Force Transmission Error Analysis for a High-Pressure Single-Sinker Magnetic Suspension Densimeter

被引:24
作者
Cristancho, Diego E. [1 ]
Mantilla, Ivan D. [1 ]
Ejaz, Saquib [1 ]
Hall, Kenneth R. [1 ]
Iglesias-Silva, Gustavo A. [2 ]
Atilhan, Mert [3 ]
机构
[1] Texas A&M Univ, Artie McFerrin Dept Chem Engn, College Stn, TX 77843 USA
[2] Inst Tecnol Celaya, Dept Ingn Quim, Guanajuato 38010, Mexico
[3] Qatar Univ, Dept Chem Engn, Doha 2713, Qatar
关键词
Densimeter; Force transmission error; High pressure; Magnetic suspension coupling; ACCURATE DENSITY-MEASUREMENTS; EQUATION-OF-STATE; THERMODYNAMIC PROPERTIES; MELTING LINE; MPA; TEMPERATURES; RANGES;
D O I
10.1007/s10765-010-0702-3
中图分类号
O414.1 [热力学];
学科分类号
摘要
The magnetic suspension densimeter (MSD) is a sophisticated, state-of-the-art device that provides extremely accurate results for density measurements. The MSD uses a magnetic technique to couple a mass inside a measurement cell with an external mass balance for mass measurement. This article presents a force transmission error (FTE) analysis for a high-pressure, single-sinker MSD. Due to the magnetic working principle of the apparatus, magnetic properties of the high-pressure cell and external magnetic fields affect the measurements slightly. For the analysis, McLinden et al. suggest making measurements using two different sinkers, a titanium sinker and a copper sinker, having the same mass. The measurements cover densities for methane, ethane, carbon dioxide and nitrogen over the temperature range from 265 K to 450 K (+/- 5 mK stability) up to 180 MPa (uncertainty of 0.01 % full scale: 200 MPa). Comparing and manipulating the measurements permit determination of apparatus and fluid specific effects that contribute to the FTE. For this MSD, the apparatus effect is about 200 ppm, which effectively masks any fluid specific effect. A comprehensive analysis of the FTE produces a uniform deviation for density values of about 0.05 % at 2 sigma across the full range of pressure.
引用
收藏
页码:698 / 709
页数:12
相关论文
共 14 条
[1]  
[Anonymous], THESIS TEXAS A M U C
[2]  
[Anonymous], THESIS TEXAS A M U C
[3]   A reference equation of state for the thermodynamic properties of ethane for temperatures from the melting line to 675 K and pressures up to 900 MPa [J].
Bücker, D ;
Wagner, W .
JOURNAL OF PHYSICAL AND CHEMICAL REFERENCE DATA, 2006, 35 (01) :205-266
[4]   A new method for correcting a force transmission error due to magnetic effects in a magnetic levitation densimeter [J].
Kano, Yuya ;
Kayukawa, Yohei ;
Fujii, Kenichi ;
Sato, Haruki .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2007, 18 (03) :659-666
[5]  
Lemmon E., 2007, NIST standard reference database 23- NIST thermodynamic and transport properties REFPROP
[6]   Force transmission errors in magnetic suspension densimeters [J].
McLinden, M. O. ;
Kleinrahm, R. ;
Wagner, W. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2007, 28 (02) :429-448
[7]   A liquid density standard over wide ranges of temperature and pressure based on toluene [J].
McLinden, Mark O. ;
Splett, Jolene D. .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 2008, 113 (01) :29-67
[8]  
Patil, 2005, COMMISSIONING MAGENT
[9]   Accurate density measurements for a 91% methane natural gas-like mixture [J].
Patil, Prashant ;
Ejaz, Saquib ;
Atilhan, Mert ;
Cristancho, Diego ;
Holste, James C. ;
Hall, Kenneth R. .
JOURNAL OF CHEMICAL THERMODYNAMICS, 2007, 39 (08) :1157-1163
[10]   A NEW EQUATION OF STATE AND TABLES OF THERMODYNAMIC PROPERTIES FOR METHANE COVERING THE RANGE FROM THE MELTING LINE TO 625-K AT PRESSURES UP TO 1000-MPA [J].
SETZMANN, U ;
WAGNER, W .
JOURNAL OF PHYSICAL AND CHEMICAL REFERENCE DATA, 1991, 20 (06) :1061-1155