Active content determination of non-coated pharmaceutical pellets by near infrared spectroscopy: Method development, validation and reliability evaluation

被引:54
|
作者
Mantanus, J. [1 ]
Ziemons, E. [1 ]
Lebrun, P. [1 ]
Rozet, E. [1 ]
Klinkenberg, R. [2 ]
Streel, B. [2 ]
Evrard, B. [3 ]
Hubert, Ph. [1 ]
机构
[1] Univ Liege, CIRM, Analyt Chem Lab, B-4000 Liege, Belgium
[2] Galephar Res Ctr MF, B-6900 Marche En Famenne, Belgium
[3] Univ Liege, CIRM, Pharmaceut Technol Lab, B-4000 Liege, Belgium
关键词
Near infrared spectroscopy; Process Analytical Technology; Pharmaceutical pellets; Active content; Validation; Accuracy profile; Multivariate analysis; QUANTITATIVE ANALYTICAL PROCEDURES; PROCESS ANALYTICAL TECHNOLOGY; SFSTP PROPOSAL; FT-NIR; HARMONIZATION; REFLECTANCE; STRATEGIES; ASSAY; SPECTROPHOTOMETRY; CHEMOMETRICS;
D O I
10.1016/j.talanta.2009.10.019
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A robust near infrared (NIR) method able to quantify the active content of pilot non-coated pharmaceutical pellets was developed. A protocol of calibration was followed, involving 2 operators, independent pilot batches of non-coated pharmaceutical pellets and two different NIR acquisition temperatures Prediction models based on Partial Least Squares (PLS) regression were then carried out. Afterwards, the NIR method was fully validated for an active content ranging from 80 to 120% of the usual active content using new independent pilot batches to evaluate the adequacy of the method to its final purpose. Conventional criteria such as the R-2, the Root Mean Square Error of Calibration (RMSEC), the Root Mean Square Error of Prediction (RMSEP) and the number of PLS factors enabled the selection of models with good predictive potential However, such criteria sometimes fail to choose the most fitted for purpose model. Therefore, a novel approach based on accuracy profiles of the validation results was used, providing a visual representation of the actual and future performances of the models. Following this approach, the prediction model using signal pre-treatment Multiplicative Scatter Correction (MSC) was chosen as it showed the best ability to quantify accurately the active content over the 80-120% active content range. The reliability of the NIR method was tested with new pilot batches of non-coated pharmaceutical pellets containing 90 and 110% of the usual active content, with blends of validation batches and industrial batches All those batches were also analyzed by the HPLC reference method and relative errors were calculated: the results showed low relative errors in full accordance with the results obtained during the validation of the method, indicating the reliability of the NIR method and its interchangeability with the HPLC reference method. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:1750 / 1757
页数:8
相关论文
共 50 条
  • [1] Moisture content determination of pharmaceutical pellets by near infrared spectroscopy: Method development and validation
    Mantanus, J.
    Ziemons, E.
    Lebrun, P.
    Rozet, E.
    Klinkenberg, R.
    Streel, B.
    Evrard, B.
    Hubert, Ph.
    ANALYTICA CHIMICA ACTA, 2009, 642 (1-2) : 186 - 192
  • [2] Active content determination of pharmaceutical tablets using near infrared spectroscopy as Process Analytical Technology tool
    Chavez, Pierre-Francois
    Sacre, Pierre-Yves
    De Bleye, Charlotte
    Netchacovitch, Lauranne
    Mantanus, Jerome
    Motte, Henri
    Schubert, Martin
    Hubert, Philippe
    Ziemons, Eric
    TALANTA, 2015, 144 : 1352 - 1359
  • [3] Development and validation of a method for active drug identification and content determination of ranitidine in pharmaceutical products using near-infrared reflectance spectroscopy: A parametric release approach
    Rosa, Silvia S.
    Barata, Pedro A.
    Martins, Jose M.
    Menezes, Jose C.
    TALANTA, 2008, 75 (03) : 725 - 733
  • [4] Development and Validation of a Near-Infrared Spectroscopy Method for Multicomponent Quantification during the Second Alcohol Precipitation Process of Astragali radix
    Li, Wenlong
    Luo, Yu
    Wang, Xi
    Gong, Xingchu
    Huang, Wenhua
    Wang, Guoxiang
    Qu, Haibin
    SEPARATIONS, 2022, 9 (10)
  • [5] Development of a comprehensive near infrared spectroscopy calibration model for rapid measurements of moisture content in multiple pharmaceutical products
    Mainali, Dipak
    Li, Jane
    Yehl, Peter
    Chetwyn, Nicholas
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 95 : 169 - 175
  • [6] Validation of a multipoint near-infrared spectroscopy method for in-line moisture content analysis during freeze-drying
    Kauppinen, An
    Toiviainen, Maunu
    Lehtonen, Marko
    Jarvinen, Kristiina
    Paaso, Janne
    Juuti, Mikko
    Ketolainen, Jarkko
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 95 : 229 - 237
  • [7] Determination and Validation of Nimesulide in Pharmaceutical Formulation by Near Infrared Spectroscopy
    Rocha, Werickson F. C.
    Rosa, Andre Luis
    Martins, Jose Antonio
    Poppi, Ronei Jesus
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2010, 21 (10) : 1929 - 1936
  • [8] Development and validation of a method for the polymorphic analysis of pharmaceutical preparations using near infrared spectroscopy
    Blanco, M
    Villar, A
    JOURNAL OF PHARMACEUTICAL SCIENCES, 2003, 92 (04) : 823 - 830
  • [9] Evaluation of Transmission and Reflection Modalities for Measuring Content Uniformity of Pharmaceutical Tablets with Near-Infrared Spectroscopy
    Xiang, Dong
    Lobrutto, Rosario
    Cheney, James
    Wabuyele, Busolo Wa
    Berry, Joseph
    Lyon, Robbe
    Wu, Huiquan
    Khan, Mansoor A.
    Hussain, Ajaz S.
    APPLIED SPECTROSCOPY, 2009, 63 (01) : 33 - 47
  • [10] Determination of the thickness of plastic sheets used in blister packaging by near infrared spectroscopy:: development and validation of the method
    Laasonen, M
    Harmia-Pulkkinen, T
    Simard, C
    Räsänen, M
    Vuorela, H
    EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2004, 21 (04) : 493 - 500