Adaptive calibration of a capacitance tomography system for imaging water droplet distribution

被引:49
作者
Yang, WQ
Chondronasios, A
Nattrass, S
Nguyen, VT
Betting, M
Ismail, I
McCann, H
机构
[1] UMIST, Dept Elect Engn & Elect, Manchester M60 1QD, Lancs, England
[2] Shell Global Solut, Shell Res Ltd, Chester CH1 3SH, Cheshire, England
[3] Twister BV, NL-2289 CC Rijswijk, Netherlands
关键词
capacitance measurement; impedance analyser; sensor calibration; wet gas;
D O I
10.1016/j.flowmeasinst.2004.07.001
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
A highly sensitive electrical capacitance tomography (ECT) system based on an HP4284 impedance analyser has been developed and used to quantify low concentration multi-phase flows in wet gas separation processes. The system hardware provides high accuracy (0.05%) and high resolution (10(-17) F). The sensor was calibrated in an environmental chamber with solid samples of known permittivity over ranges of temperature and humidity. Adaptive calibration and adjacent electrode pair correction techniques were applied to image very low concentration profiles. This paper describes the techniques used and presents the experimental results obtained from a test flow rig called Twister, which has been designed to separate liquid droplets from wet gas streams. The test results over a range of operating conditions (20-95% humidity) demonstrate that the ECT system is capable of reconstructing clear images of the droplet distribution inside Twister. Changes as small as 1 gWater/kgAir in the form of liquid droplets were detectable. It has also been shown that the concentration of the condensable phase can be estimated quantitatively within 20% in comparison with the reference measurements. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:249 / 258
页数:10
相关论文
共 15 条
[1]  
[Anonymous], 2000, HDB CHEM PHYS
[2]   Calibration of capacitance tomography systems for liquid-liquid dispersions [J].
Bolton, GT ;
Korchinsky, WJ ;
Waterfall, RC .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (11) :1797-1800
[3]  
CHONDRONASIOS A, 2001, 2 WORLD C IND PROC T, P573
[4]   A HIGH-FREQUENCY STRAY-IMMUNE CAPACITANCE TRANSDUCER BASED ON THE CHARGE-TRANSFER PRINCIPLE [J].
HUANG, SM ;
GREEN, RG ;
PLASKOWSKI, A ;
BECK, MS .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1988, 37 (03) :368-373
[5]   A dual sensor flow imaging tomographic system [J].
Johansen, GA ;
Froystein, T ;
Hjertaker, BT ;
Olsen, O .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (03) :297-307
[6]  
OKIMOTO FT, 2001, P 51 L REID GAS COND
[7]  
PICKUP E, 1995, P ECAPT 95 BERG APR, P256
[8]   Recent developments and industrial research applications of capacitance tomography [J].
Reinecke, N ;
Mewes, D .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (03) :233-246
[9]  
Reinecke N, 1998, CHEM ENG TECHNOL, V21, P7, DOI 10.1002/(SICI)1521-4125(199801)21:1<7::AID-CEAT7>3.3.CO
[10]  
2-B