RF electrode-tissue coverage significantly influences steam pop incidence and lesion size

被引:23
作者
Bourier, Felix [1 ]
Popa, Miruna [1 ]
Kottmaier, Marc [1 ]
Maurer, Susanne [1 ]
Bahlke, Fabian [1 ]
Telishevska, Marta [1 ]
Lengauer, Sarah [1 ]
Koch-Buettner, Katharina [1 ]
Kornmayer, Marielouise [1 ]
Risse, Elena [1 ]
Brkic, Amir [1 ]
Reents, Tilko [1 ]
Hessling, Gabriele [1 ]
Deisenhofer, Isabel [1 ]
机构
[1] Tech Univ Munich, Dept Electrophysiol, German Heart Ctr Munich, Lazarettstr 36, D-80636 Munich, Germany
关键词
lesion size; radiofrequency ablation; steam pop; PULMONARY VEIN ISOLATION; CONTACT FORCE; ABLATION CATHETER;
D O I
10.1111/jce.15063
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Steam pops are a rare complication associated with radiofrequency (RF) ablation and are hard to predict. The aim of this study was to assess the influence of coverage between the RF ablation electrode and cardiac tissue on steam pop incidence and lesion size. Methods and results An ex vivo model using porcine cardiac preparations and contact force sensing catheters was designed to perform RF ablations at different coverage levels between the RF electrode and cardiac tissue. During coverage level I, only the distal part of the ablation electrode was in contact with tissue. During coverage level II half of the ablation electrode, and during coverage level III the entire ablation electrode was embedded in tissue. RF applications (n = 60) at different coverage levels I-III were systematically performed using the same standardized ablation protocol. Ablations during coverage level III resulted in a significantly higher rate of steam pops (100%) when compared to ablations during coverage level II (10%) and coverage level I (0%), log rank p < .001. Coverage level I ablations resulted in significantly smaller lesion depths, diameters, and impedance drops when compared to higher coverage level ablations, p < .001. In the controlled ex vivo model, there was no difference in applied contact force or energy between different coverage levels. Conclusions The level of coverage between RF electrode, cardiac tissue, and the surrounding fluid significantly influenced the incidence of steam pops in an ex vivo setup. Larger coverage between RF electrode and tissue resulted in significantly larger lesion dimensions.
引用
收藏
页码:1594 / 1599
页数:6
相关论文
共 16 条
[1]   Impedance, power, and current in radiofrequency ablation: Insights from technical, ex vivo, and clinical studies [J].
Bourier, Felix ;
Ramirez, Francisco Daniel ;
Martin, Claire A. ;
Vlachos, Konstantinos ;
Frontera, Antonio ;
Takigawa, Masateru ;
Kitamura, Takeshi ;
Lam, Anna ;
Duchateau, Josselin ;
Pambrun, Thomas ;
Cheniti, Ghassen ;
Derval, Nicolas ;
Denis, Arnaud ;
Sacher, Frederic ;
Hocini, Meleze ;
Haissaguerre, Michel ;
Jais, Pierre .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (11) :2836-2845
[2]   High-power short-duration versus standard radiofrequency ablation: Insights on lesion metrics [J].
Bourier, Felix ;
Duchateau, Josselin ;
Vlachos, Konstantinos ;
Lam, Anna ;
Martin, Claire A. ;
Takigawa, Masateru ;
Kitamura, Takeshi ;
Frontera, Antonio ;
Cheniti, Ghassen ;
Pambrun, Thomas ;
Klotz, Nicolas ;
Denis, Arnaud ;
Derval, Nicolas ;
Cochet, Hubert ;
Sacher, Frederic ;
Hocini, Meleze ;
Haissaguerre, Michel ;
Jais, Pierre .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2018, 29 (11) :1570-1575
[3]   Same-day discharge following catheter ablation of atrial fibrillation: A safe and cost-effective approach [J].
Creta, Antonio ;
Ventrella, Nicoletta ;
Providencia, Rui ;
Earley, Mark J. ;
Sporton, Simon ;
Dhillon, Gurpreet ;
Papageorgiou, Nikolaos ;
Chow, Anthony ;
Lambiase, Pier D. ;
Lowe, Martin ;
Schilling, Richard J. ;
Finlay, Malcolm ;
Hunter, Ross J. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (12) :3097-3103
[4]   Complications of pulmonary vein isolation in atrial fibrillation: predictors and comparison between four different ablation techniques: Results from the Mlddelheim PVI-registry [J].
De Greef, Yves ;
Stroker, Edwin ;
Schwagten, Bruno ;
Kupics, Kaspars ;
De Cocker, Jeroen ;
Chierchia, Gian-Battista ;
de Asmundis, Carlo ;
Stockman, Dirk ;
Buysschaert, Ian .
EUROPACE, 2018, 20 (08) :1279-1286
[5]   Biophysics of ablation: Application to technology [J].
Haines, D .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (10) :S2-S11
[6]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[7]   High-power short duration and low-power long duration in atrial fibrillation ablation: A meta-analysis [J].
Kewcharoen, Jakrin ;
Techorueangwiwat, Chol ;
Kanitsoraphan, Chanavuth ;
Leesutipornchai, Thiratest ;
Akoum, Nazem ;
Bunch, Thomas J. ;
Navaravong, Leenhapong .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (01) :71-82
[8]   Same-day discharge after cryoballoon ablation of atrial fibrillation: A multicenter experience [J].
Kowalski, Marcin ;
Parikh, Valay ;
Salcido, Jose R. ;
Chalfoun, Nagib ;
Albano, Alfred ;
O'Neill, Padraig G. ;
Bowers, Mark R. ;
Gauri, Andre ;
Braegelmann, Kendra M. ;
Lim, Hae W. ;
Akhrass, Philippe ;
Shah, Rina ;
Bekheit, Soad ;
Epstein, Laurence M. ;
Aryana, Arash .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (02) :183-190
[9]   Relationship between the ablation index, lesion formation, and incidence of steam pops [J].
Mori, Hitoshi ;
Kato, Ritsushi ;
Sumitomo, Naokata ;
Ikeda, Yoshifumi ;
Goto, Koji ;
Tanaka, Sayaka ;
Asano, So ;
Tahara, Mai ;
Nagase, Takahiko ;
Iwanaga, Shiro ;
Muramatsu, Toshihiro ;
Matsumoto, Kazuo .
JOURNAL OF ARRHYTHMIA, 2019, 35 (04) :636-644
[10]   Randomized comparison of contact force-guided versus conventional circumferential pulmonary vein isolation of atrial fibrillation: prevalence, characteristics, and predictors of electrical reconnections and clinical outcomes [J].
Nakamura, Kohki ;
Naito, Shigeto ;
Sasaki, Takehito ;
Nakano, Masahiro ;
Minami, Kentaro ;
Nakatani, Yosuke ;
Ikeda, Kentaro ;
Yamashita, Eiji ;
Kumagai, Koji ;
Funabashi, Nobusada ;
Oshima, Shigeru .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2015, 44 (03) :235-245