An Intraocular Pressure Measurement Technique Based on Acoustic Radiation Force Using an Ultrasound Transducer: A Feasibility Study

被引:1
作者
Lee, Hee Su [1 ]
Jeong, Eun Young [1 ]
Sung, Jin Ho [1 ]
Choi, Bo Eun [1 ]
Jeong, Jong Seob [1 ]
机构
[1] Dongguk Univ, Dept Med Biotechnol, Seoul 04620, South Korea
基金
新加坡国家研究基金会;
关键词
acoustic radiation force; ultrasound transducer; intraocular pressure; tonometry; eye phantom; CENTRAL CORNEAL THICKNESS; OPTICAL COHERENCE ELASTOGRAPHY; GOLDMANN APPLANATION TONOMETER; ICARE(R) REBOUND TONOMETER; SMALL DISPLACEMENTS; REMOTE PALPATION; GLAUCOMA; ELASTICITY; GUIDELINES; MODEL;
D O I
10.3390/s21051857
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High intraocular pressure (IOP) is one of the major risk factors for glaucoma, and thus accurate IOP measurements should be performed to diagnose and treat glaucoma early. In this study, a novel technique for measuring the IOP based on acoustic radiation force was proposed, and its potential was experimentally demonstrated. The proposed technique uses the acoustic radiation force to generate axial displacement on the ocular surface while simultaneously measuring the degree of deformation. In order to verify that the ocular displacement induced by the acoustic radiation force is related to the IOP, the experiment was conducted by fabricating a 5 MHz single element transducer and gelatin phantoms with different stiffness values. Our experimental results show that there is a close relationship between the ocular displacement by the acoustic radiation force and the IOP obtained by a commercial tonometer. Therefore, the proposed acoustic radiation force technique can be a promising candidate for measuring the IOP.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 54 条
[1]   Comparison of the ICare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists [J].
Abraham, L. M. ;
Epasinghe, N. C. R. ;
Selva, D. ;
Casson, R. .
EYE, 2008, 22 (04) :503-506
[2]   Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients [J].
Brusini, P ;
Salvetat, ML ;
Zeppieri, M ;
Tosoni, C ;
Parisi, L .
JOURNAL OF GLAUCOMA, 2006, 15 (03) :213-217
[3]   Clinical evaluation of rebound tonometer [J].
Davies, LN ;
Bartlett, H ;
Mallen, EAH ;
Wolffsohn, JS .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (02) :206-209
[4]   Central corneal thickness, radius of the corneal curvature and intraocular pressure in normal subjects using non-contact techniques:: Reykjavik Eye Study [J].
Eysteinsson, T ;
Jonasson, F ;
Sasaki, H ;
Arnarsson, A ;
Sverrisson, T ;
Sasaki, K ;
Stefánsson, E .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2002, 80 (01) :11-15
[5]   Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population [J].
Fernandes, P ;
Díaz-Rey, JA ;
Queirós, A ;
Gonzalez-Meijome, JM ;
Jorge, J .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (05) :436-440
[6]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[7]  
GROLMAN B, 1972, AMER J OPT ARCH AM A, V49, P646
[8]   Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes [J].
Herndon, LW ;
Choudhri, SA ;
Cox, T ;
Damji, KF ;
Shields, MB ;
Allingham, RR .
ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (09) :1137-1141
[9]   24-hour monitoring of intraocular pressure in glaucoma management: A retrospective review [J].
Hughes, E ;
Spry, P ;
Diamond, J .
JOURNAL OF GLAUCOMA, 2003, 12 (03) :232-236
[10]   Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness [J].
Iliev, M. E. ;
Goldblum, D. ;
Katsoulis, K. ;
Amstutz, C. ;
Frueh, B. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2006, 90 (07) :833-835