Percutaneous US-guided Renal Biopsy: A Retrospective Study Comparing the 16-gauge End-cut and 14-gauge Side-notch Needles

被引:32
作者
Constantin, Andre [1 ]
Brisson, Marie-Laure [2 ]
Kwan, Janet [1 ]
Proulx, Francesca [1 ]
机构
[1] McGill Univ, Jewish Gen Hosp, Dept Radiol, Montreal, PQ H3T 1E2, Canada
[2] McGill Univ, Jewish Gen Hosp, Dept Pathol, Montreal, PQ H3T 1E2, Canada
关键词
TRANSPLANT BIOPSY; CUTTING NEEDLE; ULTRASOUND; GUN;
D O I
10.1016/j.jvir.2009.11.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: Assess glomerular yield and safety profile of two different types of needles for percutaneous ultrasound-guided kidney biopsy. MATERIALS AND METHODS: Over 24 months, 121 ultrasonographic ultrasound-guided renal biopsies were performed on native kidneys of 121 adults: 66 with 16-gauge, 29-mm end-cut (BioPince) needles and 55 with 14-gauge, 1.9-mm side-notch (Tru-Cut) needles. RESULTS: The mean number of complete glomeruli harvested per biopsy was 21.0 and 19.3, respectively, and the mean number of core samples required to obtain a satisfactory biopsy was 1.8 and 2.6, respectively. The ratio of glomeruli harvested to core samples needed with the end-cut needle was 58% greater than that with the side-notch needles (11.7 vs 7.4, respectively; difference of 4.3; 95% confidence interval: 2.0, 6.8). Procedures performed with end-cut needles were associated with fewer major complications (1.5% vs 7.3% with side-notch needles). CONCLUSIONS: Compared to the 14-g Tru-cut needle, the 16-g end-cut needle provided better glomerular yield per core sample, required fewer cores for satisfactory tissue specimen, and resulted in fewer major complications.
引用
收藏
页码:357 / 361
页数:5
相关论文
共 11 条
[1]   NEEDLE-BIOPSY OF RENAL-ALLOGRAFTS - COMPARISON OF 2 TECHNIQUES [J].
BOGAN, ML ;
KOPECKY, KK ;
KRAFT, JL ;
HOLLADAY, AO ;
FILO, RS ;
LEAPMAN, SB ;
THOMALLA, JV .
RADIOLOGY, 1990, 174 (01) :273-275
[2]   CONVENTIONAL 15G NEEDLE TECHNIQUE FOR RENAL BIOPSY COMPARED WITH ULTRASOUND-GUIDED SPRING-LOADED 18G NEEDLE-BIOPSY [J].
COZENS, NJA ;
MURCHISON, JT ;
ALLAN, PL ;
WINNEY, RJ .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (775) :594-597
[3]  
Diederich S, 2006, Cancer Imaging, V6, P43, DOI 10.1102/1470-7330.2006.0008
[4]   A new core-biopsy instrument with an end-cut technique provides prostate biopsies with increased tissue yield [J].
Häggarth, L ;
Ekman, P ;
Egevad, L .
BJU INTERNATIONAL, 2002, 90 (01) :51-55
[5]   Percutaneous renal biopsy [J].
Korbet, SM .
SEMINARS IN NEPHROLOGY, 2002, 22 (03) :254-267
[6]   SAFETY AND EFFICACY OF KIDNEY-TRANSPLANT BIOPSY - TRU-CUT NEEDLE VS SONOGRAPHICALLY GUIDED BIOPTY GUN [J].
MAHONEY, MC ;
RACADIO, JM ;
MERHAR, GL ;
FIRST, MR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (02) :325-326
[7]   A prospective randomized trial of three different sizes of core-cutting needle for renal transplant biopsy [J].
Nicholson, ML ;
Wheatley, TJ ;
Doughman, TM ;
White, SA ;
Morgan, JDT ;
Veitch, PS ;
Furness, PN .
KIDNEY INTERNATIONAL, 2000, 58 (01) :390-395
[8]  
Ori Y, 2002, ISRAEL MED ASSOC J, V4, P698
[9]  
Sacks David, 2003, J Vasc Interv Radiol, V14, pS199
[10]  
Ubhayakar GN, 2002, BJU INT, V89, P40, DOI 10.1046/j.1464-410X.2002.02531.x