lock, Samantha [1 ]
Williams, Jessica [1 ]
Newbury, Chloe [1 ]
Foster, Joseph [1 ]
Farrington, Martin [1 ]
机构
[1] Plymouth Univ, Cognit Inst, Plymouth PL4 8AA, Devon, England
关键词
Auditory alarms; Clinical alarms; Clinical safety; Patient safety; Auditory processes; Alarm fatigue; MEDICAL EQUIPMENT ALARMS; INATTENTIONAL DEAFNESS; SETS;
D O I
10.1016/j.apergo.2018.05.006
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Very little is known about people's ability to localize sound under varying workload conditions, though it would be expected that increasing workload should degrade performance. A set of eight auditory clinical alarms already known to have relatively high localizability (the ease with which their location is identified) when tested alone were tested in six conditions where workload was varied. Participants were required to indicate the location of a series of alarms emanating at random from one of eight speaker locations. Additionally, they were asked to read, carry out mental arithmetic tasks, be exposed to typical ICU noise, or carry out either the reading task or the mental arithmetic task in ICU noise. Performance in the localizability task was best in the control condition (no secondary task) and worst in those tasks which involved both a secondary task and noise. The data does therefore demonstrate the typical pattern of increasing workload affecting a primary task in an area where there is little data. In addition, the data demonstrates that performance in the control condition results in a missed alarm on one in ten occurrences, whereas performance in the heaviest workload conditions results in a missed alarm on every fourth occurrence. This finding has implications for the understanding of both 'in-attentional deafness' and 'alarm fatigue' in clinical environments.
引用
收藏
页码:88 / 93
页数:6
相关论文
共 30 条
[1]  
[Anonymous], 6060118 IEC
[2]  
[Anonymous], 1982, 82017 CIV AV AUTH
[3]  
[Anonymous], THESIS
[4]  
[Anonymous], SOUND EFF
[5]  
[Anonymous], 2011, BIOMED INSTRUM TECHN
[6]   Comparison of the identification and ease of use of two alarm sound sets by critical and acute care nurses with little or no music training: a laboratory study [J].
Atyeo, J. ;
Sanderson, P. M. .
ANAESTHESIA, 2015, 70 (07) :818-827
[7]  
Blauert J., 1997, Spatial hearing: the psychophysics of human sound localization
[8]   HUMAN PROBABILITY MATCHING BEHAVIOR IN RESPONSE TO ALARMS OF VARYING RELIABILITY [J].
BLISS, JP ;
GILSON, RD ;
DEATON, JE .
ERGONOMICS, 1995, 38 (11) :2300-2312
[9]   Behavioural implications of alarm mistrust as a function of task workload [J].
Bliss, JP ;
Dunn, MC .
ERGONOMICS, 2000, 43 (09) :1283-1300
[10]   Localizable auditory warning pulses [J].
Catchpole, K ;
McKeown, JD ;
Withington, DJ .
ERGONOMICS, 2004, 47 (07) :748-771